| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹7.1 LAccepted-AOC | L1 | Accepted-AOC OK | |
| 2 | L2₹9.7 L+₹2.6 L (36.5%)Rejected-Finance | L2 | Rejected-Finance REJECT | |
| 3 | L3₹10.2 L+₹3.1 L (43.1%)Rejected-Finance | L3 | Rejected-Finance REJECT |
Tender Value
₹9.3 L
EMD Value
₹18,520
Closing Date
1 Jul 2020, 3:00 pmClosed
EXECUTIVE ENGINEER
28 PARK ROAD AMBALA CANTT
Rural NRDWP WQM and S Ambala PHED Supplying of Laboratory Instruments to make PH Lab at Ambala Cantt upto NABL accreditation and all other works contingent thereto.
2020_HRY_133237_1
SE MEMO NO. 5007 DATED 30.04.2020
Open Tender
Civil Works
Works
60 days
Rural NRDWP WQM and S Ambala PHED Supplying of Lab
Rural NRDWP WQM and S Ambala PHED Supplying of Laboratory Instruments to make PH Lab at Ambala Cantt upto NABL accreditation and all other works contingent thereto.
2 documents required · 2 mandatory
₹1,000
Engineer in Chief
₹18,520
Yes
2 Sept 2020
24 Jun 2020
1 Jul 2020
24 Jun 2020
1 Jul 2020
24 Jun 2020
Amount
Supply of Turbidity Meter Make- Lovibond/ HACH complete in all respect (with calibration certificate). Technical specification given as under:- Measurement cycle approx. 8 seconds Display backlit LCD (on keypress) Optics temperature-compensated LED ( l= 860 nm) and photosensor amplifier in water proof sample chamber, infrared light Keypad polycarbonate membrane, splash proof Power supply 9 V power pack battery Auto - OFF automatic switch-off Storage internal ring memory for 16 data sets Additional feature real time clock and date Range 0,01 - 1100 NTU Resolution 0.01 - 9.99 NTU = 0.01 NTU 10.0 - 99.9 NTU = 0.1 NTU 100 - 1100 NTU = 1 NTU Accuracy ± 2,5 % of reading or ± 0.01 NTU (0 - 500 NTU) ± 5 % (500 - 1100 NTU) Housing ABS Ambient conditions Temperature: 5 – 40 °C rel. humidity: 30 – 90% Confirmatory CE
Supply of Ph,Conductivity,TDSMake- Lovibond/ HACH complete in all respect (with calibration certificate). Technical specification given as under:- Parameter Range, Resolution, Accuracy pH 0-14, 0.01, ± 0.02 + 2 digits ORP -1999 to +1999 mV, 1 mV, ± 0.5% + 2 digits 0-200 μS/cm, 0.1 μS/cm, ± 2% + 1 digit 200-2000 μS/cm, 1 μS/cm, ± 2% + 1 digit Conductivity 20-200 mS/cm, 0.1 mS/cm, ± 2% + 1 digit TDS 0-132 mg/l, 0.1 mg/l, ± 2% + 1 digit Technical specification given as under:- 132-1320 mg/l, 1 mg/l, ± 2% + 1 digit TDS 1320-13200 mg/l, 10 mg/l, ± 2% + 1 digit TDS 13200-132000 mg/l, 100 mg/l, ± 2% + 1 digit Ambient Conditions 0-50 ° C; 0-80% rel humidity (non condensing) Display Size 58mmX34 mm Power Supply Battery 9V & Adapter Dimensions 208 x 110 x 34 Weight 380g InterfaceUSB, RS 232 Datalogger Real Time Data Logger Software Software Data Memory Auto or Manual date memory 16000 data sets only with additional software Confirmatory CE Protection Class IP51Auto Off Yes.
Supply of Capacity-220 g, Minimum disply-0.1 mg, Repeatability (standard deviation) < 0.1 mg, Linearity - + 0.3 mg, Response time-3 Secs, UniBloc weighing technology, Build-in calibration, Built-in clock, GLP / GMP / ISO calibration report, Analog bar graphic display, Specific gravity measurement mode, below weight measurement, Formulation mode, Formulation mode- included, Operating temperature range-5 to 40 degree C, Sensitivity stability against temperature range -12 ppm / 0 C (10 to 30 degree C), pan size- Approx, Dia. 80mm, Body dimensions WxDxH- Approx. 220 x 330 x 310 mm, Weight- Approx. 7.0 Kg, Backligh display- Not Available, Input / Outpur terminal-RS232 I/F (with calibration certificate). Auto Internal Calibration, Touch-key Calibration. RS-232C interface GLP/GMP/ISO calibration report Interval time Specify gravity measurement {iece counting and unit conversioon.
Supply of Water bath 12 holes (with calibration certificate). Supply of Medium size at least with 12 holes.
Supply of Magnetic stirerMedium size.
Supply of Weight box F class (with calibration certificate) with calibration certificate 20 gm + certificate.
Supply of Double door fridge. 300liter of standard brabd, double door.
Supply of Hot Air Oven Ambassador company (with calibration certificate).Hot Air Over-Make ISO certified company (Specifications) Innder chamber size: 18" x18"x18" (LxWxD) Double walled inner chamber available in stainless steel with grade-304. Outer chamber made of stainless steel (Mild steel) Glass wool / PUF insulation to prevent tempereature loss. Outer double wall door in provided with gas kit Stainless steel removable & adjustable shelves. No. of shelves-3 Nos. Digital display with mains On/Off, swith indicating lamps & fuse. Digital microprocessor based PID Temp controller with LED display. Accuracy of temperature +/- 0.5 degree. Temperature range adjustable from Room Temp. to 2600 C (resoluation 0.10 C) Air flow systems provided blower for even distribution of temperature. Tt. Safety tehemostat to prevent over heat. Electric supply : 220 / 230 V AC, 50 / 50 Hz, 2500 watt. Calibration certioficate from NABL certified Lab required.
Supply of Voltage steplizers 0.5 kilowatts
Supply of Heating mental (with calibration certificate).Heating mantal capacity : 1 Ltr. Make - Any standard Made
Supply of Mambarain filter (with calibration certificate).Membrance filtration assembly complete consist of: Filter funnel glass with aluminium clamp, neoprene cover & cork capacity-250 ml (FFG47) Membrance filter C.N. 47 mm / 0.45 um pack of 100 Vacuum flask one litre Vacuum pump disphragm type-oil free. Air vent filter-Hydrophobic PTFE 50mm / 0.2um PTFE syringe filter 25mm / 0.2 um with needle. Flexible silicon tubino 1 Mtr. Walfhuegel glass grid with focusing facility (Auto marker pen)
Supply of Digital auto burette (with calibration certificate) (Bottle top burette). Digital auto burette (Bottle top burettle) Make: Mettler Toledo / Tarson Cap. 10 Ml. Self Zero Discription Precision titrations within the error limits of Class-A Large, easy-grip hand wheels for sensitive, drop-wise titration. Titrating tube can be adjusted horizontally and vertically-providding flexibility when positioning the instrument. No switching needed when changing between filling and titration. Calibration scheduler saves a date for the next calibration, Auto-Power-Off, change decimal place settings. Easy cleaning, maintenance and part replacement. NOTES :- i) AMC/CMC any type of part of instruments shall be cover within 2 years of warranty. ii) GST extra applicable as per Govt. Rules.
RAJAN SCIENTIFIC TRADERS
DARSHAN ENGINEERING WORKS
stage.html
html • 0.04 MB
tech_eval.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .