Tender ID
Eligibility, documents, scope and key termsEligibility, required documents, scope of work and key terms of this tender, in one click.
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹39.5 LAccepted-AOC 9TH FLOOR C 0901 TOWER C ATS BOUQUET PLOT A2 2 SECTOR 132 NOIDA GAUTAM BUDDHA NAGAR UTTAR PRADESH 201305 | NOIDA | GAUTAM BUDDHA NAGAR | UTTAR PRADESH | 201305 | L1 | Accepted-AOC Excluding GST | |
| 2 | L2₹89.3 L+₹49.3 L (123.1%)Rejected-Finance | L2 | Rejected-Finance as per the Purchase Committee decision | |
| 3 | Rejected-Technical EL 91 TTC INDUSTRIAL AREA ELECTRONIC ZONE MIDC MAHAPE NAVI MUMBAI THANE MAHARASHTRA 400710 | THANE | MAHARASHTRA | 400710 | - | Rejected-Technical technically not qualified |
Tender Value
Refer Docs
EMD Value
₹45,000
Closing Date
8 Feb 2021, 3:00 pmClosed
Eligibility, documents, scope and key termsEligibility, required documents, scope of work and key terms of this tender, in one click.
Registrar
Kerala Forest Research Institute , Peechi P.O., Thrissur District, Kerala, Pin-680653
Bench-top X-ray Diffractometer
2021_KCSTE_407817_1
KFRI/174/2021-E4-01
Open Tender
Equipments
Item Rate
15 days
KFRI, Peechi
Please refer Tender documents.
5 documents required · 5 mandatory
₹6,750
Yes
₹45,000
Yes
6 Sept 2021
21 Jan 2021
9 Feb 2021
21 Jan 2021
8 Feb 2021
21 Jan 2021
180 days
Amount
Bench-top X-ray Diffractometer
Compact Bench-top X-ray Diffractometer System. The instrument specifications are outlined below. X-ray Generator Tube load: 300 Watts or higher Tube Voltage: Min 40 kV (Variable) or Higher Current: Min 7.5 mA (Variable) or Higher Stability: +/-0.05 % or better (for a ± 10% variation in line voltage) Tube cooling provisions (external or internal). Gonimeter High Precision, vertical type with theta - theta geometry or theta – 2 theta geometry Range: –3° to 140° 2Theta or better Radius: 140mm or more Scanning Speed: 0.01 to 100 degree/min Step Size: ≤ 0.005 degree two theta or better Necessary beam slits and soller slits for Bragg-Brentano Geometry measurements should be provided. Suitable Ni K-beta filter should be provided. Sample Holders : - Standard sample holders for analysis of Powder and Small Solids. Minimum 25 pcs should be provided with the instrument. The sample holders should be re-usable and long lasting. Detector High speed High Resolution Solid State based detector or equivalent Detector should work in 0D and 1D modes Detector shouldn’t require any external cooling. Software & Database The software should be capable of simultaneous Diffractometer control, data collection and analysis, peak search, search-match and elaborate pattern treatment such as data smoothing, background subtraction, Kα2 stripping, crystallinity determination etc. Crystallographic and Rietveld based quantitative phase analysis software. Software should cover: Qualitative analysis Phase quantification Peak/Line Profile analysis Lattice parameter determination Crystallite size determination Percentage of Crystallinity Determination Phase quantification based on Rietveld method Unit Cell determination Space Group Determination COD database should be compatible with the OEM Analysis Software and should be provided with the instrument Standard - Si Standard (NIST Traceable) Computer Branded PC with Windows 8 Professional or 10 Licensed version of latest type with 22” LED monitor (or better) suitable for the above XRD applications Color Printer. Safety features The system should be completely safe for x-ray radiations and should comply to all international safety norms for x-ray safety, electrical and mechanical safety The XRD should be type approved by AERB. Type Approval should be provided with the bid. Warranty and Maintenance contract - 3 years warranty and 2 years Annual Maintenance Contract post warranty including maintenance kits for the entire instrument Power supply Single Phase, 200 to 230 V, 50/60 Hz Installation Commissioning and Training After receipt of the item at purchaser’s site the complete system shall be integrated installed and commissioned at the designated place (at purchaser’s site) by vendor’s representative. The vendor’s representative should also provide complete hands-on training to the purchaser after installation and commissioning. User Base – The Vendor should be in active business of XRD system in India and should have carried out minimum 20 installations of XRD instruments at various institutions in India. The Vendor should submit their users list for the quoted model with year of installation, addresses, e-mail, phone/mobile nos. etc.
Astron Electronic Corporation
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart shows the information on this page on a best-effort basis and does not guarantee that it is accurate, complete or current. Check it against the buyer's official notice and documents before you act on it; the official record prevails. TenderKart is not liable for any loss or claim arising from the use of this information. If you notice any error or omission, please let us know at .