Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹1.1 LAccepted-AOC | L1 | Accepted-AOC L1 | |
| 2 | Rejected-Technical | - | Rejected-Technical as per technical evaluation report |
Tender Value
₹1.2 L
EMD Value
₹2,925
Closing Date
3 Jun 2023, 6:00 pmClosed
Director
Malabar Cancer Centre, Thalassery
Rate contract for validation of Operation Theatres, clean rooms and medical equipments
2023_MCC_567443_3
MCC/585/2021-SUP-EL
Open Tender
Miscellaneous Works
Item Rate
365 days
Malabar Cancer Centre, Thalassery
Please refer Tender documents.
5 documents required · 5 mandatory
₹500
Yes
₹2,925
Yes
Malabar Cancer Centre, Thalassery
18 Mar 2024
25 May 2023
6 Jun 2023
25 May 2023
3 Jun 2023
25 May 2023
30 May 2023
Amount
Validation test for HEPA Filter Integrity Test, Air Particle count test, Air Velocity measurement, Air changes calculation tests Differential Pressure Measurement, and Fresh Air Calculations. The test should be as per ISO 14644-1, 2 & 3 and NABH norms. Detailed in specific conditions
Operation theatre
BMT Room
Validation test for Air Velocity, Differential Pressure Measurement, filter leakage Test, and Particle count test for cleanliness level of the following equipment as per ISO Class 5 - ISO 14644-1:2015, 2 & 3 and NABH norms
Cytotoxic Cabinet
Biosafety cabinet
Validation test of Air Flow velocity, Airflow pattern test, differential pressure test, Filter leakage test, and Particle test of the following equipment as per ISO class 5- ISO 14644-1:2015,2,&3 and NSF 49/EN12469 and NABH norms.
PCR Cabinet
Pass box
Validation test of Air Velocity Test, Differential Pressure Measurement, filter leakage Test, and Particle count test (Test should conduct as per ISO Class 5 - ISO 14644-1:2015, 2 & 3) as per NABL norms. Detailed in clause no.2 of the scope of work in the Tender document
Cytotoxic Cabinet
Biosafety cabinet
Validation test of Air Flow velocity test (Downflow/Inflow), Airflow pattern test, differential pressure test, Filter leakage test, and Particle test for following equipment as per ISO class 5- ISO 14644-1:2015, 2 & 3 and NSF 49/EN12469 and NABL norms
Laminar Air Flow Cabinet
PCR Cabinet
The rate for replacing the following size HEPA filters of AHU in OT & BMT Rooms as per standard procedures. The rate includes all costs of labour, materials, and scaffolding
Dimension600x600x150 MM
Dimension 450mm x 450mm x305 MM
The rate for leak-arresting work in case validation failed due to leakage
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .