| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹56.4 LAccepted-AOC 32 FATEH ENCLAVE OFFICERS ENCLAVE PHASE II PATIALA 147001 PB | PATIALA | PATIALA | PUNJAB | 147001 | L1 | Accepted-AOC Being lowest | |
| 2 | L2₹60.4 L+₹3.9 L (7.00%)Rejected-Finance 303 GHS 888 SECTOR 20 PANCHKULA 134116 HARYANA | PANCHKULA | PANCHKULA | HARYANA | 134116 | L2 | Rejected-Finance Being not lowest | |
| 3 | L3₹66.4 L+₹10.0 L (17.8%)Rejected-Finance 154 GANDHI MARKET AMBALA CANTT HARYANA 133001 | AMBALA CANTT | AMBALA | HARYANA | 133001 | L3 | Rejected-Finance Being not lowest |
Tender Value
Refer Docs
EMD Value
₹1.0 L
Closing Date
23 Sept 2021, 6:00 pmClosed
Eligibility, documents, scope and key termsEligibility, required documents, scope of work and key terms of this tender, in one click.
CWE Ambala
Headquarter Commander Works Engineer Ambala Mansfield Road Ambala Cantt
PROVISION OF INFRASTRUCTURE (B/R AND E/M) FOR THE 250 LPM (EX TURKEY) CAPACITY PSA OXYGENERATION PLANT NEAR BLDG NO. P 4 (FAMILY WARD II) UNDER GE (UTILITY) AMBALA CANTT
2021_MES_472043_2
CWE/AMB/UTILITY/T-37 OF 2021-22
Limited
Electrical and Maintenance Works
Item Rate
90 days
Ambala Cantt
All as per tender documents
9 documents required · 9 mandatory
₹1,000
GE UTILITY AMBALA CANTT
₹1.0 L
Yes
28 Sept 2021
18 Sept 2021
25 Sept 2021
18 Sept 2021
23 Sept 2021
18 Sept 2021
18 Sept 2021 - 18 Sept 2021
60 days
Amount
SCHEDULE A PART I (B/R WORKS) Excavation in trenches not exceeding 1.50 metre wide and not exceeding 1.50 metre in depth for foundation etc or shaft, wells, cesspits, manholes, pier holes, cable laying in trench etc n.exc 10sqm on plan and n.exc 1.5mtr in depth and getting out in soft/loose soil complete all as specified and as directed.
Material and labour for PCC (1:4:8) type D2 (using 40mm graded stone aggregates), 75 mm thick, as in foundation filling and mass concrete complete all as specified and as directed.
Returning filling in foundation/trenches including spreading, leveling, watering and well ramming in layers not exceeding 25 cm thick in soft loose soil complete all as specified and as directed.
Removing surplus excavated soil/materials to a distance not exceeding 100 metre and depositing where directed at level not exceeding 1.5 metre from the starting point complete all as specified and as directed.
Material and labour for brickwork, with subclass 'B' bricks, straight or curved on plan to any radius, exceeding 6 metre mean radius built in cement mortar 1:6 complete all as specified and as directed.
Material and labour for rendering 15 mm thick in cement mortar 1:4 with water proofing compound @2% by weight of cement including finished the surface even and smooth without using extra cement complete all as specified and as directed. Note: (i) WPC shall be measured and paid separately. (ii) To use only Yamuna sand.
Material and labour for preperation of newly plastered surfaces of walls and applying two coats of cement base paint complete all as specified and as directed.
Supply only integral water proofing compound complete all as specified and as directed.
Supply and fixing of framed work such as in hung shutter, gates including pintle hinges, stops, hand made sliding door bolts, grills,gratings, etc. with ends of bars shouldered and/or riveted,or forged into spikes; framedguard bars; barred iron doors; ladders; framed balusters; walkways; railings; framework of water tanks and similar work con- forming to Fe.290.Gde-E-165 complete all as specified and as directed.
Preparing surfaces of wood and wood based materials, steel and iron work, new or old treating such as surfaces of hold fast, etc. area tarred n.exc 0.1sqm each, treated with tar or coal tar black paint complete all as specified and as directed.
Material and labour for PCC (1:2:4) type B1 using 20mm graded stone aggregates as in lintels upto 1.5m clear span, cills, steps; seismic and other similar bands, plinth courses, string courses, lacing courses, parapets and railings upto 60 cm in height, copings, fixing holdfasts, kneelers, apex stones, bed plates, water troughs and the like including use and waste of formwork including weathering, slightly rounded or chamfered angles and throating, as directed complete all as specified and as directed.
Material and labour for preparation of new steel surfaces, of any description not otherwise described, over 10 cm width or girth and applying two coats of approved colour synthetic enamel paint, over a coat of red oxide zinc chrome primer complete all as specified and as directed.
Material and labour for bands, straps, hold fasts; clamps, supports or stays for rainwater, ventilating pipes, including bolts, if required: hooks (stay), heavy, for large gates; hook and pintle hinges: large hasps and staples; plate washers and similar small articles conforming to Fe-290 Gde-E-165 complete all as specified and as directed.
Material and labour for hardcore of gauge not exceeding 63mm, deposited, spread and levelled in layers not exceeding 15cm thick, watered and rammed to a true surface complete all as specified and as directed.
Material and labour for PCC (1:4:8) type D2 (using 40 mm graded stone aggregates), 100 mm thick, as in sub base of floors, finished rough and even to receive floor topping complete all as specified and as directed.
Material and labour for PCC (1:2:4) type B0 (using 12.5 mm graded stone aggregates), 25 mm thick, as in levelling course, laid and compacted in slope in floors, finished rough and even to receive floor topping complete all as specified and as directed.
Material and labour for machine cut kota stone slab 16-20 mm thick ( size of each stone not less than 55cm x 55cm), laid in flooring, over and including cement screed bed 15mm thick in cement mortar (1:4) base and jointed & pointed with grey cement slurry mixed with pigment to match shade of the slab including cutting, waste, rubbing and polishing (Granite finish) complete all as specified and directed.
Excavating in post holes (or similar holes) each not exceeding 0.5 cu. m, getting out, returning, filling in and ramming earth or broken rubble mixed with earth around, in layers not exceeding 25 cm, watering, and removing surplus soil to a distance n exc 50 m and making good surface complete all as specified and directed.
Material and labour for PCC (1:2:4) type B2 using 40mm graded stone aggregates as in foundations, filling and mass concrete complete all as specified and directed.
Supply and fixing of pre-painted galvalume, aluminuim zinc coating, GI based corrugated steel sheet, 0.50mm thick, AZ-150 conforming to IS-15961, of red colour, having tensile strength of 550 MPa, as in roof covering/cladding to wall, fixed with self tapping screws complete all as specified and as directed. Note- The sheets shall have super durable polyester quality paint of approved colour, 150 gm/sqm total on both sides, suitable for exterior application conforming to IS-15965.
Material and labour for rolled mild steel sections (channel, angle or tee sections) fixed independently, cut to length, as in hand rails, half round or other stock section or core for wooden hand-rail straight portion, drilled and countersunk for balusters running bars, for doors, balusters, nosing to steps including embeding in concrete and screwing etc. wind ties for roof including securing bolts;posts and struts in fencing including cutting tops to shape as directed, drilling holes, notching for wire, etc.; flat iron fillets (for securing wire mesh etc.) including all necessary drilling, bolting or riveting, etc. and similar work conforming to Fe-290 Gde E-165 complete all as specified and directed.
Material and labour for rolled mild steel sections as in main and cross beams, hip and jack rafters; purlins where connected to common rafters and running rails and guides for sliding doors; fixed with angle cleats or other connecting plates conforming to Fe 410-W (Gde-E- 250) quality-A complete all as specified and directed.
Material and labour for rolled mild steel sections as in plates (plain or chequered) channels, square cut or notched (ordinary and counter sunk) forming holes, without any attachment conforming to Fe 410-W (Gde-E- 250) quality-A complete all as specified and directed.
Material and labour for mild steel screw bolts (machine made) with hexagonal or square heads at one end and screwed the other, or screwed both ends, each screwed end to be fitted with washers and one hexagonal or square nut, exceeding 5cm and not exceeding 10cm long with dia of neck 11 to 15mm complete all as specified and directed.
SCHEDULE A PART II (MEDICAL GAS PIPELINE SUPPLY SYSTEM) (MGPS System): Refer PS Clauses hereinafter for notes, instructions and responsibility of bidder and for design, supply, handling, installation, testing and commissioning of complete Medical Gas Pipeline Supply (MGPS) system including necessary accessories required in coordination with the GE and Users for this MGPS System work
Supply, installation, commissioning and testing of Oxygen Manifold: Main confirming to ISO standards with Middle Frames 2 Banks x 6 Cylinder per Each bank Oxygen Manifold (without cylinders), suitable to withstand a pressure of 145 Kg/cm2, along with high-pressure copper annealed tail pipes with end Brass adapter suitable for Oxygen Cylinders and manifold.Top frame comprising of high pressure copper pipes of size 5/8” I.D. x 15swg as per BSEN 1057 with high pressure brass fittings made of high tensile brass equally spaced at 250mm fitted on powder coated MS frame and connections through non- return valves; 1.25 metre long high pressure copper tail pipes, made of high pressure copper pipe of size 5/16” I.D. x 16 swg. The design of middle and bottom frames should be provided to fit both round and flat bottom cylinders safely. The manifold shall be tested (hydraulically) at 300 bar and necessary test certificates accompany along with the supply. Only Non Halogenated Polymer materials are to be used in the Non Return Valves supplied along with the manifold. The middle frame shall be powder coated with cylinder holding chains for holding cylinders safely complete all as directed by Engineer-in-charge. Make: Oxicareengineers/Globalmedlife Solution/Lifeline Medisol/ Aktiv/ LMPL/ MK/ BSI/ MR Engg
Supply , installation, commissioning and testing of Oxygen Manifold: Main confirming to ISO standards with Middle Frames 2 Banks x 2 Cylinder per Each bank Oxygen Manifold (without cylinders), suitable to withstand a pressure of 145 Kg/cm2, along with high-pressure copper annealed tail pipes with end Brass adapter suitable for Oxygen Cylinders and manifold.Top frame comprising of high pressure copper pipes of size 5/8” I.D. x 15swg as per BSEN 1057 with high pressure brass fittings made of high tensile brass equally spaced at 250mm fitted on powder coated MS frame and connections through non- return valves; 1.25 metre long high pressure copper tail pipes, made of high pressure copper pipe of size 5/16” I.D. x 16 swg. The design of middle and bottom frames should be provided to fit both round and flat bottom cylinders safely. The manifold shall be tested (hydraulically) at 300 bar and necessary test certificates accompany along with the supply. Only Non Halogenated Polymer materials are to be used in the Non Return Valves supplied along with the manifold. The middle frame shall be powder coated with cylinder holding chains for holding cylinders safely complete all as specified and directed by Engineer-in-charge. Make: Oxicareengineers/Globalmedlife Solution/Lifeline Medisol/Aktiv/LMPL/MK/BSI/MR Engg
Notes for item No.26 and 27 : The sizes of copper pipe mentioned are of minimum size and the quoted rate deemed to be inclusive of provn of higher sizes if any as per OEM requirement.
Supply, installation, testing and commissioning of Fully Automatic Oxygen Control Panel with automatic changeover cover made up of high quality Stainless steel, with built in LED display to indicate high normal and low line pressure, Capable to supply and distribution flow rate in excess of 250 LPM. Panel cover fitted with lock to prevent unauthorized access and can be swung open for maintenance. The panel includes pressure gauge (63 mm) that indicates the gas pressure of each header and pipe line distribution pressure form Running Bank to Reserve Bank of Cylinder with timer complete all as directed by Engineer-in-charge.
Notes: (a) Automatic control system consists two first stage regulators and consisting high pressure reducer panel, Alarm and control panel, left / right Ramp pressure sensors, pipeline pressure sensor, high pressure input valve, input filter, 315 Bar input gauge, high pressure reducer body, pressure set up screw, high pressure reducer drain, high pressure gauge, 2nd stage regulator, regulator set up, automatic change -over, automatic output pressure gauge in emergency outlet, back up system input, plant output regulators, plant output pressure gauges, plant output connected piping, control/power wiring and all necessary arrangement complete and should comply Indian Medical Standards and should comply standard : HTM 02-01/DIN/EN/ISO-7396-1 and should be CE certified.
(a) 40 micron filter should be provided at the inlet of each high pressure regulators of the oxygen control panel.
(b) The control panel will provide for two individual content contact pressure gauges to indicate the cylinder pressure in the two wings of the manifold and common pressure gauge to indicate the delivery / line pressure.
(c) The control panel will have built in audio-visual signal lamp indications for bank Changeover.
(d) The control panel will be covered with aesthetically suitable cover for safe operation indicating the respective services.
(e) Control panel will have built in protection system to ensure safe operation by low voltage and all protections of surge, overload and earthfault. The Control panel shall include two pressure relief valves, one high pressure and one low pressure.
(f) The panel should have the capacity to withstand maximum operating pressure of 220 bar for safety purposes.
(g) The main control unit must be protected by a mild steel with epoxy powder coated with a viewing window to monitor the digital display. Make: MEC Medical Equipment Co Ltd UK, Reanimed, CPX (UK), AKTIV, MPS
Supply, Installation, testing and commissioning of oxygen gas outlet points complete with matching probe/adopter with one end suitable for Hose & other end suitable for terminal units complete set for Oxygen and Vacuum system as per technical specification confirming to CE certified/ HTM 02-01/EN/ISO/DIN and NFPA 99: 2005, HTM 01-02, BS EN ISO 9170-1: 2008 British Standard as mentioned below all as directed by Engineer-in-charge:- (a) The outlets shall be modular singles and shall incorporate name plate color-coding. The single modular outlets shall be of design that provides for ganging of rough-in plates in the field to form multiples. (b) The gas services shall be sequentially arranged and located as shown on the plans with a minimum centerline spacing of 5 inches (12.5 cm) between outlets. (c) The outlets shall be capable of supporting dispensing equipment without the use of slide brackets. The medical gas outlets shall be designed so that, once installed, routine service of both the primary and secondary check valves can be accomplished without removing the nameplate or gas specific portions.
(d) The primary check valve shall be unitized and of the cartridge type so that the primary check valve can be removed for service without shutting off the gas supply to the outlet, the secondary check valve shall operate automatically to stop the free flow of the pressure gas. (e) There shall be no secondary check for vacuum or gas evacuation service. Medical gas outlets that require the removal of the nameplate or gas specific components for routine service shall not be acceptable. (f) The outlets shall conform to CE certified (CE 0482)/ HTM 02-01/EN/ISO/DIN and should meet with the provisions of Council Directive of Medical Devices Directive. (g) All outlets are 100% tested for flow and leaks. (h) The outlet nameplate shall be permanently color coded with a durable scratch-resistant and protective label. (j) Medical gas outlets shall be cleaned for oxygen service as per Medical Standards, capped and placed in a protective container for shipment. (k) The outlets shall be installed in strict accordance with manufacturer’s instructions and tested before use in conformance with local and federal codes.
(l) Outlets shall be Double Lock Outlet type manufactured with an approx 165 mm length, Copper inlet pipe stub which is silver brazed to the outlet body. Body shall be of one piece brass construction. For positive pressure gas services, the outlet shall be equipped with a primary and secondary check valve and the secondary check valve shall have break safe mechanism and also comply to pressure test standards and rated at minimum 200 psi in the event the primary check valve is removed for maintenance.The outlet assembly should have separate colour coding for each services and will accept only corresponding gas specific adaptors. All outlets shall be cleaned and de-greased for medical gas service, factory assembled and tested. The medical gas outlets should be of quick connecting and wall mounted modular type. Make: MEC MEDICAL EQUIPMENT CO LTD UK/REANIMED CPX (UK), AKTIV, MPS
All as per srl item no 39 to 40 here-in-before but Vacuum outlet
Supply and installation BPC Flow meter (0-15 LPM) with Humidifier Bottle, 600 ml reuseasle collection Jar, unbreakable polycarbonate material and autoclavasle at 134 degree centigrade temperature and flow meter made up of chrome plated brass body graduated flow tube cover and humidifer bottle made of unbreakable polycarbonate with adjustable flow by needle valve with micrometrical regulation and the range should be controlled with in a range of 0-15 lpm (Calibration with in +/- 10%) complete all as directed by Engineer-in-charge.
Supply, installation, testing and commissioning of Medical Vacuum Plant Reciprocating Vacuum 250 LPM Duplex system (1 working + 1 standby), with reciever tank 750 Ltr, bacteria filters, vacuum filters complete all as per technical specifications and as per the following all as directed by Engineer-in-charge :-
(a) Vaccum Pumpos Single stage, minimum capacity 60 cfm (piston displacement) or 5.0 HP each motor. Each pump should be suited for maximum working pressure of 29 inch mercury (650 mm mercury) Each pump should be fitted with MS channel frame complete having vibration isolators with 3 Nos.silencers at vacuum pump discharge end. The pump should be air cooled. 3 sets of isolation valve, on return valve and filter should be provided with pumps and the system should be expandable to receive one additional pump in future. (b) Drive Assembly :The electrical drive for vacuum pump should consist of suitable TEFC induction motor operating at 1440 RPM and connected through a V-belt. The motor should be suited for 415 V +10% 50 HZ.3 Phase AC supply.Capacity : 5 HP-2 Nos (1 working+1 standby) Make-Crompton Greaves / Kirloskar/OEM approved, 3 phase, Starter - Control & Switchgear/ L & T/BCH/Havells/Schinider/OEM approved with overload protection, panel for automatic switching of pumps. (c) Vacuum Receiver : 1 No vacuum receiver of 750 liters capacity The receiver should be fabricated from MS plates as per IS:226 and manufactured as per IS:2825.The receiver should be supplied with safety valve, pressure gauge and drain valve etc.manufacture’s test certificate to be provided. Reciever should have feature to pass autocut signal to the plant in case it is full.
(d) Vacuum Switch :Vacuum switch of 0-760 mm range should be provided, should have both NO/NC contacts and suitable for 230 V 50 Hz AC supply. The vacuum switch should be housed in weather proof casing with 3/8”BSP gas service connection at the bottom of the assembly. (e) Vacuum Filter :2 Nos vacuum filters with drain valve. This filter should be non-ferrous and online. The discharge to autosphare should be through micron filter cartridge, and suitable capacity bacteria filter & secretion trap (f) The medical vacuum plant shall be CE Marked/ISO comprising of air cooled, oil lubricated rotary vane/reciprocating vacuum pumps to provide a flow rate. Make: Anesta / Wata /Ingersol Rand/ Boge/Coaire/Frank
Suply, installation testing and commissiong of Ward suction unit or Ward Vacuum Unit with above 1000 ml suction collection jar-1 Nos as per specifications mentioned below all as directed by Engineer-in-charge:-
(a) The vacuum regulator should have a vacuum gauge with analog display.The scale should be in mbar or mmhg as required by the end user. (b) 1-No of Suction Regulator and 1-No of 1000 ml polysulfone /polycarbonate collection jar. (c) Suction regulator should be supplied with a safety jar, including an antibacterial filter and an anti-overflow safety device and should have wide membrane continuous suction controller. (d) Should have vacuum levels: 0-760 mm of Hg and Should have vacuum gauge fitted with a protective bumper device. (e) Should have on/off knob allowing for the quick restoration of a readjusted vacuum level. (f) Must have central adjustment knob with a color coded for 0 to 760 mm of Hg. Should have Polysulfone/polycarbonate 1700cc safety jar, autoclavable at 121º C at 5mins, unbreakable, fitted with an anti-overflow safety device and equipped with a plastic antibacterial filter and should be totally transparent, to ensure perfect sucked liquid visibility. Make- Aneer/SSS Tech/LMPL
Supply, Installation, testing and commissiong of Copper Piping of following dia and thickness with complete fittings/fixtures as per technical specification and BS: 13348 and as mentioned below all as directed by Engineer-in-charge:- (a) Installation of piping should be carried out as per approved medical standards with utmost cleanliness. (b) Only pipes, fittings and valves which have been degreased as per approved medical standards should be used . (c) Pipe fixing clamps for all diameter pipes. (d) Pipes should be non ferrous suitable for the diameter of the pipe and wherever required rigid metallic hanging or cemented supports to be used. (e) The main lines to the building to be taken overhead through heavy grade metallic poles or GI tubes with inspection points in metallic boxes wherever required . Adequate supports shall be provided while laying pipelines to ensure that the pipes do not sag. (f) All copper pipes and fittings should be made of phosphorous de-oxidized, non arsenic, and fully degreased. All pipes should be of Copper and copper alloys Seamless, solid drawn round copper tubes for Medical gases or vacuum. Suitable sleeves shall be provided wherever pipes cross through walls / slabs. All pipe clamps shall be non-reactive to copper.
(g) All pipe joints should be made using inert gas fluxless brazing method. (h) All brazed joints in the piping shall be made up using brazing filler metals that bond with the base metals being brazed and that comply with Specification for Brazing Filler Metal, ANSI/AWS A5.8. Copper-to-copper joints shall be made using a copper-phosphorus brazing filler metal (BCuP series) without flux. Dissimilar metals such as copper and brass shall be joined using an appropriate flux with either a copper-phosphorus (BCuP series) or a silver (BAg series) brazing filler metal. Apply flux sparingly to the clean tube only and in a manner to avoid leaving any excess inside of completed joints. All joints should be of copper to copper and should be brazed brazing filler material without flux while being brazed joints shall be continuously purged with oil free dry nitrogen to prevent the formation of copper oxide on the inside surface of the joint. (j) All pipes should be installed without springing or forcing. (k) All pipes should be protected against mechanical injury in a manner satisfactory to authorities having jurisdiction. (l) Painting: All existing and proposed exposed pipes/should be painted with two coats of Synthetic enamel paint & color codification as per IS : 2379 of 1963. and approved medical standards. (m) All pipe clamp shall be non-reactive to copper. After erection the pipe should be flushed and then pressure tested with dry air at a pressure equal to 1.5 times of the working Pressure or 150 psig, whichever is higher for a period of not less than 24 hours in presence of medical authority/ MES authorities.
(n) All copper pipes should be inspected and certified for medical use before dispatch and the pipe should be delivered plugged or capped at both ends. Notes: (i) The tenderer shall design & submit complete layout drawings with proper colour code, fittings, fixtures complete for approval of GE or rep of accepting Officer before placing order and the quoted rate includes for the same. (ii) The quoted rate inclusive of all mechanical protections with metallic pipes to pipe required. (ii) Contractor is required to visit the site prior to bidding Make: Mex flow/Janya/ Metal Gem / Cubex/Metalco
(a) 42mm OD Copper pipe with fittings 1 mm thick
(b) 28mm OD Copper pipe with fittings 1 mm thick
(c) 22mm OD Copper pipe with fittings 1 mm thick
(d) 15mm OD Copper pipe with fittings 1 mm thick
Material and labour for Area gas alarm panel two lines, with audio Visual alarms for each gas used on the floor and come fitted with test and mute button Running on electricity 220 V, these units are suggested to be fitted for heavy usage.it would be fitted with a pressure gauge.It indicates low,medium & high pressure sensors all as directed by Engineer-in-Charge. The gas alarm system should have :- (a) Two gas alarm. (b) Individual gas status LED indicator-showing NORMAL, HIGH and LOW conditions for each service. (c) Audio alarm for high and low pressure condition. (d) Display of line Pressure for all the services with factory calibrated pressure sensors. (e) Alarm test / check / mute facility needed. (f) Small and compact in design. (g) Mounted on a powder coated MS box. (h) Nut and Nipples are to be provided for connection with pneumatic supply line. (j) Low voltage internal operation with input power supply of 220 V AC. (k) Battery Backup. (l) Easy wall mounting facility. (m) The model should be IEC 60601-1-2 compliance. (n) It should be compliant to ISO 7396-1/NFPA 99/HTM standards. (o) The product should be BIS / CE Certified. MAKE : BSI/LMPL/MR ENGG
Supply, Installation, testing and commissioning of 15mm OD Medical Line Valve or Isolating Valve as per CE Ccertified /HTM -2022/02-01/DIN/ISO/ BS EN 1057 or ball valves as per BS 1057, rated at 400 psig, as well as full vacuum ( 29.9" Hg). Valves go from full ON to full OFF by turning the vinyl-gripped valve handle 90 degree Locking-type handles may be ordered separately when required all as directed by Engineer-in-Charge. MAKE- CONEX/ RB/ AKTIV
All as per srl item no 56.00 here-in-before but 22 mm OD
All as per srl item no 56 here-in-before but 28 mm OD
All as per srl item no 56 here-in-before but 42 mm OD
Material and labour for H.P. tube for Oxygen, air 4 bar, rubber beaded complete all as directed by Engineer-in-Charge.
Material and labour for cable tray made up of MS angle of 50x50x6mm each and 150mm long welded/fixed with nut bolts to any steel frame or embedded in brick wall/concrete using PCC (1:2:4) typr B1 using 20mm graded aggregates complete all all as specified and directed.
Notes for Item No.25 to 61: The quoted rate shall be inclusive for the following:- (a) The Contractor shall provide comprehensive on-site warranty (Including All Spares, Accessories and Labour) for a period of 3 years from the date of final acceptance of the complete system after successful and complete installation and commissioning with regular updation of newer technology as and when evolved. (b) The Contractor shall, depute trained manpower to maintain manifold and medical gas pipeline system in case of any breakdowns within 24 Hrs. The technical maintenance personnel of the Contractor shall be responsible for supervision and maintenance shall be available to reach the site(s) within 1 hour's notice. (c) If the performance of any individual equipment or system is not satisfactory, the same shall be replaced by the contractor free of cost.
(d) If it is found that to meet the performance criteria, any extra equipment is required the same will be provided free of cost by the contractor. (e) All faults appearing and their rectification shall be periodically advised to the hospital, the period being not more than a month. (f) Any lacuna or lacunae noticed in the functioning of the installation as a result of any design feature shall be rectified by thecontractor free of cost. (g) The Contractor shall fully associate the engineers and technicians for MH Ambala during installation, testing, commissioning, operation and maintenance period. The testing of the entire system shall be carried out in the presence of GE / AGE and Officer rep (Min 02 Nos.) from Hospital authorities. (h) The Contractor shall train the staff of MH Ambala the operation of complete medical gas and oxygen plant system fopr a period of two weeks. (j) The Contractor shall submit MOU with authorised Medical Service engineer/ registered Bio medical engineer services experienced in the relevant field for not less than 5 years immediately before placing work order. (k) All electrical wiring/cables,cable tray, earthing, Control panels, connected electrical & pipeline distribution system with all connected accessories of complete medical gas system and oxygen plant system is deemed to be included in the quoted rates.
(l) The MGPS system installation shall be deemed to have been completed only on satisfactory performance of the MGPS system and acceptance of the satisfactory test results by the representative detailed by the Accepting Officer. In case of any failure during the testing the contractor shall be responsible to carryout necessary rectification/replacement of defective items/system parts thereof as directed by the Garrison Engineer immediately next day at no extra cost to Government. The decision of the GE shall be final and binding in this regard. The installation shall be re-tested after rectification/ replacement of the defective components, the installation/ equipments shall be finally taken over after the contractor has given satisfactory test as certified by the GE.
(m) The entire installation (after satisfactory completion of all tests and taking over) shall be deemed to be guaranteed by the contractor for efficient performance for thirty six months from the certified date of completion of work. In case of any fault arising during this period the contractor shall do the rectification/ replacement of defective parts, as the case may be, on his own expense. The defect liability period referred to in condition 46 of IAFW- 2249 (General condition of contracts) for the purpose of this contract, shall be deemed to be the period covered by the guarantee given by the contractor.
SCHEDULE A PART III (E/M WORKS) Material and labour for concealed point wiring using approved colour coding as per IS, with multi stranded copper conductor 1.5 sqmm (Nominal area), single core, FRLS-H, 1100 volts, ISI marked, PVC insulated and unsheathed stranded cable drawn through and including non metallic, medium grade PVC concealed conduit not less than 25.00mm to 32.00mm dia and its fittings, including cutting chasing on wall/ceiling, including all tees, bell mouth pieces and fixing accessories such as couplers, saddlers, pipe hooks etc, PVC junction boxes, ISI marked including mild steel sunken boxes, made out of 1.6mm thick mild steel sheet, with earth dolly, coated with two coats of anti corrosive paint over a coat of red oxide primer internally, including 3mm thick plastic laminated sheet top cover extended upto 25mm over sunken box for housing the switch, sockets, fan regulators etc. with brass screws and brass washers, including 1.5 sqmm single core PVC insulated and unsheathed FRLS-H cable, 1100 volt grade with multi stranded copper conductor green colour as earth continuity conductor upto common earth and connecting to common earth dolly with suitable size of lugs complete all as specified and directed for one light point (For flourescent Tube/ LH, Bldg Security light, Mirror light)/fan point/bell point controlled by one one way switch complete all as specified and as directed. Notes :- (i) Switch, socket, ceiling rose, lamp holder, bell, bell push and fan regulators to be measured and paid for separately. (ii) PVC Conduit conforming to IS 9537 part I & III. (iii) All chases will be cut by chase cutter only and surfaces shall be make good as per existing surfaces.
Description all as per Srl item No.66.00 here in before but for one 3 pin socket outlet point, 5 Amp, on same board.
Description all as per Srl item No.66.00 here in before but for one 3 pin socket outlet point, 5 Amp, on Independent board.
Description all as per serial item No. 66.00 above, but for one socket point, 3 pin,15 Amps, on independent board, with 2.5 Sqmm PVC, FRLS-H, unsheathed, multistranded cable, 1100 volts grade, ISI marked, including 2.5 Sq mm size, green colour, PVC insulated FRLS-H, continuous copper earth wire and connected to common earth dolly complete all as specified and as directed.
Description all as per serial item No. 66.00 above, but for one socket point, 3 pin,15 Amps, on independent board, with 4.0 Sqmm PVC, FRLS-H, unsheathed, multistranded cable, 1100 volts grade, ISI marked, including 4.0 Sq mm size, green colour, PVC insulated FRLS-H, continuous copper earth wire and connected to common earth dolly complete all as specified and as directed.
Material and labour for sub main wiring with two single core, 6 sqmm, FRLS-H, 1100 volts grade, ISI marked, with multistranded, PVC insulated unsheathed copper conductor cable, drawn in and including rigid non metallic PVC conduit, medium grade, not less than 25 mm dia, ISI marked for housing cables, concealed in wall or ceiling including necessary cutting, chasing with machine and making good disturbed surfaces as per existing, with accessories with and including one additional 6 sqmm, PVC insulated, stranded copper conductor, FRLS-H, green colour for earth continuity conductor connected to common earth dolly complete all as specified and as directed. Note- Three runs of cable and one run of concealed conduct shall be measured as one unit.
Supply and fixing of switch piano, polycarbonate, 5 Amps, one way, flush type, single pole, 230 volt, ISI marked complete all as specified and directed.
Supply and fixing of switch piano, polycarbonate, 15 Amps, one way, single pole, flush type, 230 volts, ISI marked complete all as specified and directed.
Supply and fixing of socket outlet, polycarbonate, 2 in 1, 3 pin, 5 amps, flush type, 230 volts complete all as specified and directed.
Supply and fixing of socket outlet, 2 in 1, 6 pin, 6/16 Amps, multipin, flush type,complete all as specified and directed.
Supply and fixing of ceiling rose, surface bakelite, 65x50mm, three terminals along with back plates complete all as specified and as directed.
Supply and fixing of angle lamp holder, PVC/Polycarbonate type, with back plate suitable for batten fittings complete all as specified and as directed.
Supply and fixing of indoor decorative surface mounting LED tube light fitting, batten type, with poal diffuser, high efficient aluminium housing body in CRCA body suitable for LED tube, high efficient, AC 220-240 volts, 50 Hz with and including 18/20 watt LED tube rod luminaries (warm/cool as per site requirement), environment friendly, energy saving, high efficient PMMA diffuser, T8 with aluminium extrusion, heat sink and connecting with copper conductor flexible cable of size 0.50 Sqmm from ceiling rose to connector of the light fitting complete all as specified and as directed.
Supply and fixing of exhaust fan, heavy duty, 380mm sweep, minimum 900 RPM, AC, single phase, 230 volts, metal body with louvers complete and connecting up from ceiling rose to fan by three core copper wire of suitable capacity, fixing exhaust fan and louvers with suitable size MS bolts and nuts in drilled holes of existing brick/PCC wall, suitable for mounting exhaust fan over and including supply and fixing of ply wood sheet/plastic sheet on brick/PCC wall surface for mounting of exhuast fan using fixing arrangement with anchor fastner/steel frame sheet etc and testing complete including making good the disturbed surfaces to match with existing by PCC/cement mortar work and finishing with white washing/colouring/distempering etc complete all as specified and directed.
Supply, installation, commissioning and testing of distibution board for MCBs, SPN, 4 way, made of CRCA sheet, dust and vermin proof, double door, IP 43, as per IS 8623, 240 volts, with and including 200 amps fully insulated copper bus bar fixed on wall duly grouted in solid block wall with and including copper lugs/thimbles duly crimped for connection complete all as specified and directed.
Supply and fixing of MCB SP, 6 to 32 Amps, 240 volt, 10 KA, IP-20, transperent label holder, fully insulated safety shutters, suitable for cable upto 35 Sqmm cross sectional area, ISI marked, connected with suitable lugs and testing complete all as specified and directed.
Supply and fixing of MCB SPN, 40 Amps, 240 volt, 50 Hz, breaking capacity 10 KA, conforming to IS-13947-3 and IEC-60947-3:2001, double pole complete all as specified and directed.
Supply and fixing of sheet metal enclosures vertical load line TPN distribution board, 4 ways, double door powder coated alongwith 250 Amp rated busbar with and including MCCB 3 pole 160 Amp, adjustable type as incomer with suitable size of MS cabinet fabricated by 2mm thick MS sheet, 415 Volts, 10 KA and making all type of connections using proper size of gland for incomer cable complete loadline TPN concealed on wall surface & making good the disturbed surfaces as per existing complete all as specified and as directed.
Supply and fixing of MCCB, 100 Amps, 4 pole, 16 KA, 415 volt, connected with suitable lugs and testing complete all as specified and directed.
Material and labour for earthing with galvanised earth plate electrode 600mm x 600mmx 6mm thick surrounded by charcoal and salt, buried directly in ground vertically in a pit of depth not less than 2.25m below ground level with top edge of the earth plate not less than 1.5m below normal ground level, connected to 4.00 mm GI wire earth lead by means of bolts, nuts, checks nut and washers made out of galvanized iron or steel, earth wire and thimbles protected by galvanized steel pipe 15mm dia (light grade) and connected to main switch or main control board all as shown in Electrical Plate No.3 of SSR Part-I & including necessary excavation and earth work in any type of strata, PCC (1:3:6) type C1 in pit, RCC cover slab 50mm thick in 1:2:4 type B-1 using 8mm dia TMT bars @10cm c/c both way with one 8mm dia bar as handle for lifting, 20mm dia GI medium grade watering pipe, funnel, wire wesh, charcoal or coke & salt in alternate layers, etc.all as indicated including testing on completion complete all as specified and directed.
Supply and fixing in replacement of LED security light luminaire 45 watt, 240 volts, AC, 50 Hz, made up of die cast aluminum housing with sealed toughened glass cover, top opening separate driver compartment with frame, safety input voltage range 160 to 270 Volts, IP66 protection, suitable for wall mounting, SMD led array with lense, cool white 6500K, LED efficacy>120 lm/W and system efficacy 90 lm/W, CRI>70%, power factor> 0.95, working temperature range -20 degree centigrade to 60 degree centigrade, built in surge protection of more than 5 kv with facility of power supply including service brackets of GI pipe 40mm bore, light grade, of length 0.90 metre, bent to required shape, fixed with MS clamps of size 40 x 6mm flat iron strip, 150mm long and 4 Nos.8mm dia stainless steel bolts and nuts and washers and suitable reducer for connecting to GI pipe as required at site and making all required electrical connections using 3 core copper flexible cable including cutting,drilling holes, welding etc complete all as specified and as directed.
Material and labour for fine and dry sand, cushioning for underground cable, by providing a layers of sand 8cm deep before laying and 15cm deep after laying the cable including stretching the cable and punning sand cushioning complete all as specified and as directed. Note - Punned thickness of sand layer shall only be measured and paid to contractor.
Supply and laying un-reinforced, precast concrete cable cover, class HV type 1, Flat, with peak, size 300 mm x180 mm x40 mm complete all as specified and as directed.
Supply and laying burnt clay bricks, sub class 'B', old size bricks, designation 75, laid in trenches, in transverse direction to cable, as protection to cables complete all as specified and directed.
Supplying installing testing and commissioning diesel engine driven generating set 415 V, 3 phase, 4 wire as per BS 5514/IS-10002 complete with standard control panel, of capacity continuous rating 100 KVA/80 KW at NTP conditions with acoustic enclosure and canopy cover considering all derating factors and comprising of the following :- (a) ALTERNATOR : Self regulated with brushless self excitation 100 KVA 415 V, 3 phase, 4 wire, 50 HZ, 0.8 P.F.,1500 RPM, with voltage variation plus or minus 2.5% of rated voltage from no-load to full-load foot mounted, screen protected drip proof type with class 'F' insulation, closed coupled single bearing alternator, with necessary controls as recommended by the manufacturers, conforming to BS-5000/IS-4772, capable of supplying unbalanced load. (b) Engine : Multi/cylinder,compression ignition, heavy duty,compact diesel engine liquid cooled with standard accessories, all as per BS-5514/ISO-3046,1500 RPM,control gauges, RPM/Hours meters etc; shaft H.P. suitable for generating application and capable of giving an output of 100 KVA/80 KW for coupled alternator under NTP conditions, complete with following:- (i) Flywheel housing with flywheel to suit closed coupled single bearing alternator. (ii) Air-cleaner dry type with vacuum indicator. (iii) Fuel pump and fuel piping 15mm M.S. pipe with bends, flanges etc. (iv) Fuel filter with water separator and filter element. (v) Lub-oil filter with filter element (vi) D.C. starter 12 volts. (vii) Storage Batteries of suitable voltage and ampere-hours as recommended by the manufacturers. (viii) Residential silencer as per CPCB norms (x) Supporting/fixing arrangement for muffler with celiling through antivibration pads. (xi) Engine overspeed stop/trip switch. (xii) Radiator with fan. (xiii) Dynamo for battery charging. (xiv) Corrosion inhibitor coolant. (xv) Engine control panel electronic digital display type consisting of battery voltage indicator, lub oil temp and pressure gauges coolant water temp gauge, RPM indicator cum hour meter etc. (xvi) Safety controls for alarm due to high water temp (HWT) and low lub oil pressure (LLOP). (xvii) Any other devices/accessories recommended/supplied by the manufacturers as standard eqpt. Engine.
(xiv) 3 Phase electronic energy meter for measuring meter for measuring generated units. (xv) Frequency meter (Digital type) of suitable rating. (xvi) MCCB 200 Amps 4 Pole 25 kA & MCCB 200 Amp 4 Pole 25 kA breaking capacity with fixed thermal and magnetic release, for main supply & DG Supply adjustable over load and adjustable short circuit each 01 No (xvii) The AMF panel shall have manually by pass arrangement in case of failure of AMF Panel. (xviii) Other accessories, if any , as per manufacturer instructions. (e) Diesel tank of capacity for 8 hours continuous run on full load, with caliberated scale (fuel level indicator) dip stick, fuel inlet/outlet , Air vent, drain plug and inlet arrangement for direct filling, including suitable size gate valve in outlet, tank to be fitted at suitable height. (f) Sound proof enclosure with residential silencer (CPCB) approved) the sound less than 75 db at a distance of 1 meter from canopy in free field zone. (g) PCC/RCC foundation complete as per manufacturers own design/ Or as per directions of Engineer-in- charge. (h) Documents : Three sets of following documents :- Maintenance and operation manuals for engine and alternator, spare parts catalogue, warranty/test certificates, schematic diagrams of control panels. (j) other items : List of tools and two years normal maintenance spares as per manufacturers recommended lists, duly priced, item wise. Notes : (i) Contractor will satisfactorily conduct load test of DG set for 12 hours including run on 10% overload for 1 hour. (ii) Testing equipment will be arranged by the contractor. (iii) Fuel/oil required for load test shall be arranged by the contractor and cost thereof is deemed to be included in the lump sum amount quoted by the contractor.(iv) AMF panel and DG set shall be procured from same OEM i.e make of both shall be same.
Supply, install, test and commissioning of step down transformer, out door type conforming as per IS 1180 (Part 1):2014 energy efficiancy level-2) complete with standard fittings and accessories, HT 11 KV, 3 Phase, LT 433 Volt, 3 Phase, 4 wire with off load tap changer on primary side to carry secondary voltage + 5% to -10% with tappings at every 2.5 % all as per IS 2026 of capacity 100 KVA, copper wound, oil cooling with fins vector group : DYn 11, including initial oil filling complete all as specified and directed including the following accessories :- (a) Three Nos porcelain bushing HT 11 KV with brass bushing rods, nuts, bolts & washers etc inside cable end box. (b) One cable box LV side consisting of 4 Nos porcelain with brass bushing rods complete with nuts and washers suitable for runs of 3.5 core LT UG cable. (c) Oil conservator with filler type drain valve and oil level gauge. (d) Dehydrating breather with first fill of dehydrating agent. (e) Explosion vents. (f) Thermometer pocket on tank cover. (g) One No 100mm dia, dial type thermometer vapour pressure type having range 0 to 120 degree centigrade. (h) Rating plate and terminal marking plate.(j) Earthing arrangement.(k) Oil level gauge. (l) Oil drain filter valve.(m) Oil drain valve.(n) Lifting hook. (o) Oil conservator with filling hole and cap. (p) Air relief valve plug.(q) Four plate solid cast steel detachable rollers. (r) Earth bushing for neutral outside the cable box for providing independent earthing for neutral. (s) Necessary channel on foundation for fixing the transformer with joint and stopper. Note:-Necessary civil work like as foudation/ platform for the equipment of appropriate size as reccomended by OEM or as indicated/directed by the Engr-In-Charge for the connected work is required deemed to be included in the unit quoted rate.
Supply, install, testing and commissioning of Servo control voltage stabiliser, of capacity 100 KVA (conforming as per BIS : 9815: Part 1 : 1994 reaffirmed 2009), 3-phase, 50 Hz, naturally Oil Cooled, indoor/outdoor, Unbalanced Type, copper wounded, Digital Display Type, Relay Free smooth and silent, Operation with 1st filling of Transformer oil (IS 335:1993 reaffirmed 2005) having following specifications complete all as specified and directed:- (a) Input voltage 300V-480V or as per ground requirements. (b) Output voltage stablized at any designed/desired voltage 415v +/- 1% (Adjustable) (c) Rated frequency 50 hz (d) No distortion in wave form. (e) Response time not to exceed 10 mili seconds (f) Sensitivity /-) 1% (g) Rate of correction 8 volt/second for single phase and 20 volt/second for phase to phase (h) No load losses less than 0.5% (j) Full load losses less than 2.5% (k) Duty cycle - continuous 24 x 7 (l) Efficiency more than 95% (m) Type of load, 3 phase, unbalanced type (n) Digital display voltmeter/Ammeter (o) Capacity 100 KVA (p) Overload/Short circuit protection (q) High/Low voltage cut off (r) By pass arrangement in case of emergency (s) Single phase sequence change protection.
Supply, Install, Testing & Commissioning of 11KV, 400 Amps, both side extensible outdoor [IP-54] protection, powder coated painted, light smoke gray colour, manually operated, SF-6 gas insulated RMU (Ring Main Unit), 3 way comprising of 3 Nos, 400 Amps vaccum circuit breaker of rupturing capacity 21 KA/3sec, inside SF-6 gas filled stainless steel (non ferride and non magnetic grade) tank having minimum thickness 2mm having IP-67 degree of protection and leak rate of SF-6 gas less than 0.1%. RMU should be provided with copper Busbar, Scada compitable relays with each incomming / outgoing, mechanism for interlocking arrangements for safety and operations, pad lock provisions, self powered & complete with all accessories RMU shall be type tested for internal arc test of 20 KA /3 Sec as per IEC : 62271-200. The unit shall include the following with all necessary standards fitting as required at site complete all as specified and directed. Note:-Necessary civil work like as foudation/ platform for the equipment of appropriate size as reccomended by OEM or as indicated/directed by the Engr-In-Charge for the connected work is required deemed to be included in the unit quoted rate. Make :-ABB/ Siemens/L&T /C&S Electric/Schneider
Material and labour for earthing complete with copper plate, with 60cm x 60cm x 6mm thick electrode, surrounded by charcoal and salt, buried in ground vertically not less than 2.25mtrs deep below ground level with top edge of the plate not less than 1.5mtrs below ground level connected to copper strip of size 20x3mm as earthing lead with bolts, nuts, checknuts and washers of galvanised iron or steel, earth strip protected by galvanized iron pipe 32mm dia (medium grade) from earth pit to test point and connected to main switch or main control board all as shown in Electrical plate No. 3 connected to earthing test point all as specified including testing on completion including necessary excavation and earth work (in any strata) providing charcoal dust, salt mixed with fine sand in continuous layers, with CI frame with cover, PCC (1:3:6) type C1 in pit, RCC cover slab 50mm thick in 1:2:4 type B-1 using 8mm dia TMT bars @10cm c/c both way with one 8mm dia bar as handle for lifting, 20mm dia bore light grade GI watering pipe, PGI sheet funnel etc complete all as specified and directed.
Material and labour for earthing complete with 60cm x 60cm x 6mm thick GI plate electrode surrounded by charcoal and salt, buried in ground vertically not less than 2.25mtrs deep below ground level with top edge of the plate not less than 1.5mtrs below ground level connected to GI strip of size 32x6mm as earthing lead with bolts, nuts, checknuts and washers of galvanised iron or steel, earth strip protected by galvanized iron pipe 40mm dia (medium grade) from earth pit to test point and connected to main switch or main control board all as shown in Electrical plate No. 3 connected to earthing test point all as specified including testing on completion including necessary excavation and earth work (in any strata) providing charcoal dust, salt mixed with fine sand in continuous layers, with CI frame with cover, PCC (1:3:6) type C1 in pit, RCC cover slab 50mm thick in 1:2:4 type B-1 using 8mm dia TMT bars @10cm c/c both way with one 8mm dia bar as handle for lifting, 20mm dia bore light grade GI watering pipe, PGI sheet funnel etc complete all as specified and directed.
Material and labour for boring of holes under road, pavement, hard standing by trench less technology using ultra sound technique and necessary machine including providing HDPE pipe of 80mm dia suitable for 0.4 mpa working pressure complete all asspecified and directed at site. Note:- (i) The boring shall be carried out 3 metre deep below the railway track level and NH 1A line. (ii) During carrying out the work below railway track and NH 1A, the contractor shall adhere to the instructions issued and to be followed by the respective authorities. The work will be carried out under the supervision of MES and Railway/Highway authorities. (iii) Any damage to the existing property like sewage pipes, cables, water pipes etc of all the authorities will be repaired/replaced to the entire satisfaction of respective authorities without any extra cost. Department in no way will be responsible for such damages. (iv) The quoted rate shall include all the above and excavation, working space for trenchless boring, HDPE pipe, machinary, transportation, DG Set, labour charges etc. complete and nothing extra is admissible on this account.
Supply, install, testing and commissioning of air breaker gang operater, triple switch ( GODO) outdoor type, vertical mounting type of capacity 200 Amp with drop out fuse consisting 12 insulators ISI marked mechanically operated mounted on insulator and frame work with operating GI rods and mechanical work from ground level with locking arrangement including Godrej pad lock with two keys complete with fuse carried and HT fuse element having upto 200 Amp rating suitable for HT 11 KV 50 HZ AC Supply including necessary electric connection complete all as specified and directed .
Material and labour for aluminium conductor steel reinforced (ACSR) including fixing, stretching on poles and proper noseling with binding wire with insulator, of any size as required at site complete all as specified by Engr-in-charge.
Material and labour for lightning arrester, expulsion type, suitable for 11000 volts, with all fittings, trasmission class, discharge capacity 65 KA including connecting with HT line complete all as specified and directed.
Supply and laying of LT underground cable, XLPE insulated, screened, PVC bedded, galvanised steel strip or wire armoured, electric power cable (heavy duty), 1100 volts grade, with stranded aluminium conductor of size 3.5 core, 95 Sqmm and laying in trenches, floors, ducts, passing through pipe for road crossing etc, including connecting and testing of cable complete all as specified and as directed. Note:- Excavation, earth work, brick protection, sand cushioning and GI pipe etc shall be measured & paid for separately.
Supply and laying of LT underground cable, XLPE insulated, screened, PVC bedded, galvanised steel strip or wire armoured, electric power cable (heavy duty), 1100 volts grade, with stranded aluminium conductor of size 3.5 core, 25 Sqmm and laying in trenches, floors, ducts, passing through pipe for road crossing etc, including connecting and testing of cable complete all as specified and as directed. Note:- Excavation, earth work, brick protection, sand cushioning and GI pipe etc shall be measured & paid for separately.
Supplying and laying XLPE insulated, screened, PVC bedded, galvanised steel strip or wire armoured, electric power cables (Heavy duty) with aluminium conductor, 11000 Volts grade, cross sectional area 95 Sqmm 3 core, suitable for 11 KV earthed system, in trenches, floors, ducts passing through pipe for road crossing etc, including connecting and testing of cable complete, all as specified and directed. Note : Excavation, earth work, brick protection, sand cushioning and GI pipe etc shall be measured & paid for separately.
Supply and fixing of GI tubing 80 mm bore light grade on poles, on walls, under floors, road crossing as required at site with and including clamps, nuts, bolts, saddles etc complete all as specified and directed.
Material and labour for Cable jointing kit suitable for LT 1100 volts (Earthed) grade cable outdoor end termination, heat shrinkable type joint complete with jointing material and accessories suitable for 3.5 core LT XLPE armoured aluminium conductor cable of size 120 to150 sq mm complete all as specified and as directed.
Material and labour for Cable jointing kit suitable for HT 11 KV (Earthed) grade cable, outdoor end termination, heat shrinkable type joint complete with jointing material and accessories suitable for 3 core HT XLPE armoured aluminium conductor cable of size 50 to 95 sq mm complete all as specified and as directed.
Material and labour for Cable jointing kit suitable for HT 11 KV (Earthed) grade cable, indoor end termination, heat shrinkable type joint complete with jointing material and accessories suitable for 3 core HT XLPE armoured aluminium conductor cable of size 50 to 95 sq mm complete all as specified and as directed.
Supply, install, testing & commissioning of factory fabricated, out door type cubical panel board, of size 1.2 metre x 1.0 metre x 0.5 metre, outer shell made out of 3.15 mm thick MS sheet including door and inner shell made out of 2 mm thick MS sheet and MS angle section 50x50x6 mm, mounted on PCC 1:2:4 type B-1 platform, 60 cm above and 45 cm below the ground level, frame factory fabricated with front & back openable, double door system and cable alloy (Each door hinged with body at 3 places with body and hinges should be welded with body and door) lockable/openable by Keys having chromium plated 19 mm dia fancy lock, “U” type gasket shall be used to make the outer shell completely vermin proof, suitable for 3 phase, 4 wire, 415 Volts, 50 Hz, tested at factory, approved from CPRI, having modern fabrication and testing process followed by powder coat painting having thickness not less than 100 microns, provided with suitable gasket to make it dust/vermin proof and general construction confirming to IS:8623/93 comprising of the following :-
Note :- (i) The quoted rate shall includes minor items such as lugs, nuts & bolts, washers, top insulators, sleeves,link spreater provision of extra holes in bus bar etc. (ii) All internal wiring shall be done with suitable size of PVC insulated solid copper conductor FRLS cable conforming to IS-694. (iii) Panel drawing shall be got approved before placing of order from Engr-in-Charge/GE. (iv) The bus connection for cable from main bus bar to outgoing MCCB will be connected through suitable rating bus bar strip type duly insulated. (vi) The Panel shall be surrounded with 9 inch brick wall upto 0.5 mtr height all around the panel and blank portion between panel and wall shall be filled with PCC 1:2:4 incl masonary work leaving ducts for cables complete all as directed.
Supply and fixing of sheet metal enclosures for distribution board for MCCB's, Vertical, double door, IP-43 type, 4 ways, MCCB +12 module suitable for and including MCCB, Four pole, 125 Amps as incomer and fixing only MCB's as out going, made up of 2mm thick MS sheet of suitable size and making all type of connections with suitable size of gland for incomer LT cable complete, concealed on the wall surface and making good the disturbed portion with adjoining surfaces to match with existing finishes and dismantling complete all as specified and as directed.
Supply and fixing of room window type air conditioner, electric driven (Single phase), cooling capacity 6000K Cal per hour (2 Ton), BEE 5 star rating with remote complete all as specified and directed.
Supply and fixing of voltage stabilizer for automatic operation, single phase, naturally air cooled, indoor type and designed for input variation between 170V to 260 V with output voltage stabilizer at 230 V with accuracy of + 2.5% and of capacity 4.0 KVA aluminium wound complete all as specified and directed.
Supply and fixing of single pole and neutral enclosure with a two pin and earth plug and socket complete with one single pole MCB SP 20 Amps, ISI marked complete all as specified and directed.
R D Bansal
Esha Engineers
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .