RequiredCertificate (Requested in ATC) Bid to RA enabled No Hematology Analyzer ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Set 1 Usage: To count
Requiredmeasure and analyse red blood cells
Requiredwhite blood cells and platelets
Requiredto measure the amount of hemoglobin in the blood and within each red blood cell. " * Haematology Analysis is based on the principle of counting and sizing: Yes * 1/6 "Analyser provides complete blood cell counting including: 3 part WBC differential 3 part WBC differential
Required5 part WBC differential without Reties and NRBC enumeration capability
Required5 part WBC differential with Reties and NRBC enumeration capability Or higher Analyses Leucocytes in their native state through: Spectrum photometry based analysis Laser based scatter analysis
RequiredSpectrum photometry based analysis Applicable analyses available: Specify the number of parameters analysed by the machine: 19 19
Required20
Required21
Required22
Required23
Required24
Required25
Required26
Required27
Required28
Required29
Required30 Or higher Upgradable to attach with fully automated slide- maker & slide stainer in future: No Yes
RequiredNo Capability to automatically enumerate
RequiredNucleated Red Blood Cells in the CBC Difference Mode and without additional reagents: No Yes
RequiredNo Facility for PC based data management with all scatter plots
Requiredhistograms and display and in print: Yes Yes Auto loader facility with mixer capable of loading min 90 samples at any time (with Open vial
RequiredPre dilute & Closed vial modes) STAT capability with Positive Bar code identification facility for samples loading/unloading available No Yes
RequiredNo Facility for User defined rules & flagging limits: Yes * "Database capability of storing sets of >=5000 to 10000 >=5000 to 10000
Required>=10000 to 20000
Required>= 2/6 results and graphics
Requiredplease specify sets of results and graphics storable in database: 20000 Or higher Facility for workload recording
Requiredavailable: Yes Yes Have unlimited number of user-definable control files: Yes * Auto stop function in event of unacceptable control data
Requiredavailable: Yes Yes Ability to transmit results to host computer: Yes Yes Throughput capacity of analyser in Primary mode (samples/hour): 33 33.0 - 200.0 Or higher Linearity of Platelet to be from 0 to 3000 * 1000 cells/micro litres: Yes * WBC linearity to be from 0 to 3000 * 1000 cells/micro litres: Yes * Testing mode Selection facility available for: CBC mode CBC mode
RequiredCBC plus Differential & Reticulocyte testing mode Or higher Analyses time for cytopenic samples (RBC
RequiredPlt and WBC): Yes * Directly measures MCV: Yes * Capability to differentiate between smaller RBCs and Larger Platelets: Yes Yes
RequiredNo Have auto purge function in the analyser’s software: No * Power Backup facility: "Through suitable UPS (using maintenance free batteries) ensuring at least 60 minutes power backup * Sufficient Quality Yes Yes 3/6 control files available (which store >= 100 runs each XB analysis): Printer Facility available through Inbuilt printer or external printer unit (to be supplied along with equipment): Yes Yes Offered equipment unit to be supplied with sufficient consumables (with at least 5/6th of total shelf life) required for
Requiredsufficient to carry out haematological testing of samples (please specify number of samples): 500 500
Required1000 Or higher Operating condition: Capable of operating continuously in ambient temperature of 10 to 50 deg C and relative humidity of 15 to 90% in ideal circumstances. * Storage condition: Capable of being stored continuously in ambient temperature of 0 to 50 deg C and relative humidity of 15 to 90%. * CERTIFICATION
RequiredSTANDARDS AND SAFETY Certification of product available US FDA
RequiredCE European
RequiredBIS US FDA
RequiredCE European
RequiredBIS Equipment shall meet internationally recognized for Electromagnetic Compatibility (EMC) for electro medical equipment: 61326-1 Yes * Certified to be compliant with IEC 61010-1
RequiredIEC 61010-2-281
Required61010-2- 101 for safety: Yes * Manufacturer have ISO certification for quality standards: Yes Yes 4/6 Warranty Period 3 3
Required5 Or higher * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.