RequiredCertificate (Requested in ATC) Bid to RA enabled No Semi Automatic Bio Chemistry Analyser ( 1 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Configuration & Physical Characteristics Usage: The equipment; Semi automatic bio chemistry analyser is used for measurements of bio-chemistry tests
Requiredusing wet chemistry reagent(s) and have ability to use external cuvettes and integrated flow cell. Semi automated Bio Chemistry Analyser is with built in software * 1/5 for the calculation and curve plotting . User Interface: Mechanical key board LCD display with Touch Key board
RequiredMechanical key board Flow cell type: Permanent flow cell Permanent flow cell Light Source: Halogen Lamp * Wavelength Selection & Measurement range (wavelengths): Automatic (wavelength) selection facility with 8 position filter wheel (with all filters in position)
Requiredcovers standard wavelengths in the range of 340 to 700 nm 6 fixed and 2 free positions Automatic (wavelength) selection facility with 8 position filter wheel (with all filters in position)
Requiredcovers standard wavelengths in the range of 340 to 700 nm 6 fixed and 2 free positions
RequiredAutomatic Wavelength selection facility with 6 position static photometer covers standard wavelength in the range of 340 to 630 nm
RequiredAutomatic Wavelength selection facility with 8 position static photometer covers standard wavelength in the range of 340 to 700 nm 6 fixed and 2 free positions Or higher Absorbance range: 0 to 2.5 unit 0 to 2.5 unit
Required0 to 3 unit Or higher Analyzer resolution: 0.0001 absorbance unit or better * Measurement range (temperature): 25
Required30
Required37 degree Celsius with 1 degree increment
RequiredPeltier * Memory for storing test results
Requiredsufficient for: >= 200 test results and 5000 patient results * Heat Dissipation ability: Analyser shall maintain nominal temperature
Requiredwith suitable heat dissipation/cooling mechanism. * Installation Type: Stationary installation in laboratory * Number of programmable Channels for >=100 to 200 >=100 to 200
Required>= 200 Or higher 2/5 chemistries: Assay/Calculation types available in analyser: End point with factor or standard
RequiredFixed time with factor or standard
RequiredKinetic; Enzyme kinetics with factor or standard
RequiredAbso rbance
RequiredSample blanking
RequiredDifferential mode with factor or standard End point with factor or standard
RequiredFixed time with factor or standard
RequiredKinetic; Enzyme kinetics with factor or standard
RequiredAbsorbance
RequiredSample blanking
RequiredDifferential mode with factor or standard Calibration Types available: Linear factor
RequiredMulti point
RequiredPoint to point
RequiredLog- Logit
RequiredSpline
Required1- point calibration
Required2-point calibration
RequiredPolygonal multi standard (Calibration Curve ) Linear factor
RequiredMulti point
RequiredPoint to point
RequiredLog-Logit
RequiredSpline
Required1- point calibration
Required2-point calibration
RequiredPolygonal multi standard (Calibration Curve ) 3 levels control facility with day to day Levy-Jennings curve stored and displayed
Requiredavailable: Yes Yes Sample incubator facility available: Yes Yes Flow Cell Flow cell material Quartz Quartz
RequiredQuartz + Metallic Flow Cell Volume(minimum): - * Flow Cell Optical Path: 10 mm * Aspiration Volume - * ENERGY SOURCE Power Source: Compatible with electric power from by 220V-240V 50Hz (+/-10%) AC source * Power Back up with provision of UPS (with maintenance fee batteries) for each equipment unit: At least 30 minutes power back up * CONSUMABLES Equipment is suitable for use of consumables under: Open System Open System Or higher ACCESSORIES Accessories to be supplied with each unit of Analyser equipment : Micro pipettes (5 No.) - 2 variable (5-50 )
Required(100-1000)
RequiredSmall Tips- 500 nos. & Large Tips- 500 nos. * OPERATING Operating & Storage Equipment is suitable * 3/5 CONDITIONS Conditions: for storage and capable of operating continuously in ambient room temperature of 8 degree C to 37 degree C and relative humidity of 15 to 90% in ideal circumstances. Parts of the Device that are designed to come into contact with the patient or the operator should either be capable of easy disinfection or be protected by a single use/disposable cover: Yes * PRINTER AND COMPUT ER/LABORATORY INTERFACE Printer facility: Yes
RequiredIn-built Thermal Printer * Computer interface: By RS232 port/USB/HDMI port facility * Lab Interface system (LIS) compatible viz analyser equipment preloaded with Software needed for lab interface: Yes Yes Quality Certifications & Warranty Certification available for the product: CE CE Equipment shall meet internationally recognised for Electromagnetic Compatibility (EMC) for electro medical equipment: 61326-1 and certified to be compliant with IEC 61010-1
RequiredIEC 61010-2-281 Yes Yes Equipment Manufacturer have ISO 13485 certification for quality standards: - Yes Warranty Period: 3 year 3 year
Required5 year Or higher * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer. 4/5