RequiredExperience with Government Bid to RA enabled No Veterinary Hematology Analyser ( 3 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Certifications for the product CE * Certification number and date - * Confirmity to quality management standards ISO 9001 & ISO 13485 * Confirmity to electrical safety standards IEC 61010-1
RequiredIEC 61010-2-081
RequiredIEC 61010-2-101 or BIS equivalent * 1/8 Performance Parameters Clinical Purpose Veterinary Hematology Analyzers are used to count and characterize blood cells for disease detection and monitoring. Basic analyzers return a complete blood count(CBC) with a three-part differential white blood cell(WBC) or five part differential white blood cell(WBC) count.Sophisticated analyzers measure cell morphology and can detect small cell populations to diagnose rare blood conditions * Type of Configuration Benchtop * Type of sampling system Closed system * Type of operation Fully automatic * Compatible Species Cat
Requireddog
RequiredRabbit
Requiredpig
Requiredgoat
Requiredhorse
Requiredmonkey
Required+4 user defined * Automatic Start Up
Requiredshut down and sample analysis Yes * Analysis principle Electrical impedance
RequiredCalorimetry
RequiredLaser scattering method Electrical impedance
RequiredCalorimetry
RequiredLaser scattering method Type of cell counting 3 part WBC differential 3 part WBC differential Analysis parameter available WBC
RequiredLYM#
RequiredMID#
RequiredGRA#
RequiredLYM%
RequiredMID%
RequiredGRA%
RequiredRBC
RequiredHGB
RequiredMCHC
RequiredMCH
RequiredMCV
RequiredRDW- CV
RequiredRDW-SD
RequiredHCT
RequiredPCT
RequiredMPV
RequiredPDW
RequiredPCT
RequiredP-LCR * RBC histogram Yes Yes Or higher PLT histogram Yes Yes Or higher P-LCC histogram Yes Yes Or higher WBC histogram Yes Yes Or higher Minimum sample volume required µL 9 * Maximum sample volume needed in any of modes 20 * 2/8 Throughput (samples per hour) 60 * Counting mode of the analyzer Whole blood
Requiredperipheral and prediluted * Time taken by the analyser to produce the test results(Analysis time) in seconds 60 * Capability to automatically enumerate
RequiredNucleated Red Blood Cells in the CBC/Difference Mode and without additional reagents Yes * Auto loader facility with mixer capable of loading min 90 samples at any time (with Open vial
RequiredPre dilute & Closed vial modes) STAT capability with Positive Bar code identification facility for samples loading/unloading available Yes Yes Or higher Facility for User defined rules & flagging limits Yes * Facility for workload recording Yes * Have unlimited number of user-definable control files Yes * Quality assurance system with calibration and controls Yes * Quality control X-R
RequiredL-J * Type of Calibration Automatic * Automatic probe wipe Yes * Directly measures MCV Yes * Floating discriminator for WBC/PLT Yes * Linearity of Platelet in cells/microlitres 0 to 2000 * 10³ cells micro litres * Linearity of WBC in cells/microlitres 0 to 3000 * 10³ cells micro litres * 3/8 Linearity of hemoglobin in gram per litre 0 to 25 gm per litre * Precision of WBC in % 3.00% 3.00% Precision of RBC in % 1.50% * Precision in HGB in % 1.50% 1.50% Precision of PLT in % 3.00% 3.00% Carryover of RBC
RequiredWBC
RequiredHGB
RequiredPLT in % ≤1 * Capability to differentiate between smaller RBCs and larger platelets Yes * Auto purge function in the analysers software Yes * Availability of Auto dilution Yes * Barcode reader facility for sample Yes * "Auto stop function in event of unacceptable control data" Yes * Analyses time for cytopenic samples Yes * Separate diluting nozzles for RBC and WBC Yes * Optimized reagents and controls Yes * Automatic electric clog removal Yes * DATA MANAGEMENT AND DISPLAY Type of data management Inbuilt system * Inbuilt monitor type LCD * Inbuilt monitor size in inches 8 * Data management systems Provide scatterplots
Requiredhistograms in display and print * Type of user Interface or data entry Touchscreen Touchscreen No of USB Port 1 * "Database capability of storing sets of results 50000 * 4/8 and graphics" Have auto cleaning function in the analyser’s software Yes * Printer facility Yes * Type of printer unit Inbuilt and external * Power Requirements Type of power supply in VAC
RequiredHz 220-240V
Required50/60Hz single phase * Power Backup facility Yes * Rating of UPS in KVA 0.5 KVA 0.5 KVA Back up time in minutes 15 min 15 min Accessories
RequiredSpare Parts and Consumables Should provide the rate and packing of all reagents for one year Yes * One set of all reagents and controls at the time of delivery Yes * Reagent expiry time should be minimum of 1 year Yes * Alerts for operator for level of reagents and to empty waste when indicated Yes * Staged supply of controls over 1 year from the date of installation Yes Yes Or higher Cost of reagents
Requiredcontrol & calibration for price to be declared involve cost/cycle for 3 years Yes * Availability of micro capillary adaptor Yes * 3 level QC should be demonstrated to the technical committee during the technical demonstration of the offered equipment Yes * Operating temperature and humidity Capable of operating continuously in ambient temperature of 15 to 35 deg C and relative humidity of 15 to 85% in ideal circumstances * 5/8 Miscellaneous Parameters Number of installations in Central/State/PSU Govt Hospitals ( Hint: Seller should supply a performace certificate of the device to the buyer incase demaneded after placement of order) 0 0 OEM/Reseller ( if supplied by reseller shall ensure uninterputed availabilty of all spares for 10 years) Yes * Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes * During the warranty period it shall be ensured that all the break down calls are attended with in 24 hrs and the complete details of service agents/contact details should be furnished to buyer and consignee at time of supplies Yes * User/Technical/Mainten ance manuals to be supplied in English in hard and soft copy Yes * Details of equipments and procedures required for local calibration and routine maintenance to be supplied and advanced maintenance task documentation also to be furnished Yes * List of important spares and accessories
Requiredwith their part numbers to be supplied to the buyer at the time of supplying the equipment Yes * 6/8 Installation and demonstration of equipment and training to be provided after completing supplies before acceptance Yes * The Principal Manufacturer must have direct Presence/approved service center In India Yes * Copies of reports and certifications to be furnished to buyer on demand at time of supplies Yes * Calibration certificates as per NABH requirement Yes * Warranty in years 3 3 * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer. Additional Specification Documents