Loading…
Loading…
| # | Company | Offered Item | Amount | Rank | Status |
|---|---|---|---|---|---|
| 1 | L1₹1.0 LQualified Make: Meditec
Model: Make: Meditec International England Ltd. & Model: E-402 203 SWASTIK BHAWAN COMMERCIAL COMPLEX CENTRAL DELHI RANJEET NAGAR CENTRAL DELHI DELHI 110008 | WEST DELHI | DELHI | 110008 | Make: Meditec
Model: Make: Meditec International England Ltd. & Model: E-402 | ₹1.0 L | L1 | Qualified |
| 2 | L2₹1.1 L+₹3,669 (3.52%)Qualified Make: Contec
Model: Contec CMS100G A 38 1ST FLOOR NARAINA INDUSTRIAL AREA PHASE 1 NARAINA INDUSTRIAL AREA PHASE 1 SOUTH WEST DELHI DELHI 110028 | NEW DELHI | DELHI | 110028 | Make: Contec
Model: Contec CMS100G | ₹1.1 L+₹3,669 (3.52%) | L2 | Qualified |
| 3 | L4₹1.4 L+₹30,669 (29.4%)Qualified Make: LIFE PLUS
Model: LIFEPLUS , 1203 18 18 GANESH CHANDRA AVENUE KOLKATA CALCUTTA CENTRAL DIVISION WEST BENGAL 700013 | KOLKATA | WEST BENGAL | 700013 | Make: LIFE PLUS
Model: LIFEPLUS , 1203 | ₹1.4 L+₹30,669 (29.4%) | L4 | Qualified |
| 4 | L5₹1.4 L+₹36,069 (34.6%)Qualified Make: BPL
Model: BPL CARDIART 9108 PHASE1 PANCHKULA HARYANA 134113 INDIA | PANCHKULA | HARYANA | 134113 | Make: BPL
Model: BPL CARDIART 9108 | ₹1.4 L+₹36,069 (34.6%) | L5 | Qualified |
| 5 | L6₹1.4 L+₹39,669 (38.0%)Qualified Make: NA
Model: RMS 255 A VIMAN NAGAR KANPUR KANPUR NAGAR UTTAR PRADESH 208007 | KANPUR NAGAR | UTTAR PRADESH | 208007 | Make: NA
Model: RMS | ₹1.4 L+₹39,669 (38.0%) | L6 | Qualified |
Tender Value
Refer Docs
EMD Value
Exempted
Closing Date
7 Jun 2018, 3:00 pmClosed
electrocardiography ekg/ecg units Electrocardiography Ekg/ecg Units ( 3 ) Technical Specifications Specification Specification Name Values Allowed Values FEATURES Mobility
portability Protective splashproof case for clean storage and safe transport - Heart rate measurement range to be at least 30 to 250 bpm
with accuracy better than ± 5 bpm yes - Heart rate trend display of at least previous 24 hours yes - In built Software and/or standard of communication yes - Heat Dissipiation: Should maitain nominal Temp and the heat should be disbursed through a exhaust cooling fan yes - User interface manual - Noise Lavel (dba) - - Electrical protection provided by fuses in both live and neutral yes - 1/4 supply lines Operating condition: Capable of operating continuously in ambient temperature of 0 to 50 deg C and relative humidity of 15 to 90% in ideal circumstances yes - Arrhythmia detection facility required; minimum gradation of 1 bpm yes - Functional requirements: electricity;AC-powered; bedside device designed to continuously detect
measure
and display a patient’s electrocardiogram through leads and sensors attached to the patient
it also typically display heart rate
the device is typically equipped with audible and/or visual alarms that are triggered when the patien's parameters drop below or exceed pre-set limits: Continuous display of patient ECG and heart rate on screen
Allows display of single
5 lead ECG or simultaneous display of at least 5 waves selected from up to 12 points Operator can set audiovisual alarm levels for low or high heart rate Operates from mains voltage or from internal rechargeable battery yes - Power consumption - - weight - - Power cable length (meter) 3 METER OR MORE - Display must allow easy viewing in all yes - 2/4 ambient light levels Audiovisual alarms required: high and low heart rate; operator variable settings
cardiac arrhythmia
sensor/wire disconnected
low battery yes - STANDARDS AND SAFETY
CERTIFICATIONS Certification(s) Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (Essential Performance Of Electrocardiographs)
C E Approved product - MISCLANEOUS Warranty (years) 3 3.0 power input and frequency 220 To 240V
50 Hz AC - Certificate of Calibration and Inspection to be provided YES - * Bold specifications are the golden parameters. Consignees and Quantity S.No. Consignee Address Quantity Delivery Days 1 K Chandra Mohan 743144
PO - Ichhapur Nawabganj
North 24 Parganas 3 30 EMD Required No ePBG Required No
139826
GEM/2018/B/19954
electrocardiography ekg/ecg units
30 days
West Bengal; 24 Paraganas North
PRODUCT
| Item | Delivery Location | Qty | Delivery |
|---|---|---|---|
| electrocardiography ekg/ecg units Electrocardiography Ekg/ecg Units ( 3 ) Technical Specifications Specification Specification Name Values Allowed Values FEATURES Mobility | - | - | - |
| portability Protective splashproof case for clean storage and safe transport - Heart rate measurement range to be at least 30 to 250 bpm | - | - | - |
| with accuracy better than ± 5 bpm yes - Heart rate trend display of at least previous 24 hours yes - In built Software and/or standard of communication yes - Heat Dissipiation: Should maitain nominal Temp and the heat should be disbursed through a exhaust cooling fan yes - User interface manual - Noise Lavel (dba) - - Electrical protection provided by fuses in both live and neutral yes - 1/4 supply lines Operating condition: Capable of operating continuously in ambient temperature of 0 to 50 deg C and relative humidity of 15 to 90% in ideal circumstances yes - Arrhythmia detection facility required; minimum gradation of 1 bpm yes - Functional requirements: electricity;AC-powered; bedside device designed to continuously detect | - | - | - |
| measure | - | - | - |
| and display a patient’s electrocardiogram through leads and sensors attached to the patient | - | - | - |
| it also typically display heart rate | - | - | - |
| the device is typically equipped with audible and/or visual alarms that are triggered when the patien's parameters drop below or exceed pre-set limits: Continuous display of patient ECG and heart rate on screen | - | - | - |
| Allows display of single | - | - | - |
| 5 lead ECG or simultaneous display of at least 5 waves selected from up to 12 points Operator can set audiovisual alarm levels for low or high heart rate Operates from mains voltage or from internal rechargeable battery yes - Power consumption - - weight - - Power cable length (meter) 3 METER OR MORE - Display must allow easy viewing in all yes - 2/4 ambient light levels Audiovisual alarms required: high and low heart rate; operator variable settings | - | - | - |
| cardiac arrhythmia | - | - | - |
| sensor/wire disconnected | - | - | - |
| low battery yes - STANDARDS AND SAFETY | - | - | - |
| CERTIFICATIONS Certification(s) Electrical Safety Conforms To Standards For Electrical Safety IEC-60601-1 Shall Meet IEC-60601-1-2 (General Requirements For Safety - Electromagnetic Compatibility) And IEC 60601-2-25 (Essential Performance Of Electrocardiographs) | - | - | - |
| C E Approved product - MISCLANEOUS Warranty (years) 3 3.0 power input and frequency 220 To 240V | - | - | - |
| 50 Hz AC - Certificate of Calibration and Inspection to be provided YES - * Bold specifications are the golden parameters. Consignees and Quantity S.No. Consignee Address Quantity Delivery Days 1 K Chandra Mohan 743144 | - | - | - |
| PO - Ichhapur Nawabganj | - | - | - |
| North 24 Parganas 3 30 EMD Required No ePBG Required No | - | - | - |
Exempted
7 Jun 2018
28 May 2018
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
bid_139826.pdf
GEM_BID
termsCondition.pdf
OTHER
2657c7ca70287299b5181580fdb5a7ab.pdf
OTHER
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .