RequiredOEM Authorization Certificate Bid to RA enabled No Anaesthesia Workstation ( 2 ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to standards CE European CE
RequiredUSFDA
RequiredBIS
RequiredCE Certification date and number - * Certification of manufactures ISO 9001 & ISO 13485 ISO 9001 & ISO 13485 ISO certification number and date - * Confirmity to medical electrical safety standards IEC60601-2-13 and BS EN 6060-1-1 IEC60601-2-13 and BS EN 6060-1-1 1/9 Performance Parameters Purpose of anesthesia workstation Anesthetic machine in used to provide an accurate and continuous supply of medical gases mixed with an accurate concentration of anaesthetic vapour and deliver this to the patient at a safe pressure and flow. Anesthetic machine in used to provide an accurate and continuous supply of medical gases mixed with an accurate concentration of anaesthetic vapour and deliver this to the patient at a safe pressure and flow. Components of Anesthesia workstation Anesthesia machine
RequiredInbuilt ventilator
Requiredvaporisor
RequiredGas delivery system
RequiredMulti para monitor Anesthesia machine
RequiredInbuilt ventilator
Requiredvaporisor
RequiredGas delivery system
RequiredMulti para monitor Type of mounting for the Anesthesia workstation Trolley Mounting Trolley Mounting Material used for Anesthesia workstation Non ferrous metals Non ferrous metals
RequiredPolymer
RequiredEpoxy powder painted steel
RequiredStainless steel
Requiredrust-proof ABS plastic with metal reinforcements Mode of integrated scavenging system with pressure relief valve Passive Active
RequiredPassive Minimum Battery backup for anesthesia workstation 2 hour 45 Minute
Required1 hour
Required2 hour Minimum Battery backup for Ventilator 2 hour 45 Minute
Required1 hour
Required2 hour Minimum Battery backup for monitor 1 hour 45 Minutes
Required1 hour
Required2 hour Facility of manual ventillation in case of electricity and battery failure Yes Yes
RequiredNo Facility of gas and agent delivery in case of electricity and battery failure Yes Yes
RequiredNo All the components likes anesthesia machine
Requiredintegrated ventilator
Requiredvaporizer from the same manufacturer Yes * 2/9 Patient monitor from same manufacturer Yes Yes
RequiredNo Workstation should have minimum of two drawers to keep accessories
Requiredgood mobility
Requiredanti static caster wheels with locking facility & conveniently placed handles for easy movement of the machine Yes * All software updates should be provided free of cost during Warranty period Yes * The anesthesia machine can be used safely up to 1000 gauss line Yes Yes
RequiredNo Should have Air/N₂O interlock Yes Yes
RequiredNo Inspiratory time in sec 0.2 to 3 sec * Expiratory time 0.5 to 6 sec * Flow rate in lpm 0.2 to 6 lpm * Pressure relief 05 to 75 cm H20 * Warranty in years 5 3
Required5 Gas Delivery System Availability of Pin index yokes for gases Oxygen and Nitrous oxide * Availability of Separate Pin index yokes for gases Oxygen
RequiredNitrous oxide
RequiredAir * Availability of cascading type flow meters for O2
RequiredN2O and air Yes * Mode of gas mixing Pneumatic Pneumatic Availability of rotameter control guards in anesthesia workstation Yes * Availability of pressure gauges for cylinder and pipelines in anesthesia workstation Yes * Availability of audible and visual oxygen failure alarms Yes * 3/9 Minimum emergency oxygen flow in liter per minute 35-70 * Type of hypoxic guard with automatic cutoff of N2O Mechanical Electronic
RequiredMechanical Vaporizer Number of quick mount type vaporizer for easy interchangeability at a time 2 * Number of selectable back bar 2 * Facility of interlock for vaporizer Yes * Volume percentage of delivery range 0 to 10 * Minimum agent capacity for free volatile anesthetic agent in ml 225 * Sensor connection shall be internal to prevent disconnection Yes * Vaporizer compatible to gases Sevoflurane
RequiredIsoflurane
RequiredDesflurane
RequiredHalothane Sevoflurane
RequiredIsoflurane
RequiredDesflurane
RequiredHalothane The vaporizer design should be maintenance free Yes * Vaporizer calibration requirement After every 5 years * Breathing System Breathing system should have fresh gas de-coupled compensation
Requiredclosed circle absorber system Yes * Range of pressure relief valve in mbar of breathing system 5 to 75 * Number of steps in change over from spontaneous to bag ventilation single step * Number of reusable autoclavable closed silicone circuit supplied 2 for adult and 2 for pediatric * Number of bains circuit (adult) supplied with 15 10
Required15
Required20 4/9 workstation Number of Jackson Rees circuit for pediatric patients supplied with workstation 15 10
Required15
Required20 Volume capacity of breathing system in liters 1.5 1.5
Required1.7 Breathing system should have leak and compliance test Yes * Anesthesia Ventilator Available modes of operating ventilator Manual/spont
Requiredvolume controlled
Requiredpressure controlled
RequiredSIMV/PS Manual/spont
Requiredvolume controlled
Requiredpressure controlled
RequiredSIMV/PS
RequiredCMV
RequiredSIMV
RequiredAC
RequiredSign Ventilator should be capable to pressure support Yes Yes
RequiredNo Integrated ventilator shall not require bellows to be change for adults and infants Yes * Tidal volume of ventilator in ml 20 to 1400 ml * PEEP of ventilator in mbar 0 to 20 mbar * Breathing frequency in BPM 4 to 60 BPM * Inspiratory pause percentage of Ti 0-50% * Screen size of Ventilator monitor in inches 10 8
Required10
Required12
Required15
Required7 Integrated ventilator shall be able to provide facility to monitor ETCO2
RequiredAutomatic identification of agents Yes * Ventilator should be able to ventilate with atmospheric air if total gas supply failure Yes Yes Monitor Type of monitor LCD TFT
RequiredLCD Size of monitor in inches 12 12
Required15
Required19 Monitor should not require lengthy start up Yes * 5/9 procedure or calibration for display of parameters The graphical and numerical trend with split screen duration 24 hours * Patient multipara monitor shall be able to display parameters HR
RequiredSpO2
RequiredNIBP
RequiredECG
RequiredTemp HR
RequiredSpO2
RequiredNIBP
RequiredECG
RequiredTemp
RequiredHR
RequiredSpO2
RequiredNIBP
RequiredECG
RequiredTemp
RequiredAgent identification (MAC)
RequiredHR
RequiredSpO2
RequiredNIBP
RequiredECG
RequiredTemp
RequiredAgent identification (MAC)
RequiredNIMT/BIS
RequiredEntrop y NMT to be displayed Yes Yes
RequiredNo Dual temperature to be displayed Yes Yes
RequiredNo BIS/ENTROPY to be displayed Yes Yes
RequiredNo Capnography to be displayed Yes Yes
RequiredNo EtCO2 to be displayed Yes Yes
RequiredNo IBP1 and IBP2 to be displayed Yes Yes
RequiredNo CO(cardiac output) to be displayed Yes Yes
RequiredNo ECG monitoring with standard lead configuration
RequiredI
RequiredII
RequiredIII
RequiredavL
RequiredavF&V filters so that it is protected against defibrillator
Requiredelectro surgery potentials
Requiredarrhythmia analysis and ST segment analysis with trends Yes * NIBP should be determine with oscillometric method
Requiredshould read systolic
Requireddiastolic and mean arterial pressure and pulse rate Yes * Pulse oximetery with sensors suitable for adults
Requiredpediatric& neonate Yes * ETCO2 measurement Yes * 6/9 should be by infra –red absorption technique
Requiredmeasuring mode should be side stream or main stream and response time should be less than 3 seconds Monitor should be able to monitor and display all parameters in single screen Yes * Make of monitor if monitor provided is not from the same OEM and if from same OEM put NA - * System to be supplied with oxygen monitor Yes Yes
RequiredNo Alarms Number of alarms 16 * Alarms shall be adjustable to high/low limits with audio and visual alarms Yes * Power supply Length of main cable in meter with power socket of standard indian make to be provided 6 * Power supply 220-240V AC
Required50 Hz single phase * Additional Accessories Transparent re-usable silicon face mask size 0
Required1
Required2
Required3
Required4 & 5 2 each * Breathing bag 2 Litre & 500 ml 2 each * 3/5 ECG lead 2 * Nasopharyngeal/rectal and skinTemperatureprobes should be Provided 2 * Re usable adult cuffs
Requiredreusable Paediatric cuff and sets of Disposables cuff for neonatal 2 * ETCO2 line and adapters 15 each * Mox regulator with 3 meter high Pressure tubing 1 * 7/9 Provision for laryngscopes Yes (adult and pediatric) Yes (adult and pediatric)
RequiredNot Provided Reusable patient circuits 2 1
Required2 Reusable masks 3 2
Required3
RequiredNot provided
RequiredAdult & Pediatric i piece on each sizes 0
Required1
Required2
Required3
Required4 & 5 SPO₂ sensor provided 1 1
RequiredNot provided
Required2 NIBP cuff 1 1
RequiredNot Provided Skin and nasopharyngeal temperature probe 1 1
RequiredNot provided ETCO2 sample lines 1 1
RequiredNot provided
Required10
Required25
Required10 Reusable IBP interface cables 1 1
RequiredNot provided Disposable IBP transducers 1 1
RequiredNot provided
Required5
Required10 BIS sensors 10 10
Required20
Required25
RequiredNot provided Accessories for wall mounting Yes Yes
RequiredNA (If trolley mounting) * Bold specifications are the golden parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer.