Loading…
Loading…
| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | L1₹8.7 LAccepted-AOC | L1 | Accepted-AOC Accepted | |
| 2 | L2₹9.9 L+₹44,200 (4.69%)Rejected-AOC | L2 | Rejected-AOC Rejected | |
| 3 | Rejected-Technical | - | Rejected-Technical Rejected due to incomplete technical bid |
Tender Value
₹8.1 L
EMD Value
₹16,140
Closing Date
17 Sept 2020, 3:00 pmClosed
Sumit Garg, Executive Engineer
Dussehra Ground, Model Town, Yamuna Nagar
Supply and Installation of UV Visible Spectrophotometer, Ph,Cond,TDS , Turbidity Meter and ION Meter complete in all respect and all other works contingent thereto
2020_HRY_143254_1
SE(A) No. 10634 Dt 6.8.2020 (YNR2)
Open Tender
Civil Works
Works
30 days
Yamunanagar
Please refer Tender documents.
2 documents required · 2 mandatory
₹1,000
Yes
Engineer in Chief
₹16,140
Yes
19 Oct 2020
9 Sept 2020
18 Sept 2020
9 Sept 2020
17 Sept 2020
9 Sept 2020
Amount
N.S.Items
Supply of UV-Visible Spectrophotometer Pre- Programmed Water analyzer complete in all respect. Make- HACH Model 6000 & Lovibond Model XD-7500 (with calibration certificate)- Technical specification given as under:- (Parameters- Specifications) : Wavelength range-190 – 1100 nm (scan range), Optics- Monochromator with grating and step motor / reference beam, Lamp- Xenon Flash Lamp (500 million flashes possible) with five years warranty, Display- Backlit 7” colour display for extensive graphical evaluation, Operation- Membrane Keyboard (Alpha Numeric Membrane Keypad), Stray Light-< 0.05 % Transmission at 340 and 408 nm, Bandwidth- 4nm, Measurement- concentration, single and multi-wavelength measurement of absorbance and % transmission, kinetics, spectra, Turbidity correction- The instrument should have automatic turbidity correction in Conc. Mode, Scan speed-700–2000 nm/min, Scans in 1, 2, 5, 10 nm steps of wavelength range, Wavelength accuracy/Reproducibility-± 1 nm / < 0.5 nm, Photometric accuracy/Reproducibility-– 0.003 E for E < 0.600; 0.5 % of values for 0.600 < E < 2.000, Curette-16 mm round, 10 mm, 20 mm, 50 mm rectangular with automatic detection w/o adapter additional 13mm and 24 mm cells could be used, Bar code- Automatic method recognition including measurement range for all corvettes, external bar code reader could be added, Data storage-5000 measurement values, spectra and kinetics approx. 40 MB => 500 spectra (300–900 nm) and 400 kinetics with 150 measurement values, Methods and profiles-> 150 pre-programmed methods, 100 user-defined methods, comprehensive programming options , Instrument should be pre-programmed for Iron analysis at 510 nm with 1,10 Phenanthroline method as per APHA, Instrument should be pre-programmed for DO atleast upto 10ppb-800ppb. Sulphate upto 1000 ppm, Manganese, Chlorine, COD from 0 - 15000 ppm and other water analysis parameters, Interface/update-1 USB-A, 1 USB-B, 1 Ethernet(LIMS Connectivity ASCII,.CSV-files),External Barcode Reader and Scanner/ Update via Internet and USB stick, IP Class- IP 30 including drainage in optical compartment, Temperature Range- Operation: +10 °C to +35 °C, Storage: –25 °C to +65 °C, Device Check (Validation facility of Instrument and method)- Instrument should have internal Analytical quality assurance system with additional options for customising the Analytical quality assurance tools, AQA with conjunction of method and AQA for photometric determination other ingredients in sample, Certification- Instrument must have ISO certification for production for water testing kits and instruments from manufacturer, CE- CE Conformation as per EU guidelines, Wavelength- Wavelength (NM) should be displayed while measurement of sample in concentration mode, Stray light correction- at 340nm GG375 filter tolerance of T should be 0.04-0.026%, Neutral density filter- Neutral glass F2 at 546nm should be less than ± 0.24 ABS to ± 0.003 ABS, Built-in Security Level- Device supports the assignment of passwords and assignment of upto 3 different user levels(Guest,User,Admin), Manual-Hard Copy of Manual all the Pre-programmed methods test procedure with working principle of Instrument Do's and Don't during the analysis interference study and bar-coding of all the test procedure with pictorial diagram should be in the manual, Barcode Reader- External Barcode Reader Required to Scan the Pre-programmed Methods test Procedure
Supply of Ph,Cond,TDS Make- HACH / Lovibond complete in all respect (with Calibration certificate) - Technical specification given as under:- Parameter - Range, Resolution, Accuracy, pH -0-14, 0.01, ± 0.02 + 2 digits, ORP- -1999 to +1999 mV, 1 mV, ± 0.5% + 2 digits, Conductivity - 0-200 μS/cm, 0.1 μS/cm, ± 2% + 1 digit, 200-2000 μS/cm, 1 μS/cm, ± 2% + 1 digit, 2-20 mS/cm, 0.01 mS/cm, ± 2% + 1 digit, 20-200 mS/cm, 0.1 mS/cm, ± 2% + 1 digit, TDS - 0-132 mg/l, 0.1 mg/l, ± 2% + 1 digit, 132-1320 mg/l, 1 mg/l, ± 2% + 1 digit, 1320-13200 mg/l, 10 mg/l, ± 2% + 1 digit, 13200-132000 mg/l, 100 mg/l, ± 2% + 1 digit, Ambient Conditions- 0-50 ° C; 0-80% rel humidity (non condensing), Display Size- 58mmX34 mm, Power Supply - Battery 9V & Adaptor, Dimensions- 208 x 110 x 34, Weight- 380g, Conformity- CE, Interface- USB, RS 232, Datalogger - Real time Data Logger, Software- Data Acquisition Software, Data Memory- Auto or Manual data Memory 16000 data sets only with additional software, Confirmatory- CE, Protection Class- IP51, Auto Off - Yes
Supply of Turbidity Meter Make- HACH / Lovibond complete in all respect (with Calibration certificate): Technical specification given as under: (Parameters- Specifications) :- Measurement cycle- approx. 8 seconds, Display- backlit LCD (on key press), Optics- temperature-compensated LED ( l= 860 nm) and photo sensor amplifier in after proof sample chamber, infrared light, Keypad- polycarbonate membrane, splash proof, Power supply-9 V power pack battery, Auto – OFF- automatic switch-off, Storage- internal ring memory for 16 data sets, Additional feature- real time clock and date, Range-0,01 - 1100 NTU, Resolution-0.01 - 9.99 NTU = 0.01 NTU 10.0 - 99.9 NTU = 0.1 NTU,100 - 1100 NTU = 1 NTU, Accuracy- ± 2,5 % of reading or ± 0.01 NTU (0 - 500 NTU) ± 5 % (500 - 1100 NTU), Housing- ABS, Ambient conditions- Temperature: 5 – 40 °C rel. humidity: 30 – 90%, Confirmatory- CE
DARSHAN ENGINEERING WORKS
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.PDF
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .