Superintending Engineer, Project Implementation Unit, Health & Family Welfare Department
Supply, Installation, Testing and Commissioning of Semi Modular Operation Theatre at DH Gandhidham,Tal. Gandhidham Dist. Kutch
274280
138/05/2025-26
Open
Civil - All
Jamnagar
11 documents required · 8 mandatory · 3 optional
₹1,500
Yes
PIU – General Fund A/C 11-114
₹38,000
Yes
7 Feb 2026
7 Feb 2026
7 Feb 2026
16 Feb 2026
7 Feb 2026
Name of Work : Supply, Installation, Testing and Commissioning of Semi Modular
Operation Theatre at DH Gandhidham,Tal. Gandhidham Dist. Kutch
Quantities Total Amount
Quantities Total Amount
estimated In Figure According to
NBO NO. but may be Description Unit
No. In Words estimated
less Rs. Ps. quantities Rs. Ps.
Supply, Installation, Testing and Commissioning of
mm thick Pre Fabricated Modular Operation Theater
Panels (Walls and Ceiling) made out Polyurethane
foam injected at 40 Kg/m3 sandwiched between
0.8 mm thick GP and SP sheet on front side and
mm GP and SP/PPGI sheet on back-side. The inner
surface walls be fixed to the bricks wall with essential
support. There will be minimum possible cavity/gap in
between solid and steel walls and will be variable to
suit the architects lay out for the flush mounting of the
equipments. The wall and ceiling panels must comply Rupees Six Thousand Sq
with necessary provision of to JIS Z 2801/ISO 22196, Five Hundred Fifty Only Mtrs.
52210-2:1984-08, DIN 4102-1, Class A2. Any Third-
party compliance certificate mandatory. Panel should
be covered with protective sheath of 300 d.f.t Anti-
bacterial Paint having undergone Washability/Water
resistance Test JIS K 5663.7.12, permeability test less
than 200 unit of permeability, Anti-bacterial effect
against staphyloccoccus aureus, Escherichia coli (via
HALO test) and safety and causative agents such as
formaldehyde, toluene, xylene, paradichlorobenzene
The flooring within the surgical theatre should consist
of a minimum of 2mm thick antistatic seamless PVC,
ensuring durability and safety. This floor should be
smooth, non-slip, and made from an impervious
material with sufficient conductivity to dissipate static
electricity effectively, yet not so conductive as to risk
electric shock to personnel. The PVC used should be of
very high quality, particularly designed for
electrostatic charge dissipation, and should not be less
than 2 mm in thickness. The installation process
should utilize continuous rolls, and joints should be Rupees Four Thousand Sq
seamlessly welded using special PVC thermal welding Two Hundred Fifty Only Mtrs.
units with welding bars of the same color to ensure
uniformity and prevent contamination. The sheets
should be highly durable, capable of resisting shocks,
indentations, and scratches. The conductive material
should be uniformly impregnated within the PVC as
grains to enhance its antistatic properties.
Additionally, the flooring should be inert to body
fluids, chemicals, and disinfectants, and remain
unaffected by temperature variations typical within an
operating theatre. It should efficiently discharge
electric charges up to 2 kV, and the electrical
The door to the surgical theatre should be a
hermetically sealed, single sliding door measuring
meters in height and 1.5 meters in width. The door
should be equipped with a controller capable of being
operated by elbow switches, foot switches, and a
touchless sensor, providing multiple, accessible control
options. The track for the door should be made from
stainless steel or aluminum, with the running surface
for the top rollers angled appropriately to minimize
resistance and facilitate smoother movement. The
door leaf should be suspended using high-quality hard Rupees Two Lac Ninety
plastic rollers with double bearings at the top. These Five Thousand Only
rollers should run under the stainless steel or
aluminum track to ensure smooth and noiseless
operation. The opening and closing mechanisms of the
door should be controlled by a microprocessor for
precise electromechanical movement. The material of
the door should be either SMS or HPL, with a color that
matches the interior design. The door should be
designed to be strong yet lightweight for ease of use.
In case the automatic mechanism fails, the door
should still be able to open and close effortlessly
manually. The handle of the door should be made
Hatch boxes should be provided in each operating
theater to facilitate the safe and efficient removal of
waste materials to the dirty linen area or adjacent
corridor. Each hatch box should be equipped with two
doors, operated electrically or motorized, designed
such that only one door can be opened at a time to
prevent cross-contamination. UV lighting should be
installed to ensure sterilization and should remain on Rupees One Lac Fifteen
when both doors are closed, automatically turning off Thousand Only
when either door is opened. Indicators should be
provided on both sides of the operating theater to
monitor the door open/close status. The material of
the hatch box should also be SS304 grade stainless
steel, with a minimum size of 600mm x 600mm.
Supply, Installation, Testing and Commissioning of
Double-Bay Scrub Sink - The scrub station should be
compact, designed for use in the OT complex, and
facilitate pre-procedural scrubbing. It should be made
from heavy gauge type 304 stainless steel with a
minimum thickness of 1.5mm and feature seamless
welded construction polished to a satin finish. The
scrub sink should have a front access panel for easy
maintenance of water control valves, waste
connections, stoppers, and strainers. Hands-free
operation should be enabled through infrared sensors Rupees One Lac Twenty
with programmable adjustments. Thermostatic mixing Six Thousand Only
valve control should maintain constant water
temperature and be located behind the access panel.
Timing should be adjustable to meet individual usage
requirements, and units should have reduced anti-
splash fronts. An additional knee/foot-operated switch
should also be provided for enhanced functionality.
Supply, Installation, Testing and Commissioning of
Storage Units - Storage units should be constructed
with same material, as wall and ceiling panels. The
shelves, made of glass and, should be removable for
cleaning. The unit should be divided into two or more
Rupees Eighty Five
6 2.00 parts, each with individual glass doors equipped with a 85000.00 Nos.
high-quality locking system. The overall size of the
storage unit should
be approximately 200 cm (H) x 120 cm (W) x 35 cm
AIR HENDALING UNIT Make:KrVelocity: Not More
Then 10 mt/second. , Canvass Connection-Heavy duty,
non partical shreding,Fire Retardentuger,Nikotra only
Type: PLUG,Drirect Driven,Dynamically Balanced
Selection of fan-Should run comfortably at 50 HZ at
rated capecity. Motor
Make:Siemence,B.B,Crompton,ABB , Type:TEFC, IP:55,
Foot Mounted, VFD compitable and I2 generation
Safety: Motor should have interlocked with AHU
Motor door/excess/Fire DamperType-TWO TIRE, Floor
Rupees Twelve Lac
Mounted , Inner const.:0.6mm plain G.I , Outer
7 1.00 1225000.00 Twenty Five Thousand job
const.:0.8 mm G.I. Pre coated / Powder coated
Thickness of Insulation: 25mm,PUF injected having
density 40 to 42 kg/m3. ,Colour:As approved by
CLIENT CONDENSING UNIT :
Condensing unit for AHU with all assesories and
instalation (Copper pipie provide 3 mtr Only Extra will
charged extra) consider 1200 sqfeet insulation
Double Dome OT Light with Ceiling mounted Double
Dome LED OT lights having following features and
enclosed technical specifications, having Mobile App
Connectivity, Battery Backup
-Light intensity Ec - 160 000 & 160 000 lux, - Light
intensity adjustment - 10 – 100% (using Digital
Capacitive Touch (Fully Remote Controlled)
- Color temperature Tc (with adjustment)
-Light field diameter d10 at Ec - 100 - 150 mm
-Working range - 1 m Rupees Three Lac Ninety
-Color rendering index 93Ra Thousand Only
- Focusing Preset: Adjustable
- LED Average life - 50000-60000 hrs
- Diameter of Light - not less than 650 mm per dome
- Power Consumption - not more than 120W
OT Light should be BIS marked; manufacturer should
have CDSCO License for this product under Class A,
Compliant to IEC standards such as IEC60601-
1:2005+A2:2020, IEC 60601-2-41:201 (Relevant Test
Certificates to be submitted with Bid) Clause-wise
Technical Compliance to be
Single-arm surgery/anesthesia pendant designed for
Operating Theatres should feature a compact, modular
configuration to meet the precise requirements of
hospitals, enabling customization to suit specific
operational needs of both Operating Theatres and
Intensive Care Units. It should be equipped with
robust ceiling support capable of mounting a selection
of arms. The extension arm, pivotal for functionality,
should allow horizontal movement with a radius of
800-900mm and provide a level rotation ranging from
0 to 330/340 degrees, with both the extension arm Rupees Two Lac Eighty
and terminal box designed to rotate together and Four Thousand Only
independently for maximum versatility. The pendant
should support a net loading capacity of min. 100 kg,
accommodating a variety of medical instruments and
devices securely. For medical and electrical services, it
should offer configurations that include at least two
oxygen outlets, two vacuum outlets, one medical air,
one medical air at 7 bar, and an optional nitrous oxide
outlet (In-scope of MGPS vendor) Additionally, there
should be six main sockets, each capable of supporting
5 to 15 amps, conforming to universal standards to
facilitate the operation of various medical and TOTAL
Rupees Thirty Seven Lac Sixty Nine Thousand Four Hundred Fifteen Only
I / We am / are willing to carry out the work at
percent ( Should be written in figures and words) of the Estimated rates mentioned above. Amount of my/our tender works out as under.
Estimated Amount put to tender. Rs.37,69,415/-
1 All work shall be carried out as per Public Works Department Handbook and other specifiations of Division or as directed.
2 All the column is Schedule should be filled in ink and the entries in the last column should be struck by the cotractor under his signature.
Rates quoted include clearance to site (prior commencement of work and at its close) in allrespects and hold good for work under all conditions,
site, moisture, weather etc.
Serviceble material available (if any) after Demolishing / Dismentalling any item should be hand over to the concerned PIU Site Engineer In
Charge. No Extra Payments will Done for any Demolishied or Dismentaled old Struture if require.
Signature of the contractor with seal: Signature of the Contracting Authority with seal:
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
RAW_HTML
3 GTS Ele..pdf
4 GTS Civil.pdf
5 Standard of Work_Applicable_Jamnagar.pdf
6 ijaradar-ok.pdf
7 1465 કામની સાઈટ ઉપર રાખવાની થતી કાળજી બાબત. 02-03-2021.pdf
8 B1 2 Clause 8A GR PIU.pdf
11 Defects Libility GR R B.pdf
12 INSTRUCTIONS TO PERSON TENDERING.pdf
13 Labour Cess - GR - Page-01 (1).pdf
14 Labour Cess - GR - Page-02 (1).pdf
15 Labour Cess - GR - Page-03.pdf
16 SPECIAL CONDITIONS.pdf
19 6680 બ્રાન્ડ - મેકલીસ્ટ_compressed.pdf
9 B1 2 Clause 10 GR PIU.pdf
1. SBD OT DH Gandhidham.pdf
2. Schedule B OT DH Gandhidham.pdf
Download all tender documents and submit your bid