Name Of Work :- Upgradation of Exiting 4 Nos. Operation Theatre into Moduler Operation Theatre (SURGERY OT-1, Trauma OT-1, Eye OT, ENT OT) at PDU Hospital,
Memorandum Showing Items of Works to be Carried Out
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
Supply, Installation, Testing and Commissioning of 50 mm thick Pre Fabricated Modular Operation Theater Panels
(Walls and Ceiling) made out Polyurethane foam injected at 40 Kg/m3 sandwiched between 0.8 mm thick GP and SP
sheet on front side and 0.5 mm GP and SP/PPGI sheet on back-side. The inner surface walls be fixed to the bricks wall
with essential support. There will be minimum possible cavity/gap in between solid and steel walls and will be variable
to suit the architects lay out for the flush mounting of the equipments. The wall and ceiling panels must comply with
necessary provision of to JIS Z 2801/ISO 22196, BS EN 1363-1:2020, EN 1363-2:1999, EN 1364-1:2015, DIN 52210-
Rupees Six Thousand Five
1 1 644.00 2:1984-08, DIN 4102-1, Class A2. Any Third-party compliance certificate mandatory. Panel should be covered with 6550 Smt
hundred Fifty Only
protective sheath of 300 d.f.t Anti-bacterial Paint having undergone Washability/Water resistance Test JIS K 5663.7.12,
permeability test less than 200 unit of permeability, Anti-bacterial effect against staphyloccoccus aureus, Escherichia
coli (via HALO test) and safety and causative agents such as formaldehyde, toluene, xylene, paradichlorobenzene etc.
Third party (NABL approved Lab) certification for EGP and thickness test certificate to be submitted at time of supply.
Supply, Installation, Testing and Commissioning of Ceiling Filtration System - The ceiling filtration system in the surgical
theatre should ensure a unidirectional distribution of sterile air, covering all areas affected by the airflow. It should
consist of a laminar flow system incorporated within a frame made from thick extruded aluminum profiles, complete with
a sealed gasket to prevent leaks. The system should incorporate "H 14" class HEPA filters positioned in the ceiling
plenum. These filters should achieve an MPPS average efficiency greater than 99.95% and a 3-micron DOP efficiency
greater than 99.99%. The filters should also be designed to handle a final pressure drop of no more than 600 Pa,
operate at a maximum temperature of 60 degrees Celsius, and function effectively within a relative humidity range of 40-
50%. The filtration ceiling system should be capable of maintaining a constant airflow velocity of 0.25 m/s at the
diffuser. This should be accomplished through the use of a diffuser or flow equalizer that ensures uniform and constant
air distribution over the entire surface area. The structure holding the filtration system, including the filter frames and top
plenum, should be made from either aluminum or stainless steel. The system should be CE/UL certified and designed Rupees Three Lakh Ninety
to achieve an ISO 14644/1 classification of ISO 5, with a bacteriological class of B (5 CFU/m3) and particle thousand Only
decontamination kinetics CP of 5 minutes. After installation, the system's performance should be validated by a third-
party government-approved environmental lab to ensure compliance with specified standards. A positive pressure
should be maintained inside the operating theatre to prevent contamination from external air sources. The supplier
should provide test certificates for the HEPA filters and the laminar airflow systems from the original manufacturers.
Furthermore, the size of the laminar airflow system should be a minimum of 8 feet by 8 feet. Finally, the entire system
should meet CE/BIS certification standards, ensuring it adheres to both European and Indian safety and performance
regulations and including validation.
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
The flooring within the surgical theatre should consist of a minimum of 2mm thick antistatic seamless PVC, ensuring
durability and safety. This floor should be smooth, non-slip, and made from an impervious material with sufficient
conductivity to dissipate static electricity effectively, yet not so conductive as to risk electric shock to personnel. The
PVC used should be of very high quality, particularly designed for electrostatic charge dissipation, and should not be
less than 2 mm in thickness. The installation process should utilize continuous rolls, and joints should be seamlessly
welded using special PVC thermal welding units with welding bars of the same color to ensure uniformity and prevent
contamination. The sheets should be highly durable, capable of resisting shocks, indentations, and scratches. The
conductive material should be uniformly impregnated within the PVC as grains to enhance its antistatic properties.
Additionally, the flooring should be inert to body fluids, chemicals, and disinfectants, and remain unaffected by
temperature variations typical within an operating theatre. It should efficiently discharge electric charges up to 2 kV, and
the electrical resistance (point to ground) should be maintained within 2.5x10^4 to 5x10^6 ohms, ensuring no buildup of
electrical charge beyond 100 volts. Floor installation should be performed by skilled workers from accredited agencies
authorized by the PVC sheet supplier. The installation should include a conductive copper grid beneath the PVC sheet, Rupees Four Thousand Two
supported by liquid epoxy compounds to create a uniform and level surface. Copper strips should be made visible by hundred Fifty Only
grinding and not project more than 0.5mm above the floor level to prevent damage to the PVC sheet, with one earthing
lead brought out from every 150 sq.ft area and attached to the main earthing strip/ground. The flooring should also be
mechanically shockproof, scratchproof, flame retardant, and antimicrobial. Corners of the flooring should be uniformly
curved, and a concealed cove-former (aluminium) should overlap the wall panel to a height of approximately 25mm and
be sealed perfectly and uniformly. Before laying the PVC, suitable self-leveling compounds should be used to prevent
any undulations within the operating theatre. The final surface should be non-corrosive to biological fluids and
detergents, with a color that is uniform, pleasant, and matching with the ambience as approved by the respective
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
The door to the surgical theatre should be a hermetically sealed, single sliding door measuring 2.1 meters in height and
1.5 meters in width. The door should be equipped with a controller capable of being operated by elbow switches, foot
switches, and a touchless sensor, providing multiple, accessible control options. The track for the door should be made
from stainless steel or aluminum, with the running surface for the top rollers angled appropriately to minimize resistance
and facilitate smoother movement. The door leaf should be suspended using high-quality hard plastic rollers with
double bearings at the top. These rollers should run under the stainless steel or aluminum track to ensure smooth and
noiseless operation. The opening and closing mechanisms of the door should be controlled by a microprocessor for
precise electromechanical movement. The material of the door should be either SMS or HPL, with a color that matches
the interior design. The door should be designed to be strong yet lightweight for ease of use. In case the automatic
mechanism fails, the door should still be able to open and close effortlessly manually. The handle of the door should be
made from stainless steel with a matte finish and equipped with a high-quality cylindrical or ESPg lock for security. The
door leaf should incorporate a high-quality synthetic rubber gasket with long durability to ensure hermetic sealing and
maintain a 99.99% air tightness at a minimum pressure of 75Pa. The floor on either side of the door should be perfectly Rupees Two Lakh Ninety Five
4 4 8.00 level, with a maximum permissible difference of +1mm, to avoid any tripping hazards or unsealed gaps. The overall 295000 Nos.
thickness of the finished door should range from 30-60mm, with the inner part of the door filled with CFC-free
polyurethane foam, approximately 48mm thick, sealed airtight to block any microbial ingress. The door and its controls
should comply with IEE regulations, utilizing DC brushless or PMDC motors that are effectively isolated from mains
power to ensure safety and reliability. Additionally, the door should include a vision window measuring 300mm x
300mm, featuring double-glazed panels that are hermetically sealed to maintain the integrity of the seal.
The movement of the door should be as quiet as possible, and the time to start moving after
receiving a signal should be adjustable between 0.5 to 20 seconds.The door controller should be CE marked, BIS
certified, or IEC 60601 certified to meet international and national safety standards. A test certificate for the hermetically
sealed door frame should be included in the pre-dispatch documents to verify compliance with the specified standards
and functionality.
Supply, Installation, Testing and Commissioning of Membrane-Type Operation Switch and Communication Controller -
The Operation Switch and Comms Controller should encompass all necessary controls to ensure the proper operation
and monitoring of equipment and services in the operating room (OR). Positioned for optimal visibility in the surgeon's
line of sight and the OT staff, the wall-mounted membrane-buttons should be easily accessible to the nurse for
operation and assistance. In addition to hosting a digital clock and elapsed time indicator, the panel should feature a
medical gas alarm, clearly indicating high and low gas pressures for each gas service, including vacuum, supported by
an audible alarm. The Operation Switch and Comms Controller should also provide an alarm mute option for fault
annunciation, with pressure sensors strategically located near isolation valves. Controls for general lighting, including
ON/OFF and dimming, organized in groups for uniform illumination, should be present. Furthermore, the panel should Rupees One Lakh Ninety Five
facilitate control over operating lights (major and satellite) and camera functions (on/off and intensity). A hands-free thousand Only
telephone set with memory should be conveniently situated, and temperature and humidity controls for the room, linked
to the AHU (+/5%) using NO/NC or Chilled Water/Hot Water Actuator Control Valve (whichever is applicable), should
be adjustable from the Operation Switch and Comms Controller. Real-time indicators, such as a digital room pressure
indicator and HEPA filter bank differential pressure indicator, are essential for monitoring and maintaining a safe and
controlled environment in the operating room. Telephone with Alpha-numeric keyboard is required with Speed-dialing
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
In each operating theater, an operating list board should be installed. It should be made of ceramic with magnetic
properties and should be flush with the wall to maintain cleanliness and ease of use. LED type flat panel X-ray viewing
panels should be provided, complying with relevant electrical safety codes, and mounted flush with the wall to prevent
dust accumulation and microbial growth. The body should be made of extruded aluminum, powder-coated with a Rupees Twenty Eight Thousand
bacteria-resistant and disinfectant-resistant finish. The diffuser on the front panel should provide a uniformly lit screen, Five hundred Only
and dimming controls should be easily accessible. Additionally, a proper spring-loaded film clip with rollers should be
included to securely hold and easily remove films without causing scratches.
Supply, Installation, Testing and Commissioning of Double-Bay Scrub Sink - The scrub station should be compact,
designed for use in the OT complex, and facilitate pre-procedural scrubbing. It should be made from heavy gauge type
304 stainless steel with a minimum thickness of 1.5mm and feature seamless welded construction polished to a satin
finish. The scrub sink should have a front access panel for easy maintenance of water control valves, waste
connections, stoppers, and strainers. Hands-free operation should be enabled through infrared sensors with Rupees One Lakh Twenty Six
programmable adjustments. Thermostatic mixing valve control should maintain constant water temperature and be thousand Only
located behind the access panel. Timing should be adjustable to meet individual usage requirements, and units should
have reduced anti-splash fronts. An additional knee/foot-operated switch should also be provided for enhanced
Supply, Installation, Testing and Commissioning of Storage Units - Storage units should be constructed with same
material, as wall and ceiling panels. The shelves, made of glass and, should be removable for cleaning. The unit should
Rupees Eighty Five Thousand
8 11 11.00 be divided into two or more parts, each with individual glass doors equipped with a high-quality locking system. The 85000 Nos.
overall size of the storage unit should
be approximately 200 cm (H) x 120 cm (W) x 35 cm (D).
Supply, Installation, Testing and Commissioning of Peripheral Lighting - To enhance the operational environment of
each operating theatre, peripheral lighting and clean room luminaries should be installed, with a minimum of eight units
per theatre. These lights should offer an intensity of at least 500 Lux and be housed in highly specular anodized
aluminum with an optical antiglare system to minimize visual fatigue and glare. Each peripheral light should measure 2'
x 1'. The covers for these luminaries should be constructed from highly resistant, disinfectant-proof laminated safety
glass or acrylic, featuring a stylish, fine-grained surface that complements the clean and sterile environment of the
operating theatre. The body of the luminaries should be made from sheet steel, perfectly powder-coated to resist
contamination and facilitate easy cleaning, and should be ready for connection either as individual units or within a
series circuit. This setup should include digital electronic control gear utilizing multilamp technology to optimize energy
9 12 64.00 12000.00 Rupees Twelve Thousand Only Nos.
efficiency and light output. The installation of these luminaries should ensure that the recess frames are gas tight. Each
fitting should be flush with the ceiling and designed to be removable from either the top or bottom for ease of
maintenance. It is crucial that the light fittings are uniformly and aesthetically distributed across the ceiling to provide
consistent illumination throughout the OT, without interfering with visibility during procedures such as green mode
endoscopy. Furthermore, the peripheral lighting system should conform to IP65 or IP54 international protection ratings,
ensuring that the fixtures are dust tight and protected against water ingress, suitable for the demanding conditions of an
operating theatre.
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
Supply, Installation, Testing and Commissioning of Electrical Wiring inside OT - Power distribution within each
Operating Theatre (OT) should be handled via distribution boards located locally to each theatre using Fire Resistance
Low Smoke Wiring, layed in PVC conduits of reputed brand. The sub mains power to these panels should be provided
by the general electrical contractor, and all distribution services within the departments should run from these panels.
The power distribution unit should hold certifications such as EN, CE, FDA, IEC, and BIS to ensure compliance with
international safety and performance standards. Each wall of the MOT should have a minimum of two hybrid switch
sockets (6/16A) on each wall, in compliance with IEC standards. Light fittings within the clinical areas should be
recessed LED types equipped with control gear and sealed in accordance with the IP54 standard. This standard
ensures that the fittings are protected against limited dust ingress and water spray from all directions, suitable for
clinical environments. All equipment should be fully and permanently labeled to clearly identify and describe the
function, operation, and voltage of the apparatus concerned. Upon completion of the electrical installation, tests should Rupees Two Lakh Ninety Nine
be carried out in accordance with the relevant sections of local wiring regulations, with results duly recorded. Each OT thousand Only
should have a distribution board (DBs) installed, with suitable wiring. A separate enclosure should house all high-
voltage equipment to ensure safety and isolation. The remote cabinet should contain operating lamp transformers,
mains failure relays, UPS, electrical distribution equipment, and circuit protection equipment for all circuits within the
operating theatre. All internal wiring within the DBs should terminate in connectors equipped with screws and clamp
springs. Connections should be of the clip-on type mounted on a rail approved by CE/IEC/BIS standards and labeled
with indelible proprietary labels. All internal circuits should be protected by individual fuses or miniature circuit breakers
to ensure safety and manage potential electrical faults effectively.
Single-arm surgery/anesthesia pendant designed for Operating Theatres should feature a compact, modular
configuration to meet the precise requirements of hospitals, enabling customization to suit specific operational needs of
both Operating Theatres and Intensive Care Units. It should be equipped with robust ceiling support capable of
mounting a selection of arms. The extension arm, pivotal for functionality, should allow horizontal movement with a
radius of 800-900mm and provide a level rotation ranging from 0 to 330/340 degrees, with both the extension arm and
terminal box designed to rotate together and independently for maximum versatility. The pendant should support a net
loading capacity of min. 100 kg, accommodating a variety of medical instruments and devices securely. For medical
and electrical services, it should offer configurations that include at least two oxygen outlets, two vacuum outlets, one
medical air, one medical air at 7 bar, and an optional nitrous oxide outlet (In-scope of MGPS vendor) Additionally, there
should be six main sockets, each capable of supporting 5 to 15 amps, conforming to universal standards to facilitate the
operation of various medical and electronic devices. In terms of materials and construction, the pendant should be Rupees Two Lakh Eighty Four
crafted from high-strength aluminum alloy with a sealed design to enhance durability and maintain sterile conditions. thousand Only
The arm itself should be made from high-density aluminum alloy extrusion, ensuring both strength and reliability.
Installation of the pendant should be ceiling-mounted to provide a reliable and firm setup, optimizing space usage in the
Operating Theatre or ICU while seamlessly integrating with existing medical systems for managing medical gas,
electrical services, data communication, and equipment. Additionally, the pendant should include practical features
such as a stainless steel IV pole and a ward vacuum unit clamp to further enhance its utility in a medical setting. With a
net weight of approximately 105 kg, accessible by the equipment carrier, the pendant should be robust yet
manageable, ensuring it meets the rigorous demands of modern medical environments.
Medical Gas Pipe Line Of Required dia for OT oxygen , Vaccum , Air.. Etc reuired Point(from existing line ) as directed
Rupees Three Lakh Forty Five
12 17 4.00 by engineer incharge from Outer Nearest Passage to OT Pendant Rate Including , Alarm , Valve (for all line) , Point 345000.00 job
,Flowmeter and Probe . if alredy then no payment given
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
AIR HENDALING UNIT Make:KrVelocity: Not More Then 10 mt/second. , Canvass Connection-Heavy duty, non
partical shreding,Fire Retardentuger,Nikotra only , Type: PLUG,Drirect Driven,Dynamically Balanced Selection of fan-
Should run comfortably at 50 HZ at rated capecity. Motor Make:Siemence,B.B,Crompton,ABB , Type:TEFC, IP:55,
Foot Mounted, VFD compitable and I2 generation Safety: Motor should have interlocked with AHU Motor
door/excess/Fire DamperType-TWO TIRE, Floor Mounted , Inner const.:0.6mm plain G.I , Outer const.:0.8 mm G.I. Pre Rupees Twelve Lakh Thirty
13 14 4.00 coated / Powder coated Thickness of Insulation: 25mm,PUF injected having density 40 to 42 kg/m3. ,Colour:As 1237250.00 Seven thousand Two hundred Nos.
approved by CLIENT CONDENSING UNIT : Fifty Only
Condensing unit for AHU with all assesories and instalation (Copper pipie provide 3 mtr Only Extra will charged extra)
consider 1200 sqfeet insulation and ducting
Double Dome OT Light with Ceiling mounted Double Dome LED OT lights having following features and enclosed
technical specifications, having Mobile App Connectivity, Battery Backup
-Light intensity Ec - 160 000 & 160 000 lux, - Light intensity adjustment - 10 – 100% (using Digital Capacitive Touch
(Fully Remote Controlled) - Color temperature Tc (with adjustment) -3800 – 4800 K -Light field diameter d10 at Ec -
100 - 150 mm -Working range - 1 m -Color rendering index 93Ra - Focusing Preset: Adjustable - LED Average life - Rupees Three Lakh Ninety Three
50000-60000 hrs - Diameter of Light - not less than 650 mm per dome - Power Consumption - not more than 120W OT thousand Nine hundred Only
Light should be BIS marked; manufacturer should have CDSCO License for this product under Class A, Compliant to
IEC standards such as IEC60601-1:2005+A2:2020, IEC 60601-2-41:201 (Relevant Test Certificates to be submitted
with Bid) Clause-wise Technical Compliance to be
Hatch boxes should be provided in each operating theater to facilitate the safe and efficient removal of waste materials
to the dirty linen area or adjacent corridor. Each hatch box should be equipped with two doors, operated electrically or
motorized, designed such that only one door can be opened at a time to prevent cross-contamination. UV lighting
Rupees One Lakh Sixteen
15 7 4.00 should be installed to ensure sterilization and should remain on when both doors are closed, automatically turning off 116150.00 Nos.
thousand One hundred Fifty Only
when either door is opened. Indicators should be provided on both sides of the operating theater to monitor the door
open/close status. The material of the hatch box should also be SS304 grade stainless steel, with a minimum size of
Providing and fixing UPS (25KVA) Of Good Quality as directed by engineer incharge Rupees Three Lakh Seventy Five
LED type flat panel X-ray viewing panel should be supplied. This should comply with relevant electrical safely codes.
This should be a 2-panel viewing screen. Mounting should be flush with the wall to avoid dust accumulation and growth
or organisms between wall and panel. Body should be of extruded aluminium powder coated black with bacteria
Rupees Forty One Thousand
17 9 4.00 resistant and disinfectant resistant finish. The diffuser on the front panel should be a uniformly lit screen. Dimming 41500.00 Nos.
Five hundred Only
electronic control should be enclosed at the bottom of the cabinet. Proper spring-loaded film clip with rollers should be
provided to hold the films firmly and to remove the film without scratches
Providing and erecting Approved make Four pole moulded case circuit breaker having breaking capacity ICU of 35 KA.
and above at 415 V . having normal current rating 250A. with variable thermal & magnetic release suitable to work on Rupees Twenty Thousand Nine
18 3-4-4. 4.00 A.C.supply 50 c/s. with all internal connections, spreader tinned copper & complete erected in existing 16 G.M.S. 20946.39 hundred Forty Six Paise Thirty Ea.
housing. ICS=100% of ICU only Cat III Nine Only
Providing and erecting Approved make Four pole moulded case circuit breaker having breaking capacity ICU of 25 KA.
at 415 V,having normal current rating up to 25 A to 100A. with Fixed thermal & magnetic release suitable to work on Rupees Six Thousand One
19 3-4-1. 12.00 A.C. supply 50 c/s. with all internal connections, spreader tinned copper & complete erected in existing 16 G.M.S. 6173.12 hundred Seventy Three Paise Ea.
housing. ICS=100% of ICU only Cat III Twelve Only
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
Supplying and erecting triple pole & neutral 440V / 500V panel mounting Copper Busbars with four equal Nos. of
electrolyte bus having current density not more than 1.6 Amp. / sq.mm (Rated current / cross section area) duly Rupees One Thousand Nine
20 3-8-2. 4.80 wrapped with colour insulating tape for phase sequence of following current carrying capacity, erected with necessary 1961.42 hundred Sixty One Paise Forty Rn Mtr
bus bar supports /insulators, main cable socket to each bar, erected in existing cubical panel with necessary Two Only
connections.(A) Suitable for 100 Amp. Capacity
(B) Suitable for 200 Amp. Capacity Rupees Three Thousand Two
21 3-8-2. 4.00 3202.71 hundred Two Paise Seventy One Rn Mtr
Providing & erecting weather proof, dust & vermin proof, floor mounted front operated indoor type cubical panel board
necessary IP-42 and above protection as per approval from engineer incharge made from 14 SWG thick CRC M.S.
sheet for outer body & doors, 16 SWG thick CRC M.S.sheet for internal partitions with necessary accesories
supporting angles/ flats channel including cutting, bending, drilling, welding, riveting with internal partitions & cable
alley as per requirements & instruction of engineer-in-charge with erection of supplied switch gears, BUSBARS,
suitable size of inter connecting PVC copper wire / copper-aluminium strips, rubber grommets, rib, bakelite control
Rupees Ten Thousand Four
fuses/MCB for measuring instruments, earth bus & earth bolts, foundation flange - bolts-base Plates, sufficient nos. of
22 3-16-1 4.80 10468.65 hundred Sixty Eight Paise Sixty Sq Mtrs.
hinged doors, handles with locking arrangement and rubber gasket,heavy duty end terminal connection,danger notice
board,necessary ventilation,earthing strip complete. The Panel shall be painted with epoxy powder coating. (The rates
excludes the cost of switchgears, bus bars, inter connecting mains & Copper Aluminium strips, meters, Fuses etc. The
dimension shall be measured excluding base beams) The panel shall be supplied with following approved
manufacturers with following size.(A) locally fabricated panel board (i) with 350 mm depth
Supplying & erecting in earthpit of minimum bore dia. 225mm size approved make Safe Earthing Electrode consisting
Pipe-in-Pipe Technology as per IS 3043-1987 made of corrosion free hot dipped G.I.Pipes having Outer pipe dia of
mm having 80-200 Micron galvanising, Inner pipe dia of 40 mm having 200-250 Micron galvanising, connection terminal
dia of 14 mm with constant ohmic value surrounded by highly conductive compound with high charge dissipation
suitable for following type of applications with chamber and heavy duty cover (A)(approved make OEM has to submit
test certificate including value of earth resistance of installation duly stamped and signed by agency and officer
Rupees Five Thousand Eight
Incharge has to ensure the value of earthing resistane mentioned in test Certificate) & having back filling compound of
23 3-14-10 8.00 5842.85 hundred Forty Two Paise Eighty Ea.
(B) Inner chemical (CCM Compound)- Resistivity:- 0.2 ohm / meter testing as per IEC 62561-2017, Voltage drop:- <
volt at no load & dry form, Sulphar content:- <2%(C) Back fill Compound :- Earthing compound should be capable to
retain moisture for long time Necessary test report must be submitted by Agency.[A] For electrical installation up to
Length of Pipe - 1 Mtr
Back filling compound - 1 Nos. of Bag of 15 Kgs.
Providing and erecting Annealed bare Copper wire 8 to 16 SWG.
Rupees Seven hundred Ninety
Four Paise Eighty Seven Only
Providing and erecting XLPE(IS:7098)(I)-88 ISI armoured cable multistrand Aluminium conductor for 1.1 KV. to be laid
Rupees Two hundred Twenty
25 4-3-3. 300.00 on wall with necessary clamps or in existing trench / pipe of following size of cables (B) 3 1/2 core 35 Sq. mm ( 16 Sq. 220.18 Mtr
Paise Eighteen Only
Providing and erecting XLPE(IS:7098)(I)-88 ISI armoured cable multistrand Copper conductor for 1.1 KV. to be laid
Rupees Six hundred Seventy Six
26 4-4-5. 1440.00 on wall with necessary clamps or in existing trench / pipe at road crossing or floor of following size of cables. (A) 4 676.70 Mtr
Paise Seventy Only
Supplying and erecting approved make panel mounting type Digital Voltmeter having 3 digits LED display, 0 to 750 AC Rupees One Thousand Two
27 9-2-2. 4.00 Volts range erected on existing panel board with all connection, wiring etc .with manufacturers calibration certificate. 1275.63 hundred Seventy Five Paise Sixty Ea.
NBO No. Qty Item of Works. In figures Unit. Total Amount
NBO No. Qty Item of Works. In figures Unit. Total Amount
Supplying and erecting approved make panel mounting type Digital Ammeter having 3 digits LED display, external Rupees One Thousand Five
28 9-2-4. 4.00 CT operated, calibrated for 0 to 1000 Amps suitable to operate on 500 Volt AC , erected on existing panel board with 1558.43 hundred Fifty Eight Paise Forty Ea.
all connection, wiring etc .with manufacturers calibration certificate. Three Only
Supplying and erecting Ammeter / Voltmeter selector switch for 3 phase AC Supply 500 V on existing panel board with
Rupees Three hundred
29 9-2-6. 8.00 necessary connections. 317.14 Ea.
Seventeen Paise Fourteen Only
Supplying and erecting approved make suitable panel indicator LED type lamp, lens cover, complete erected with Rupees Forty Three Paise Forty
necessary connections. Three Only
Providing & erecting L.T. Current Transformer with bar primary 50/5 to 1000/5 ratio 15 VA burden erected in existing
Rupees Two hundred Seventy
31 9-2-7. 12.00 CRCA box duly secured with insulating materials connected to the meter 274.72 Ea.
Four Paise Seventy Two Only
Total of Civil work
In Words :- Rupees Two Crore Sixty Four Lakh Thirty Two thousand Three hundred Thirty Seven Paise Fifty Eight Only
I / We am/are willing to carry out the work at _______________________________________________________________________ % Above / Below
percent ( Should be written in figures and words ) of the Estimated rates mentioned above.
Estimated Rate put to tender
Add/Deduct ______________________% above/ below
1 All works shall be carried out as per Public Work Department Handbook and other specification of Division or as directed.
2 All the columns in Schedules should be filled in ink and the total of the entries in last column should be struck by the contractor under his signature
3 Rates quoted include clearance of site ( prior commencement of work and at its close ). In all respect and hold good for work under all condition, site, moisture, weather etc.
4 Serviceable material available ( if any ) after Demolishing / Dismentalling any item should be hand over to the concerned PIU Engineer In Charge as per instruction.
Dy.Executive Engineer, Executive Engineer,
P.I.U.,Zone-I Rajkot. P.I.U.,Zone-I Rajkot.
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
RAW_HTML
01 SBD MOT.pdf
02 SCH-B.pdf
3 PQ for OT.pdf
3A. Specification.pdf
3B. GTS Ele..pdf
4 GTS Civil.pdf
5 Special Condition Schedule-C.pdf
6 6372 Make List HO 15.10.2025-1.pdf
07 Special Condition for Tender.pdf
8 ijaradar-ok.pdf
9 Standard Specification for Electrical Work-1.pdf
10 Labour Cess - GR - Page-01 (1).pdf
11 Labour Cess - GR - Page-02 (1).pdf
12 Labour Cess - GR - Page-03.pdf
Download all tender documents and submit your bid