Loading…
Loading…
| # | Company | Status |
|---|---|---|
| 1 | Admitted-Finance 43 SHIV SAROVER BANK COLONY GARH ROAD MEERUT UTTAR PRADESH 250002 | MEERUT | UTTAR PRADESH | 250002 | Admitted-Finance |
| 2 | Admitted-Finance | Admitted-Finance |
| 3 | Rejected-Technical | Rejected-Technical Bid Under Evaluation |
Tender Value
₹25 L
EMD Value
₹50,000
Closing Date
1 Feb 2020, 2:00 pmClosed
Registrar / Finance officer, CCSU
Office of the FO, CCSU, Meerut
For Supply and installation of Equipments (UV-Visible Double Beam Spectrophotometer (Shimadzu))
2020_CCSUN_426136_1
FC/507 Dt. 16-01-2020
Open Tender
Miscellaneous Services
Item Wise
90 days
Meerut
Tender Fee, EMD and Other documents as per the tender document/NIT
2 documents required · 2 mandatory
₹1,050
Finance Officer, Ch. Charan Singh University
₹50,000
18 Feb 2020
21 Jan 2020
1 Feb 2020
21 Jan 2020
1 Feb 2020
21 Jan 2020
Amount
PC based Double Beam UV-VIS Spectrophotometer working on 220V, 50Hz with variable bandwidth manufactured by an ISO 9001, ISO-14001 certified company, working on 32 bit software with following features:-
1. Setting Wavelength range: 185 to 1400 nm
2. Measurement wavelength range:
Up to 900 nm from 185 nm (standard)
Up to 1400 nm from 220 nm (with Integrating Sphere attachment)
3. Wavelength accuracy: ± 0.1 nm (at 656.1 nm) & ± 0.3 nm for all range
4. Wavelength repeatability : ± 0.05 nm
5. Wavelength setting : At 1 nm units for scan start and scan end wavelengths, And 0.1 nm units for other wavelengths.
6. Wavelength scanning speed :
Wavelength slew rate : about 14000 nm/min
Wavelength scan rate : about 4000 to 0.5 nm/min
7. Spectral bandwidth : 6-steps switching among 0.1/0.2/0.5/1.0/2.0/5.0 nm
8. Resolution : 0.1 nm
9. Stray light : Less than 0.005% with Nal soln at 220 nm & at 340 nm, 370 nm using NaNO2 Soln; 1% at 198 nm with KCI soln
10. Photometric mode : Absorbance (Abs), Transmittance (%), Reflectance (%), Energy(E)
11. Photometric accuracy :
± 0.002 Abs (0 to 0.5 Abs)
± 0.003 Abs (0.5 to 1.0 Abs)
All to be tested with NIST930D/NIST1930 or equivalent filter
12. Photometric range :
Absorbance : -5 to 5 Abs
Transmittance, reflectance : 0 to 100000%
13. Photometric repeatability :
± 0.001 Abs (0 to 0.5 Abs)
± 0.001 Abs (0.5 to 1.0 Abs)
± 0.003 Abs (1.0 to 2.0 Abs)
14. Noise Level: 0.0003 Abs RMS (500nm)
15. Baseline stability : Within 0.0002 Abs/hr(700nm), 1 hours after light source is switched on
16. Baseline flatness : ± 0.0003 Abs (200 ̴ 860 nm), 1 hours after light source is switched on
17. Light source : 50W Halogen lamp, Deuterium lamp, Light source auto position built in
18. Monochromator : Single Monochromator : High performance, Lo-Ray-Ligh blazed holographic grating in aberration corrected Czerny Turner mounting
19. Detector : Photomultiplier tube (R-928)
20. Software Features :
Software Features : fully multitasking; Windows 7 professional/Vista Business(32 Bit) operation for stable speration; QA/AC functionality; GLP/GMP support functions; flexible report generator, customized formats, editing of templates, multi-page print-out support; detailed on-line help with SPECTRUM MODE, KINETICS MODE, QUANTITATION MODE etc. Normalization, point pick, peak/valley detection, area calculation, Transformation: 1St -4th derivatives, smoothing, reciprocal, square root, natural log, logarithm power, Abs to %T conversion and exponential, Kubelka-Munk Conversion. Ensemble averaging, enterpolation, data set and constants arithmetic (between spectra, between spectra and constants).
An integrating sphere is built-in to permit accurate measurement of solid samples. The sample space around the integrating sphere is ample enough to allow measurement of very large samples.
With the integrating sphere shift function, the range of applications is expanded. V stage built ti. The sample position can be adjusted vertically and laterally.
Powdered sample holder capacity of 0.16ml, 3 included.
GENTRONIX BIOSYSTEMS
stage.html
html • 0.03 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.03 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .