GEMC-511687789238858
Awarded to TAYALTECH
₹53,998
Products
| Product | Brand / Model | HSN | Quantity | Unit Price | Amount |
|---|---|---|---|---|---|
| Breath Analyzer as per RDSO Spec dated 15.03. 2018 | Make TAYALTECH Model : Alcokhoj 2018 | - | 2 NOS | ₹26,999 | 53,998 |
Loading…
| # | Company | Offered Item | Amount | Rank | Status |
|---|---|---|---|---|---|
| 1 | L1₹53,998Qualified Make: Tayaltech
Model: Make TAYALTECH Model : Alcokhoj 2018 C 39 TAYAL TECHNOLOGY HUB VYAPAR VIHAR VYAPAR VIHAR BILASPUR CHHATTISGARH 495001 | BILASPUR | CHHATTISGARH | 495001 | Make: Tayaltech
Model: Make TAYALTECH Model : Alcokhoj 2018 | ₹53,998 | L1 | Qualified |
| 2 | Disqualified Make: Alcovisior
Model: alcovisor - jupiter A 1 130 GROUND FLOOR POLE NO 32 GYAN DEEP SCHOOL PALAMNAV VIJAY ENCLAVE SOUTH WEST DELHI DELHI 110045 | SOUTH WEST DELHI | DELHI | 110045 | Make: Alcovisior
Model: alcovisor - jupiter | - | - | Disqualified |
Tender Value
Refer Docs
EMD Value
Exempted
Closing Date
5 Nov 2018, 11:00 amClosed
breath alcohol analyzers Experience with Gov. Required 2 Year (s) MSME Exemption For Years Of Experience And Turnover Yes Breath Alcohol Analyzers ( 2 ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Allowed Values TECHNOLOGY PERFORMANCE Sensor Technology ElectroChemical - Fuel Cell (alcohol specific) ElectroChemical - Fuel Cell (alcohol specific) Sampling Method Deep Lung Exhalation Method * Sampling Error Detection Alarm on wrong sampling * Sampling modes Both Active and Passive * Measuring Range [mg/100 ml BAC (Blood Alcohol Concentration)] 0 to 500 0 to 400
0 to 500
0 to 600 Accuracy +/- 5 % or better * Drift of the breath analyzer in 6 months period (mg/100 ml BAC) 1 Any applicable numeric value 1/5 Easily replaceable mouthpiece with Sputum Trap yes * Sample Taking Duration 2 * Analytical Response Time (At '0' level of Alcohol) 3 * Analytical Response Time for positive result 15 * Recovery Time at (1/2) mg/ltr 0.5 * Warm up Time (after putting on) 2 * Time required for re- calibration 12 * Barometric Pressure Range (hpa or millibar) 600 * Shelf Life of Fuel Cell 3 * RoHS Compliant true * The Equipment meet RFI requirement as per EN:5008-1/1992(emissi on) and EN 50082-1:19 97(immunity) or equivalent requirements true * PHYSICAL Weight 150 * Size ( mm x mm x mm ) - * BATTERY OF ANALYZER Power Source Rechargeable Battery and also work through automobile 12 V DC Power. * Chemistry of Rechargeable Battery Li-ion * Battery Voltage 6 * Battery Backup Time (continuous duty) 30 * Number of Spare Rechargeable Batteries supplied with the instrument 0 * Low Battery indication yes * DISPLAY/INDICATOR STORAGE/INTERFACE Display Type TFT LCD * Display Size 2.4 * 2/5 Availability of audible beeper for over range
low battery etc yes * Record Entry Through Alphanumeric KeyBoard Through Integrated/Touch Screen
Through Alphanumeric KeyBoard Minimum Record Fields Full Information as per the user's requirement such as Driver Name
Vehicle number
inspector name
test report with time
date
geolocation
instrument id
etc * Record Memory (Minimum) 10000 4000
10000
20000
30000
50000 Record Data Downloadable to PC yes * Data Interface Bluetooth * GPS false * If yes
Capable to display the geolocation of test inspection location NA * Camera false true
false If yes
Capable to capture the clear image of test target NA * PRINTER Printer With With
Without Type of Printer Internal Thermal Printer Portable Dot Matrix Printer with Wireless Bluetooth Connectivity
Portable Inkjet Printer with Wireless Bluetooth Connectivity
Portable Laserjet Printer with Wireless Bluetooth Connectivity
Internal Thermal Printer
NA Print Resolution (dpi) 300 * Printer consists of Inbuilt Rechargeable Battery
Paper for offence printouts
Charger yes * Minimum number of printouts that can be 50 * 3/5 taken with the paper available with the Printer "Minimum number of printouts that can be taken with single charge of the printer battery (Hint:- Minimum 60 prints
if 'With Printer' May indicate '0' if not applicable)" 100 * Bis Registration number for Printer under CRS of MeitY - * WARRANTY Warranty 2 1
2
3
4
5 (or higher) STATUTORY REQUIREMENTS The equipment can be used as an evidential instrument for Traffic Law Enforcement
The instrument is hand held
portable and easy to operate true * If Yes
the equipment meets the requirements of the applicable latest Central Motor Vehicle Rules and all Govt notifications yes * Immunity to RF interface
meets requirement as per EN: 5008-/1992(emission) and EN:50082-1:1997 (immunity) or equivalent yes * OPERATING CONDITIONS Minimum Operating Temperature -10 * Maximum Operating Temperature 55 * Relative Humidity Non Condensing (%) 95 * ENVIRONMENTAL CERTIFICATION Conformity to Dry Heat test at 50 degree C for 16 hours as per IS :9000 latest true * Conformity to Damp heat at 40 degree C - 95% RH (cyclic test): two cycles of 16+8 true * 4/5 hours as per IS: 9000 latest Conformity to Cold test at - 5/-10/- 15/-20 degree C (as applicable) for 4 hours as per IS: 9000 latest true * Availability of Test Reports from Central Govt/NABL approved ILAC accredited Lab to prove conformity to the specification true * Test Report Number and Date - * Name of the Lab and Address - * * Bold specifications are the golden parameters. Consignees and Quantity S.No. Consignee Address Quantity Delivery Days 1 Pradip Kumar 854101
DMM/KIR
Office of Divisonal Railway Manager
Katihar Division
2 15
780977
GEM/2018/B/109567
GeM Contract
breath alcohol analyzers
GeM Contract
15 days
Bihar; Katihar
PRODUCT
Awarded to TAYALTECH
₹53,998
| Product | Brand / Model | HSN | Quantity | Unit Price | Amount |
|---|---|---|---|---|---|
| Breath Analyzer as per RDSO Spec dated 15.03. 2018 | Make TAYALTECH Model : Alcokhoj 2018 | - | 2 NOS | ₹26,999 | 53,998 |
| Item | Delivery Location | Qty | Delivery |
|---|---|---|---|
| breath alcohol analyzers Experience with Gov. Required 2 Year (s) MSME Exemption For Years Of Experience And Turnover Yes Breath Alcohol Analyzers ( 2 ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Allowed Values TECHNOLOGY PERFORMANCE Sensor Technology ElectroChemical - Fuel Cell (alcohol specific) ElectroChemical - Fuel Cell (alcohol specific) Sampling Method Deep Lung Exhalation Method * Sampling Error Detection Alarm on wrong sampling * Sampling modes Both Active and Passive * Measuring Range [mg/100 ml BAC (Blood Alcohol Concentration)] 0 to 500 0 to 400 | - | - | - |
| 0 to 500 | - | - | - |
| 0 to 600 Accuracy +/- 5 % or better * Drift of the breath analyzer in 6 months period (mg/100 ml BAC) 1 Any applicable numeric value 1/5 Easily replaceable mouthpiece with Sputum Trap yes * Sample Taking Duration 2 * Analytical Response Time (At '0' level of Alcohol) 3 * Analytical Response Time for positive result 15 * Recovery Time at (1/2) mg/ltr 0.5 * Warm up Time (after putting on) 2 * Time required for re- calibration 12 * Barometric Pressure Range (hpa or millibar) 600 * Shelf Life of Fuel Cell 3 * RoHS Compliant true * The Equipment meet RFI requirement as per EN:5008-1/1992(emissi on) and EN 50082-1:19 97(immunity) or equivalent requirements true * PHYSICAL Weight 150 * Size ( mm x mm x mm ) - * BATTERY OF ANALYZER Power Source Rechargeable Battery and also work through automobile 12 V DC Power. * Chemistry of Rechargeable Battery Li-ion * Battery Voltage 6 * Battery Backup Time (continuous duty) 30 * Number of Spare Rechargeable Batteries supplied with the instrument 0 * Low Battery indication yes * DISPLAY/INDICATOR STORAGE/INTERFACE Display Type TFT LCD * Display Size 2.4 * 2/5 Availability of audible beeper for over range | - | - | - |
| low battery etc yes * Record Entry Through Alphanumeric KeyBoard Through Integrated/Touch Screen | - | - | - |
| Through Alphanumeric KeyBoard Minimum Record Fields Full Information as per the user's requirement such as Driver Name | - | - | - |
| Vehicle number | - | - | - |
| inspector name | - | - | - |
| test report with time | - | - | - |
| date | - | - | - |
| geolocation | - | - | - |
| instrument id | - | - | - |
| etc * Record Memory (Minimum) 10000 4000 | - | - | - |
| 10000 | - | - | - |
| 20000 | - | - | - |
| 30000 | - | - | - |
| 50000 Record Data Downloadable to PC yes * Data Interface Bluetooth * GPS false * If yes | - | - | - |
| Capable to display the geolocation of test inspection location NA * Camera false true | - | - | - |
| false If yes | - | - | - |
| Capable to capture the clear image of test target NA * PRINTER Printer With With | - | - | - |
| Without Type of Printer Internal Thermal Printer Portable Dot Matrix Printer with Wireless Bluetooth Connectivity | - | - | - |
| Portable Inkjet Printer with Wireless Bluetooth Connectivity | - | - | - |
| Portable Laserjet Printer with Wireless Bluetooth Connectivity | - | - | - |
| Internal Thermal Printer | - | - | - |
| NA Print Resolution (dpi) 300 * Printer consists of Inbuilt Rechargeable Battery | - | - | - |
| Paper for offence printouts | - | - | - |
| Charger yes * Minimum number of printouts that can be 50 * 3/5 taken with the paper available with the Printer "Minimum number of printouts that can be taken with single charge of the printer battery (Hint:- Minimum 60 prints | - | - | - |
| if 'With Printer' May indicate '0' if not applicable)" 100 * Bis Registration number for Printer under CRS of MeitY - * WARRANTY Warranty 2 1 | - | - | - |
| 2 | - | - | - |
| 3 | - | - | - |
| 4 | - | - | - |
| 5 (or higher) STATUTORY REQUIREMENTS The equipment can be used as an evidential instrument for Traffic Law Enforcement | - | - | - |
| The instrument is hand held | - | - | - |
| portable and easy to operate true * If Yes | - | - | - |
| the equipment meets the requirements of the applicable latest Central Motor Vehicle Rules and all Govt notifications yes * Immunity to RF interface | - | - | - |
| meets requirement as per EN: 5008-/1992(emission) and EN:50082-1:1997 (immunity) or equivalent yes * OPERATING CONDITIONS Minimum Operating Temperature -10 * Maximum Operating Temperature 55 * Relative Humidity Non Condensing (%) 95 * ENVIRONMENTAL CERTIFICATION Conformity to Dry Heat test at 50 degree C for 16 hours as per IS :9000 latest true * Conformity to Damp heat at 40 degree C - 95% RH (cyclic test): two cycles of 16+8 true * 4/5 hours as per IS: 9000 latest Conformity to Cold test at - 5/-10/- 15/-20 degree C (as applicable) for 4 hours as per IS: 9000 latest true * Availability of Test Reports from Central Govt/NABL approved ILAC accredited Lab to prove conformity to the specification true * Test Report Number and Date - * Name of the Lab and Address - * * Bold specifications are the golden parameters. Consignees and Quantity S.No. Consignee Address Quantity Delivery Days 1 Pradip Kumar 854101 | - | - | - |
| DMM/KIR | - | - | - |
| Office of Divisonal Railway Manager | - | - | - |
| Katihar Division | - | - | - |
| 2 15 | - | - | - |
Exempted
27 Dec 2018
25 Oct 2018
contract_GEMC-511687789238858.pdf
GEM_CONTRACT • 0.02 MB
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
bid_780977.pdf
GEM_BID
termsCondition.pdf
OTHER
2657c7ca70287299b5181580fdb5a7ab.pdf
OTHER
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .