Gmers Medical College Attached Hospital,Porbandar
Supply of Blood Bank Equipment
334275
01/2026-27
Open
Medical Equipments/Waste
Supply
Porbandar
2 documents required · 2 mandatory
₹5,000
Medical Superintendent G.M.E.R.S General Hospital,Porbandar
₹3 L
14 Aug 2026
14 Aug 2026
14 Aug 2026
31 Aug 2026
14 Aug 2026
GMERS Medical College Attached Hospital, Porbandar Near Bus Stand, Porbandar - 360575 Mob - 9979521946
BLOOD BANK EQUIPMENTS Specifications.
Sr. No Name of Equipment Nos Specifications.
1 Donor Couches 2 Specifications: • Should comply IEC safety /EMI/EMC standards, ISO 9001:2015, ISO 13485:2016 certified and CE marked as Class 1. • Should have a maximum lifting capacity of up to 135 Kg. • Should have Lockable castors for easy mobility. • Should have an adjustable head rest. • Must have variable positioning for either arm with comfortably wide armrests • which are swivel able and height adjustable. • Must have interfacing facility with blood collection monitor. • Should have BCM cum accessories stand with telescopic IV pole as accessory (Optional). • Should have comfortable chair type with soft padding for cushioning and rexin cover. The soft upholstery should have a thickness 2.5 inch. • Must be adjust stable to any position between seating / donation position and vasovagal attack position. • Must have a safety switch to quickly adjust to vasovagal position. • Must have a noise free mode of operation. • The motor movement should be initiated by the rocker switch. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
2 Blood Collection Monitor 6 Specifications: • Should comply IEC safety / ISO 9001:2015, ISO 13485:2016 certified and CE marked as Class Im (Measuring). • Must have a weighing range of 50-500 ml. • Must have Automatic tare to zero for the bag weight. • The default volume must be adjustable. • Should clamp Automatically at termination of preset volume collection. • Should have option for manual clamp for usage in case of emergency. • Must have a weighing accuracy +/-2%. • The weight of D601- 4.5kg and Power Adaptor- 550g. • Must have easily detachable tray for cleaning and disinfection which are compatible with all bag systems. • Must have 20 x 4-line character Backlit LCD display of programmed volume, collected volume, flow rate, main battery. • Alarm /Indications for high/low flow indication, end of collection, power failure and battery low. • Must have indication of time taken for collection. • Must have indication with alarm if blood flow rate is high or low. • Must have continuous display of collected volume, flow and time during collection. • Trays Should oscillate continuously at 12+/- 2 rpm during collection. • Should operate on mains as well as rechargeable battery. Lithium Ion Battery with 12 hours battery backup / 90 to 100 collections assuming an average 8-minute donation time. • Pause facility to pause during collection. • 350/450 ml easy selection switch. • Should have wight check mode. • Should have external dimension 190 mm × 500 mm × 185 mm (+/- 5mm Tolerance). • Should have following alarm & indications: • Audible/visible: Flow rate below 20 mL/min or above 180 mL/min • Audible/visible: Power failure • Audible/visible: Battery low • Audible/visible: Automatic clamping when flow rate is sustained at less than 20 mL/min for more than 2 minutes • Audible/visible: End of collection. • working environment: • Temperature: 10 °C to 40 °C • Relative humidity: 30% to 75% • Atmospheric pressure: 700 hPa to 1,060 hPa • Should have wight check mode. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
3 BB Refrigerators (2 to 8 ) 300 Itrs 2 Specifications: • Should have a chamber temperature range of 2°C to 6°C. • Should comply IEC safety /EMI/EMC standards, ISO 9001:2015 ,ISO 13485 :2016 certified. • Should have a total storage capacity of 160 nos. of 450ml blood bags or 200 nos. of 350ml bags. Storage capacity per tray: 1. 32 no of bags 450ml. or 2. 40 no of bags 350ml. • Should have effective storage volume of 300 Litre. • Should have insulation thickness of 80mm PU. • Should have a corrosion free Interior of SS 304, Thickness :0.8mm grade which helps in easy cleaning. • Should have a 18 swg CRCA/HDGI Sheet -1mm, powder coated. • Should have built in microcontroller based temperature recorder and controller unit (TRCU) positioned at eye level for better visibility and monitoring. • Must have six inch, 7-day, ink-less pressure sensitive circular chart recorder and accuracy should be +/- 1 degree Celsius. • Should have triple pane glass door fitted in PUF insulated door which insulates the refrigerator from atmospheric temperature and better visibility of blood bags with key lock. • Must have an internal evaporator fan for uniform temperature maintenance inside the chamber. It should switch off automatically whenever the door is opened. • Should have perforated sliding Stainless steel tray which allows bags to be placed upright with sufficient airspace to reduce " Sardine effect". • Should have 5 stainless steel slide out trays. • Should have separate trays fitted with acrylic panel ensures minimal loss of cooling. • Should have hotline around the mouth of cabinet to prevent moisture condensation. • Should have built in internal stabilizer of 1 KVA which ensures steady supply voltage for the whole blood storage cabinet and prevents voltage fluctuations. • Should have a factory calibrated encapsulated digital sensor dipped in liquid medium to measure temperature at an accuracy of ±0.5ºC. • Should have Provided with LED bulb for uniform lighting and better visibility inside the cabinet. • Should include castor wheels as a standard feature *2 lockable casters in the front *2 non-lockable casters in the rear • Should have refrigeration system “On” indicator provided as a standard feature. • Should have audible and visual high and low temperature alarms as a standard feature. • Should have an alarm acknowledgement button. • Should have a power fail alarm as a standard feature. • Must incorporate a inbuilt battery back-up of (12V 2.3Ah), 2hr backup with Rechargeable Sealed Maintenance free Lead Acid Battery to ensure continuous operation of the TRCU during power failure. • Should have display of 4*7 Segment LED with 0.5ºC display resolution. • Must utilize non-CFC, commercially available refrigerant R134a (CFC free). • Must keep the refrigerator free of frost without elevating the chamber temperature. • Must have a cabinet with powder coated finish to guard against rust and corrosion. • There must be eye level indications of power/Line in from inbuilt stabilizer. • Chart change and Chart set options in TRCU must be available. • Must have MCB protection in the input power line. • Should have negative temperature protection as standard feature. • Should have data logging feature with removable data logger. • Should have accuracy of chart with +/- 1°C. • Should be compatible for working environment : Temperature 22oC to 33oC Relative humidity 10 to 90% Atmospheric pressure 700 to 1060hpa. • Should comply with Storing and transportation: Temperature -20oC to 40oC Relative humidity 20to 90% Atmospheric pressure 600 to 1060hpa. • Should have user interface with membrane keypad. • Should have temperature sensing method with encapsulated digital sensor dipped in 0.25% glycerine solution and kept in a plastic bottle. • Should have following alarm & indications: *Line - Visible *Power- Visible *Compressor on- Visible * Temperature High/Low- Audible/Visible * Door Open- Audible/Visible * System Ok - Visible *Battery Low - Visible * Power Failure - Audible/visible *Sensor failure- Audible/Visible *Chart Change- Visible • Should have following dimensions: Internal dimension:510 × 490 × 1220 mm External Dimension:670 × 676 × 1785 mm with casters. (+/-5mm Tolerance) • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
4 BB Refrigerators (2 to 8 ) 600 Itrs 1 Specifications: • Should have a chamber temperature range of 2°C to 6°C. • Should comply IEC safety/EMI/EMC standards, ISO 9001:2015, ISO 13485:2016 certified. • Should have a total storage capacity of 360 nos. of 450 ml blood bags or 510 nos. of 350 ml bags. Storage capacity per tray: 72 no. of bags 450ml. or 102 no. of bags 350ml. • Should have effective storage volume of 600 Ltr. • Should have insulation thickness of 80mm PU. • Should have a corrosion free Interior of SS 304, Thickness: 0.8mm grade which helps in easy cleaning. • Should have a 18 swg CRCA/HDGI Sheet -1mm, powder coated. • Should have built in microcontroller-based temperature recorder and controller unit (TRCU) positioned at eye level for better visibility and monitoring. • Must have six inch, 7-day, ink-less pressure sensitive circular chart recorder and accuracy should be +/- 1 degree Celsius. • Should have triple pane glass door fitted in PUF insulated door which insulates the refrigerator from atmospheric temperature and better visibility of blood bags with key lock. • Must have an internal evaporator fan for uniform temperature maintenance inside the chamber. It should switch off automatically whenever the door is opened. • Should have perforated sliding Stainless steel tray which allows bags to be placed upright with sufficient airspace to reduce " Sardine effect". • Should have 5 stainless steels slide out trays. • Should have separate trays fitted with acrylic panel ensures minimal loss of cooling. • Should have hotline around the mouth of cabinet to prevent moisture condensation. • Should have built in internal stabilizer of 1 KVA which ensures steady supply voltage for the whole blood storage cabinet and prevents voltage fluctuations. • Should have a factory calibrated encapsulated digital sensor dipped in liquid medium to measure temperature at an accuracy of ±0.5ºC. • Should have Provided with LED bulb for uniform lighting and better visibility inside the cabinet. • Should include castor wheels as a standard feature * 2 lockable casters in the front * 2 non-lockable casters in the rear • Should have refrigeration system “On” indicator provided as a standard feature. • Should have audible and visual high and low temperature alarms as a standard feature. • Should have an alarm acknowledgement button. • Should have a power fail alarm as a standard feature. • Must incorporate an inbuilt battery backup of (12V 2.3Ah), 2hr backup with Rechargeable Sealed Maintenance free Lead Acid Battery to ensure continuous operation of the TRCU during power failure. • Should have display of 4*7 Segment LED with 0.5ºC display resolution. • Must utilize non-CFC, commercially available refrigerant R134a (CFC free). • Must keep the refrigerator free of frost without elevating the chamber temperature. • Must have a cabinet with powder coated finish to guard against rust and corrosion. • There must be eye level indications of power/Line in from inbuilt stabilizer. • Chart change and Chart set options in TRCU must be available. • Must have MCB protection in the input power line. • Should have negative temperature protection as standard feature. • Should have data logging feature with removable data logger. • Should have accuracy of chart with +/- 1°C. • Should be compatible for working environment: Temperature 22oC to 33oC Relative humidity 10 to 90% Atmospheric pressure 700 to 1060hpa. • Should comply with Storing and transportation: Temperature -20oC to 40oC Relative humidity 20 to 90% Atmospheric pressure 600 to 1060hpa • Should have user interface with membrane keypad. • Should have temperature sensing method with encapsulated digital sensor dipped in 0.25% glycerin solution and kept in a plastic bottle. • Should have following alarm & indications: * Line - Visible * Power - Visible * Compressor on – Visible * Temperature High/Low - Audible/Visible * Door Open - Audible/Visible * System Ok – Visible * Battery Low - Visible * Power Failure - Audible/visible * Sensor failure - Audible/Visible * Chart Change – Visible • Should have following dimensions: Internal dimension: 710 X 770 X 1210 mm External Dimension: 870 × 956 × 1865 mm with casters (+/-5mm Tolerance). • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
5 HAND Stripper 1 Specifications: • A hand sealer for use with blood tubing sets and aluminum sealing clips to provide a permanent seal of the tubing. • The sealer can also be used for stripping and cutting the tubing. • The company should have more than 100 installations in Gujarat. • Spring power Handle for uniform pressure. • Net weight- 140g. Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
6 Portable top tube Sealer 1 Specifications: • Should be a Heavy duty radio frequency sealer. • Should have Automatic detection of the tube by pressing of a lever which activates sensor. • The equipment must not have warm up time before sealing. • Should have separable rupture line to separate tube ends after sealing. • Should have Switch mode power supply of uniform sealing irrespective of power supply variations. • Standard blood bag tubes with tube diameter up to 6mm. • Should have adjustable sealing time between 1-3 seconds. • RF signal should have applied only after the tube is fully squeezed. • Should have removable splash guard to protect phlebotomist from all kinds of blood splashes. • Should have Indications for Cover open, Ready, Seal and Power. • The equipment must be protected against electric shock class I. • Must be operational on 185W; standby 20W. • Should have Easy to clean electrodes which are easily accessible & protected by cover. • Should have CE as lab device, ISO 9001:2015 ,ISO 13485:2016, EURoHS. • Should have a SMPS to withstand input voltage fluctuations. • Should be easy to handle and must weight approximately 5.25 Kg. • Should have dimension 155 x 330 x 160mm (+/-5mm Tolerance) • Should comply with working environment: Temperature: 5 °C to 40 °C Relative Humidity: 30 to 75% Atm Pressure: 700 to 1060hPa. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
7 Plasma Extractor 1 Specifications: • Should comply with safety requirements ,ISO 9001:2015 ,ISO 13485:2016 certified. CE marked under Machinery directive. • Should have a body material with spring loaded acrylic sheet. • The handle should be with ball which helps easy handling for expressing the plasma. • Should be transparent plate which allows easy visibility of the level of expressing the blood product, with bag pressing mechanism. • Should have spring force :3.25+/-0.25Kgf. • It should weigh not more than 4.25kg. • Should be user friendly. • Mode of operation should be Manual mode. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
8 Platelet agitator 1 Specifications: • Should be able to store minimum of 24/48 platelet bags. • Should have 4 trays and be able to store minimum 24 platelet bags & 8 trays to store 48platelet bags. • Should be flat bed agitator. • Should have 65 +/-7 cycles motions per min. • Should have agitation pause function for 10 sec • Should have LED indication for blown fuse. • Should have audible intermittent beeps for power failure. • Should have SS304 grade sliding stainless steel trays. • Should comply IEC safety /EMI/EMC standards, ISO 9001:2015 ,ISO 13485:2016 certified and CE marked as Class 1. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
9 Platelet incubator 1 Specifications: • Should have power supply of 230+/-10% VAC,50 Hz single phase. • Should have displacement range 36+/- 3mm. • Should have following dimension PA200-440x390x290 PA300-440x390x395 (mm) (+/-5mm Tolerance). • Should have following alarm & indication: Power failure- Audible fuse blown – visible. • Should comply with working environment: Temperature: 10 to 40 C Relative humidity: 30 to 75% Atm Pressure: 700 to 1060hPa. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
10 Deep Freezer -40 1 Specifications: • Should have a Chamber temperature range between -20 º C to -40º C • Should have an input voltage of 230+/-10%V/50Hz, Single phase Ac • Should have an inner chamber storage volume of 300L with a total storage capacity of 320 nos. of 250ml plasma bags. • Should comply IEC safety /EMI/EMC standards, ISO 9001:2015 ,ISO 13485 :2016 certified. • Should have built in microcontroller based temperature recorder and controller unit (TRCU). • Should include caster wheels as a standard feature: 2 lockable casters in the front 2 non-lockable casters in the rear • Must have six inch, 7-day, ink-less pressure sensitive circular chart recorder and accuracy should be +/- 1 degree Celsius. • Must have a Heavy gauge , cold rolled annealed exterior with powder coating-1.5mm thick HDGI sheet – Powder coated Interior- 1 mm thick Stainless-Steel sheet of 304 Grade. • Should have 3 adjustable stainless steel trays to place the bags.(4 compartment) • Should have stainless steel interior doors with magnetic latches that ensure secure storage and less cold air loss during opening and closing of doors. • Refrigerating fluid & PUF insulation should be CFC free. • The door insulation should have a PUF thickness of 125 mm with rubber gasket sealing to ensure more temperature holding time and reduced power consumption • Should have hotline around the mouth of cabinet to prevent moisture condensation • Should have 12 hours Battery back up for TRCU that keeps the temperature monitoring uninterrupted. • Should have user friendly alarms and indications to display all possible abnormalities. • Should have display of 4*7 Segment LED . • Should have audible and visual high and low temperature alarms as a standard feature. • Should have a chart recorder with chart duration of 7 days and chart resolution of 3º C • Must have a High capacity air cooled condenser • Should have controlled voltage stabilizer of 4 KVA should be excluded. • Should have hermetically sealed compressor which automatically regulate according to the freezer. • Display resolution of controller should be 0.1C. • Chart Range should be -100C to +50C • working environment should be Temperature: 22ºC to 30ºC Relative Humidity: 10 to 90% Atm Pressure: 700 to 1060hPa • Storage & Transportation: Temperature: -20ºC to 40ºC Relative Humidity: 20 to 90% Atm Pressure: 600 to 1060hPa • It should have temperature sensor RTD sensor-PT100 • Should have dimension (mm) External:805 x 895 x 1940 Internal:550 x 500 x 1100 (+/-5mm Tolerance) • Should have following alarm & Indications: Compressor on - Visible High/low temperature- Audible/visible Door open- Audible/visible Battery low- Visible Power on- Visible Power fail- Visible • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
11 Deep Freezer -80 1 • Should have a Chamber temperature range between -50 ºC to -80ºC( at 22 to 30ºC ambient temperature). • Should have an input voltage of 230+/- 10% VAC 50Hz, Single phase Ac. • Should have an inner chamber storage volume of 300L with a total storage capacity of 320 nos of 250ml plasma bags. • Should comply IEC safety /EMI/EMC standards, ISO 9001:2015 ,ISO 13485 :2016 certified. • Should have built in microcontroller based temperature recorder and controller unit (TRCU) positioned at eye level. • Should be heavy duty refrigeration system, maintenance free, with hermetically sealed refrigeration compressors reliable refrigeration with minimum noise & vibration. • Should include caster wheels as a standard feature 2 lockable casters in the front 2 non-lockable casters in the rear • Must have six inch, 7-day, ink-less pressure sensitive circular chart recorder • Must have a Heavy gauge, cold rolled annealed exterior with powder coating Interior:1.2mm thick Stainless-Steel sheet of 304 Grade Exterior:1.5mm thick HDGI sheet – Powder coated • Should have adjustable stainless steel trays to place the bags. • Should have stainless steel interior doors with magnetic latches that ensure secure storage and less cold air loss during opening and closing of doors. • Refrigerating fluid & PUF insulation should be CFC free. • Should have an ergonomically designed door handle with self-pushing mechanism. • The door insulation should have a PUF thickness of 125 mm with rubber gasket sealing to ensure more temperature holding time and reduced power consumption. • Should have hotline around the mouth of cabinet to prevent moisture condensation. • Should have triple point rubber gasket sealing which reduces impact of ambient temperature and humidity in inner chamber. • Should have 12 hours Battery backup for TRCU that keeps the temperature monitoring uninterrupted. • Should have user friendly alarms and indications to display all possible abnormalities • Should have display of 4*7 Segment LED with 0.1ºC display resolution. • Should have audible and visual high and low temperature alarms as a standard feature. • Should have a chart recorder with chart duration of 7 days and chart resolution of 3º C. • Must have a high capacity air cooled condenser. • Servo controlled voltage stabilizer of 5 KVA should be excluded. • Should have two hermetically sealed compressor which automatically regulate according to the freezer. • Should have pressure sensitive mechanism to protect the compressor in long run. • Should have temperature sensing method as RTD sensor placed in a stainless steel bracket fitted at the left side of the second compartment from the bottom side. • Should have chart range -100 to +50 C. • Display resolution of controller should be 0.1C. • should comply with Working environment : Temperature: 22ºC to 30ºC Relative Humidity : 10 to 90% Atm Pressure : 700 to 1060hPa • Should comply with Storing environment: Temperature: -20ºC to 40ºC Relative Humidity: 20 to 90% Atm Pressure: 600 to 1060hPa. • It should have following dimensions: External :805 x 825 x 1940 (mm) with castor wheel. Internal:550 x 500 x 1100(mm) (+/-5mm Tolerance). • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
12 Cryoprecipitate bath 1 Specifications: • Should be able to thaw plasma bags at preset temperature between 3°C to 56 °C. • Should have removable Stainless steel trays with individual compartments for holding 12 plasma bags. • Should have a uniform and optimum thawing mechanism – pumping mechanism by high capacity pump. • Total weight should not exceed 80 kg. • Should have castor wheels for easy transportation. • Should have a display resolution of 0.1°C. • Should have a 2 line 4x7 LED display. • Should have a microprocessor based digital controller to control temperature. • Should have ISO 9001:2015 ,ISO 13485:2016 • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
13 Thawing Bath 1 Specifications: • Should be able to thaw plasma bags at preset temperature between 37 °C to 56 °C. • Should have removable Stainless-steel trays with individual compartments for holding 12 plasma bags. • The inner tank should be made of stainless steel. • Should have a uniform and optimum thawing mechanism – pumping mechanism by high capacity pump. • Total weight should not exceed 35kg. • Should have a display resolution of 0.1°C. • Should have a 2 line 4x7 LED display. • Should have a microprocessor based digital controller to control temperature. • Should have a external dimension , not greater than 701mm x 390mm x 372mm ( L x W x H ) (+/-5mm Tolerance). • Should have ISO 9001:2015 ,ISO 13485:2016 • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
14 Compo Scale 1 Specifications: • Should comply with safety requirements of EN 61010-1, ISO 9001:2015 ,ISO 13485:2016 certified, CE marked as Lab device. • Should have user friendly system with alarms and indications. • Should have a 16X1 Line LCD Display. • Must have a weighing range up to 2.5 Kg. • Accurate measurement of volume and gravity of various blood components must be possible. • Should have provision for Zero set. • Must have a compare mode that allows balancing of buckets used in centrifuge. • Should have a keyboard with controls such as weight/volume conversion, volume setting, component selection, zero set, balancing of bags. • Should have a load cell transducer. • Must have a resolution of 1gm/1ml. • Should have weight of approximately 2.5 Kg so that it is easily portable. • Should have tolerance range Tolerance : 0-500g ±2g/ml 501-2500g ±5g/ml • Should have following dimension 185x240x50 mm(approx.) (+/-5mm Tolerance). • Comply with working environment: Temperature: 5°C to 40°C Relative humidity: 10% to 80% Atm. Pressure: 700hPa to 1060hPa • should have capability to interface with TOMEs. • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
15 Sterile Connecting Devic 1 Specifications: Tubing Material: Polyvinyl Chloride (PVC) Tubing Sizes: O.D: 3.9 – 4.5mm I.D.: 2.9 – 3.1mm Wafer: TSCD wafer / Product code SC*W017 Operating conditions Power Supply: AC 100V to 240V 10%, 50/60 Hz External Dimensions: 224(W) x 177 (H) x 342 (D) mm Weight: Approx. 6.5kg (14.33 Ibs.) Operating Conditions: The welder should be allowed to equilibrate at these conditions before use. Ambient Temperature: 10°C to 40°C (50° to 104°F) Relative humidity: 10% to 80%, non-condensing Maximum attitude: 2,000m (6,562 ft) Storage Conditions: Ambient Temperature: - 10°C to 50°C (14° to 140°F) Relative humidity: 10% to 95% non-condensing • Required Installation List in Gujarat. • More then 9-year-old machine’s satisfactory service and performance certificate (Minimum 10 Certificate) is required from blood bank or pharmaceutical. • Local engineer is must in Gujarat.
16 Refrigerated Centrifuge (12 bags) 1 Specifications: • Wind Shielded Swinging bucket rotor with Metal buckets having middle partition to hold 2 blood bag sets, Max total Capacity- 6 X 2000ml • Rotor must be suitable for 12 Quintuple blood bag systems per run: each having 450 mL capacity with filters, with SAGM bag and empty satellite bags. • Removable adapters with smooth edges to spin quadruple blood bag system 450 ml with soft filters.3 pairs must be quoted. • These adaptors must have provision to hold balanced weights in ventilated & separated compartments at the sides to avoid contact with blood bags. • Must be supplied with soft rubber balancing plates to compensate for big weight difference, one set of 2 plates each of 35g and 65 g. • Software must automatically Identify rotor & prevent over speeding with guided alert on screen. • Max RCF 5000xg or more, Programable in 1xg increments for precise control. Should be able to set run at minimum speed of 300 RPM. • Inbuilt operating software must have User access controls with master passcode & 4 different User Access Levels to ensure GMP compliant operation. • Imbalance tolerance of > 100 gram. Centrifuge must be able to run without any error even if weight difference on opposite buckets is up to 100 gm to help minimize the buckets balancing efforts and time during the peak load/ demand time. • Easy thread pull mechanism for emergency manual lid opening from 2 sides. • Easily accessible Run Log with User Id, Run parameters & blood bag ID option with USB output. • Must have option to preset system cooling by time of the day (example: start at 6 AM/Stop at 7 AM) • System should record Event log & rotor log in inbuilt software. • System should have option to count Run time: At start or at speed. Onboard Timer should have HH:MM & HH:MM:SS options to choose, with 1 sec increment option. • Should have options to Run for defined Time or Hold. • Centrifuge should have inbuilt capability to automatically adjust centrifugation run time to compensate for variations in acceleration and deceleration due to Rotor loading, Instrument Age, voltage fluctuations etc. to ensure consistent and reproducible separation of blood components. • System must have compatibility to upgrade with attached barcode scanner for tracking & traceability & optional data management software. • Programmable memory with tamper proof facility. The centrifuge must be able to store & quick recall of > 100 programs. • Step Runs can be created in single programs with more than 25 step options with varying RPM, Temp, Acceleration & Deceleration values & Display in graph curve format. • Completely smooth Stainless-steel corrosion resistant chamber for Easy cleaning. • Must have dedicated drain port, drain plug, no leak pipe and user removable drain tray for easy cleaning. • Must have complete automation in door movement. The centrifuge door should open and close automatically using simple touch on screen. • The Door must be equipped with Automatic Lid lock for safety. • The windshield lid must also close & open automatically with door. No manual lifting. • Door must be able to sense hand or object blockage while movement to avoid accidents. While movement of door, Sounds & Alert on screen must be there. • Rotor imbalance detection: Automatic shutdown of Run if rotor load difference on opposite buckets is 250 gram or more, with Appropriate error message & Sound alarms. The imbalance should be detected within 1500 RPM speed to avoid any accident. • Temperature Control Range: -200C to +400C , adjustable by 1 °C. • Should incorporate alarms for imbalance detection & door movement. Settings should have options of multiple alarm sounds & user editable sound levels. • Max Run Time- 99 hours 58 min 59 sec (1 second increment). • Warm air outlet from sides / rear of the machine. Average heat dissipation must be less than 2 KW/h. • Compressor must turn immediately off when the door is opened. • The centrifuge must have more than 9 Acceleration profiles and 9 Deceleration profiles. Noise and Vibration free operation at maximum speed. 1.Noise level must be less than 65dB at Max speed. • Power Requirement: 230V, Single Phase, 50Hz, 32 Amps. • Bidder should have minimum 10 Installations within state of Gujarat for 12/16 Bags Blood Processing Centrifuge in last 5 years. Performance Certificate or proof of satisfactory installation to be submitted along with technical bid. • Must have CE Class IIa Medical Device certificate for “Blood bank centrifuges for transfusion purposes” from reputed Internationally accredited notified bodies. • Must be certified by UL to meet the international standards of Electrical Safety 1. IEC / EN 61010-2-020 • Must be manufactured at site certified with ISO 13485:2016 certification. • 10 KVA analog stabilizer required for Centrifuge, Price exclusive. • Refrigerated Centrifuge should use Environment friendly, Natural and Green Refrigerants i.e. Ammonia (R-717), Carbon Dioxide (R-744), Hydrocarbons such as propane (R-290) and isobutane (R-600a), as well as synthetic hydrofluorpoolefins (HFOs) like R1234yf and R1234ze.
17 ELISA Reader anti washer (Semi Auto) 1 Product Name: Elisa Reader Photometric Range: 0.000-4.000Abs Resolution:0.001Abs(displayed). 0.0001Abs(calculated) Light Source: Halogen lamp, 8V/50W Wavelengths: 405,450,492,630nm, 4 more filters optional Wavelength Accuracy: + 1nm Band Width: 8nm Calculation Method: ABS, %ABS, cut-off, linear, point to point, log, exponent, power regression Accuracy: + 1% or + 0.005A Reading Speed: 5secs for 96 well plate (Single wavelength) Shaking Plate: Shaking time and speed adjustable Memory: 100 tests, up to 10000 test results Interface: RS-232, PSTN (optional) Display: 6" LCD Input: Touch panel or mouse Output: External printer (optional) Net Weight: 8KG Dimensions LXWXH (mm): 460 x 330 x 200 Power Supply: AC 110V-220V+ 10%, 50-60Hz Product Name: Elisa Washer Plate Mode: 96-or 48- well plate or strip Dispense Precision: CV<2% Residual Volume: < 1ul Washing Volume: 50-3000ul for 8-way manifold 50-2000ul for 12-way manifold Washing Cycle: Maximum 99 cycles Strip Washing: 1 to 12 channels adjustable Soaking Time: 0-24 hr Display: 5" LCD Bottle Capacity: 2 litres each Memory: 50 washing protocols Interface: RS-232 serial port Net Weight: 8KG Dimensions LXWXH (mm): 390 x 340 x 180 AC 220V + 10%, 50-60Hz, 110V on request
18 Gel system (semi auto) for Cross Match 1 Specifications • Semi-automated Immunohematology system should be able to perform centrifugation, Incubation, reading and interpretation of all the tests based on Gel Technology for Cross Matching in coombs and enzyme phase to pick both IgG & IgM Antibodies, Blood grouping, Rh typing, Antibody screening & Identification. • System should comprise of following- Gel Card Centrifuge minimum 24 gel card capacity. Gel Card Incubator Gel Card Reader Work Station • System should be table top model. • System should be based on Gel Colum agglutination technology. • Card Centrifuge should be microprocessor controlled. • Card Centrifuge should be of minimum 24 cards capacity with 85g force. • Card Centrifuge should be compatible with 6 microtubes and 8 microtubes gel cards. • Company should manufacture compatible Gel cards and reagents. • Card Centrifuge should have Electronic door lock with open lid detection. • Card Centrifuge should be preprogrammed for fixed 10 minutes time with audible alarm. • Centrifuge should have LCD type panel for display of RPM and time. • Centrifuge Power supply 110/220 volts/60-50 Hz. • Card incubator should have capacity to incubate 24 gel cards. • Card incubator should have two independent programmable timers with audible alarm providing flexibility to incubate gel cards in two batches. • Incubator should be preprogrammed at 37° C. • Gel Card Reader should be a Semi-automated system for reading and interpretation of results in Gel Cards based on column agglutination technology. • Should be compatible with 6 and 8 microtube gel cards. • Should be capable of reading Strength of reaction and interpretation of results. • Should have system for positive identification of gel card parameter through barcodes. • Should have complete traceability of Test results, Sample and Operator. • System should allow user to modify the reactions of a microtube. • System should remember the unused microtubes of a gel card and allow the user to use them later. • System should store and print test results using the original image of microtubes. • Should have LIS compatibility. • System should be preferably make in India. • Local representative of company should be available for service and application support.
19 Coagulation Analyzer (Semi) 1 Technical Characteristics: Optical Channels: Single / Duel / Four Wavelength: 620 NM Real Time Clock: Yes Display: 4” colour touch screen Dimension: 275 mm (Depth) X 140 mm (Width) X 80 mm (Height) Interface: Printer, Bar-code reader. Test Volume: 75 - 150 μl Global Coagulation Test: PT, APTT, TT, FIB Humidity: 20% - 80% without condensation Operation Temperature: 15 - 30o C Power Supply: 110 - 230V AC 50/60 Hz
20 Cell Counter 1 Specifications: Principle: Impedance method for WBC, RBC, PLT counting Cyanide free reagent for HGB test. Parameter: • 21 Parameters: WBC, Lym#, Mid#, Gran#, Lym%, Mid%, Gran%, RBC, HGB, HCT, • MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC • 3 Histograms: WBC, RBC, PLT Throughput: 70 samples per hour Sample volume: 70 samples per hour Whole blood: 10 μl Pre-diluted: 20 μl Reagent: 2 reagents+1 maintenance cleanser (Diluent, Lyse, Probe cleanser) Calibration: Manual calibration Auto calibration: calibrator, fresh blood QC: L-J, X-B Input: 10.4-inch touch screen Mouse, keyboard (optional) Output: Built-in recorder, external printer (optional) Data storage: >= 40,000 Interface: LAN port, HL7 protocol, support bilateral LIS 4 USB ports for external printer, external barcode scanner, keyboard, etc. Power requirement: 100V-240V~, 50/60Hz, < 200W Working environment: 15°C-35°C Weight: 18Kg Dimension(mm): 300(W)X400(D)X410(H)
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
RAW_HTML
Blood Bank Equipment final - 14-08-26.pdf
Blood Bank Equipment Specifications- 14-08-26.pdf
Blood Bank Equipment Specification.xlsx
XLSX
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at [email protected].