| # | Company | Amount | Rank | Status |
|---|---|---|---|---|
| 1 | 1₹8.5 LAccepted-AOC 66 4743 M G ROAD NORTH END ERNAKULAM ERNAKUALM KERALA 682035 | ERNAKULAM | KERALA | 682035 | 1 | Accepted-AOC L1 | |
| 2 | Rejected-Technical DOOR NO 19 OPP ESI DISPENSARY MAR GREGORIOUS ROAD ALUVA ERNAKULAM 683101 | ALUVA | ERNAKULAM | KERALA | 683101 | - | Rejected-Technical Technically not Feasible | |
| 3 | Rejected-Technical 3724 OPP KALI BARI MANDIR OPP KALI BARI MANDIR AMBALA CANTT HARYANA 133001 | AMBALA | HARYANA | 133001 | - | Rejected-Technical Technically not Feasible |
Tender Value
Refer Docs
EMD Value
₹23,000
Closing Date
7 Jan 2021, 3:00 pmClosed
Director SRI KWA
State Referral Institute for Water Quality Nettoor Ernakulam 682040
JJM 2020-21 Setting up Up-gradation Operation of Laboratories for WQMS Procurement of 8 nos Micro processor based bench top touch screen pH meter for various Quality Control Laboratories of Kerala Water Authority
2020_KWA_401496_1
11/ 2020-21 / SRI / NTR
Open Tender
Laboratory and scientific equipment
Item Wise
45 days
As per Annexure A of NIT
Please refer Tender documents.
10 documents required · 10 mandatory
₹1,840
₹23,000
Yes
SRI Nettoor
27 Aug 2021
21 Dec 2020
11 Jan 2021
21 Dec 2020
7 Jan 2021
21 Dec 2020
21 Dec 2020 - 26 Dec 2020
28 Dec 2020
Amount
JJM-2020-21-Setting up / upgradation / operation of Laboratories for WQM&S- Procurement of 8 nos Micro Processor based bench top touch screen pH Meter for various Quality Control Laboratories of Kerala Water Authority
Micro processor based bench top touch screen pH meter with the following specifications pH:- Range: -2 to 20.000pH Resolution: 0.001 pH Repeatability: ±0.002pH or better Accuracy: ± 0.002 pH Cal.Points 5 points Buffer options: USA, NIST, Custom ORP:- Range: ± 1999.9 mV or better Resolution: 0.1 mV Repeatability:± 0.2mV or better Accuracy: ± 0.2mV or better Temperature :Range: -5°C to 100°C ; Resolution: 0.1°C Repeatability: ±0.4°C or better Accuracy: ±0.4°C or better Electrode specification:Double junction toughened glass electrode with filling solution Shall provide ATC ;Meter and electrode shall be of same manufacturer Meter Features:(i)Should be touch screen model (ii) Graphical LCD (iii)Electrode status on screen display (iv) Calibration due alarm (v) Off set/slope should be displayed (vi) Data logging facility (vii) Output RS232/USB (viii) Memory 2000 data sets minimum (ix)Power 230V ±10%, 50Hz minimum (x)Buffer set: USA pH 4.01, 7.00, 10.01 (minimum 100ml each) (xi) Should provide electrode filling solution (minimum 100ml) (xii) Should supply non corrosive electrode stand (xiii) Compliance: GLP with NABL Calibration certificate (xiv)Should provide hard copy of user manual with instrument and electrode specification (xv) IQ/OQ/PQ documentation shall be provided (xvi) Warranty: 3 years for the meter and 1 year for the electrode
stage.html
html • 0.05 MB
tech_bid_open.pdf
tech_eval.pdf
fin_bid_open.pdf
boq_comp_chart.xlsx
xlsx
fin_eval.pdf
aoc.pdf
Tap a document below to read it instantly. You can also download everything as a ZIP if you prefer.
details.html
html • 0.04 MB
Download all tender documents and submit your bid
Disclaimer: TenderKart has made every reasonable effort to ensure that the information on this page is accurate and authentic, however it cannot be held liable for any third-party claims or losses or any damages. TenderKart makes no warranty, expressed or implied, as to the results obtained from the use of this information. If you notice any error or omission, please let us know at .