RequiredCertificate (Requested in ATC) Bid to RA enabled No Multipara Monitor - Low End ( 2 pieces ) Technical Specifications * As per GeM Category Specification Specification Specification Name Values Bid Requirement (Allowed Values) Standards Conformity to Certifications - * Confirmity to Manufacturing Standards ISO 9001 & ISO 13485 * Conformity to Safety Standards ISO 80601-2-49 and IEC 60601-1 or BIS equivalent * Performance Purpose Low-end monitors are * 1/7 Parameters designed to perform basic vital-signs monitoring. These monitors may also be appropriate for IPD as well as OPD patients Type of Monitor Preconfigured/Non Modular Preconfigured/Non Modular Monitoring Parameters ECG
RequiredHR
RequiredNIBP
RequiredPR
Requiredrespiratory rate
RequiredSpO2
Requiredand body temperature * ECG parameters Number of leads of ECG 5 5 12-lead interpretation for ECG No No Arrhythmia detection and analysis Yes * Number of leads analyzed for ECG in single time 5 5 Heart rate measurement range At-least 30 -250 BPM * Accuracy ± 5 BPM * Minimum gradation 1 bpm * Respiration Parameters Respiration Method Impedance pneumography * Display waveform of respiration Yes * Threshold control for respiration Yes * Respiration rate measurement range At-least 0-100 bpm * Minimum gradation of respiratory measurement 1 BPM * Pulse Oximetry Parameters Measurement technique Others Others Other Measuring Techniques (In case measurement technique is nelcor
Requiredmasimo
Requireddraeger
Requiredoxisure
Requiredtrusignal put NA otherwise indicate the measurement technique) - * Probe type supplied Adult Adult 2/7 Type of adult probe Reusable Reusable Type of pediatric probe NA NA Type of neonatal probe NA NA SPO2 measurement range Atleast 40-70 % & 70 to 99% * Accuracy of SPO2 measurement Better than ±1 % for 40-70 range & better than ± 3 % for 70-99 range * Minimum Gradation 1% * NIBP Parameters Measurement technique for NIBP Step-deflation oscillometric * Cuff supplied Adult
RequiredPediatric Adult
RequiredPediatric Cuff size for Adult Large Large Cuff size for pediatric Standard Standard Cuff size for neonate Not supplied Not supplied Blood pressure monitoring range 30-300mm/Hg * Gradation 1 mm/Hg or better * Temperature Parameters Number of channels for temperature 1 1 Probe type skin skin Temperature probe type Reusable * Temperature probe range Atleast 30 - 40 degree * Minimum gradation of temp probe 1/10 degreeC * Accuracy in % ±1% * Length of time of full disclosure of patient data in hr ≥48 * Operating conditions Capable of operating continuously in ambient temperature of 0 to 50 de c and relative humidity of 15 to 90% in ideal circumstances * Length of time of trending in hr ≥24 * Alarms Opeartor can set audio Yes * 3/7 visual alarm levels for low or high levels of each parameter independently Systolic BP - low and high Yes * Diastolic BP - low and high Yes * Mean BP - low and high Yes * Heart rate - low and high Yes * SPO2 - low and high Yes * Respiratory rate - low and high Yes * Temperature -low and high Yes * Alarm Volume should be adjustable Yes * Type of alarm Audio and Visual * Activating silence of alarm Yes * Duration of alarm silence in sec 120 120 Suspension of all alarms Yes * Latching of alarms Yes * Cuff leakage alarm Yes * Cuff disconnect alarm Yes * Alarm for inflation/deflation errors Yes * Low battery voltage indication Yes * Sensor/wire/probe disconnection alarm Yes * Networking Parameters Ethernet port Yes * Wireless No No USB port Yes Yes Connectivity provision with Central Nursing Station Yes Yes Display Parameters Type of display TFT TFT Display Size
Requiredinch with ±5% tolerance 10 10 4/7 Traces of display 6 or more * User Interface of display Both (Rotary knob and Touchscreen) Both (Rotary knob and Touchscreen) Inbuilt thermal Recorder (Printer) No No Display must allow easy viewing at all ambient light levels Yes * Standard Accessories Resusable SpO2 probe Adult Adult Number of adult probe 1 1 Number of pediatric probe NA NA Number of neonatal probe NA NA ECG cable 5 lead 5 lead Number of ECG cable 1 1 Skin and Rectal temperature probe 1 * Additional Parameters Accessories to wall mount the monitor Yes Yes Weight of the Machine in kg 3 * Noise level in dB 40 * Body of the monitor to be strong
Requireddurable and splash proof Yes * Should be supplied with protective case for clean storage and safe transport Yes * Power Requirements Type of power supply in VAC
RequiredHz 220-240
Required50Hz Single Phase * Mains cable to be atleast 3m length Indian type 3 pin Yes * Type of battery backup Lithium-ion
Requiredrechargeable Lithium-ion
Requiredrechargeable Backup time in minutes 120 120 Recharging time of battery in hr ≤6 ≤6 Warranty
RequiredServicing and Installation Original User Technical manuals to be supplied in English Yes * 5/7 in hard and soft copy Installation
RequiredDemonstration and training to be provided at consignee end at any part of country Yes * The Principal Manufacturer Must Have Direct Presence/approved service center In India Yes * Advanced maintenance task required shall be documented and procedures required for local calibration and routine maintenence and list of imp spare parts with accessories along with part numbers to be furnished Yes * Availability of toll free facility for technical support maintained by OEM or authorized agencies No No Comprehensive Warranty in Years 3 3 Copies of all certifications and reports to be provided to buyer on demand at time of supplies Yes * * Specifications highlighted in bold are the Golden Parameters. * Bidders may note that In respect of non-golden Parameters
Requiredthe specifications 'Values' chosen by Buyer will generally be preferred over 'Bid requirement ( allowed Values) by the Buyer. Additional Specification Documents